First application of computed radiology to mammography with synchrotron radiation

Primo utilizzo di un sistema di computed radiography in mammografia con luce di sincrotrone

  • Breast Radiology / Senologia
  • Published:
La radiologia medica Aims and scope Submit manuscript

Abstract

Objective

The aim of this study was to evaluate the feasibility of phase-contrast mammography with synchrotron radiation using a high-resolution computed radiology (CR) system devoted to mammography.

Materials and methods

The study was performed at the Synchrotron Radiation for Medical Physics (SYRMEP) beamline of the Elettra synchrotron radiation (SR) facility in Trieste (Italy); X-ray beams were in the range 16–22 keV with a high degree of monochromaticity and spatial coherence. The CR system evaluated is the FCR Profect CS by Fujifilm Global. The first images were obtained from test objects and surgical breast specimens. Images obtained using SR and both screen-film and the CR system were compared with images of the same samples acquired with digital mammography equipment. In view of the good quality of the results obtained, the CR system was used in two mammographic examinations with SR.

Results

Images acquired using SR and both screen-film and CR were obtained with the same level of delivered dose. Image quality obtained with CR was similar or superior to that of screen-film images. Moreover, the digital images obtained with SR were always better than those acquired using the digital mammography system.

Conclusions

Phase-contrast mammography with SR using the studied CR system is a feasible option.

Riassunto

Obiettivo

Lo scopo di questo studio è valutare se sia possibile effettuare mammografie con luce di sincrotrone in contrasto di fase, impiegando come rivelatore un sistema di computed radiography (CR) mammografico ad elevata risoluzione spaziale.

Materiali e metodi

Lo studio è stato condotto presso le linea Synchrotron Radiation for Medical Physics (SYRMEP; Elettra, Trieste) utilizzando fasci di raggi X monocromatici, ad alta coerenza spaziale, con energia compresa tra 16 e 22 keV. Il sistema CR è il FCR Profect CS della Fujifilm Global. Fantocci e campioni chirurgici sono stati impiegati per le prove iniziali; immagini ottenute a SYRMEP su pellicola e su CR sono state confrontate con immagini ottenute presso un mammografo digitale. Visti i buoni risultati, il sistema CR in esame è stato utilizzato per effettuare mammografie su due pazienti.

Risultati

Le immagini acquisite a SYRMEP su pellicola e su CR sono state ottenute a parità di dose; la qualità delle immagini CR è risultata confrontabile o superiore a quella con pellicola. Inoltre le immagini CR ottenute a SYRMEP sono sempre di qualità superiore a quelle ottenute con il mammografo digitale.

Conclusioni

Il sistema CR analizzato risulta adeguato per eseguire mammografie in contrasto di fase.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References/Bibliografia

  1. Gao D, Pogany A, Stevenson AW, Wilkins SK (1998) Phase-contrast radiography. RadioGraphics 18:1257–1267

    PubMed  CAS  Google Scholar 

  2. Arfelli F, Bonvicini V, Bravin A et al (2000) Mammography with synchrotron radiation: phase-detection techniques. Radiology 215:286–293

    PubMed  CAS  Google Scholar 

  3. Karellas A, Vedantham S (2008) Breast cancer imaging: a perspective for the next decade. Med Phys 35:4878–4897

    Article  PubMed  Google Scholar 

  4. Keyrilainen J, Bravin A, Fernandez M et al (2010) Phase-contrast X-ray imaging of breast. Acta Radiologica 8:866–884

    Article  Google Scholar 

  5. Fitzgerald R (2000) Phase-sensitive x-ray imaging. Phys Today 53:23–26

    Article  Google Scholar 

  6. Snigirev A, Snigireva I, Kohn V et al (1995) On the possibilities of x-ray phase contrast microimaging by coherent high-energy synchrotron radiation. Rev Sci Instrum 66:5486–5492

    Article  CAS  Google Scholar 

  7. Wilkins SW, Gureyev TE, Gao D et al (1996) Phase-contrast imaging using polychromatic hard X-rays. Nature 384:335–338

    Article  CAS  Google Scholar 

  8. Dreossi D, Abrami A, Arfelli F et al (2008) The mammography project at the SYRMEP beamline. EJR 68S:S58–S62

    Article  Google Scholar 

  9. Tromba G, Cova M, Castelli E (2011) Phase-contrast mammography at the SYRMEP beamline of Elettra. Synchrotron Radiation News 24:3–7

    Article  Google Scholar 

  10. Castelli E, Tonutti M, Arfelli F et al (2011) Mammography with synchrotron radiation: first clinical experience with phase-detection technique. Radiology 259:684–694

    Article  PubMed  Google Scholar 

  11. Arfelli F, Assante M, Bonvicini V et al (1998) Low-dose phase contrast X-ray medical imaging. Phys Med Biol 43:2845–2852

    Article  PubMed  CAS  Google Scholar 

  12. Kodera Y, Takamura M, Tsuboi E et al (2004) Determination of imaging performance of a digital mammography. Proceedings of SPIE. The International Society for Optical Engineering 5368:743–750

    Article  Google Scholar 

  13. Burns DT, Toni MP, Bovi M (1999) Comparison of the air-kerma standards of the ENEA-INMRI and the BIPM in the low-energy X-ray range. Rapport BIPM-99/11 http://www.bipm.org/utils/common/pdf/rapportBIPM/1999/11.pdf. Last access April 2012

  14. Bovi M, Laitano RF, Pimpinella M et al (2007) Absolute air-kerma measurement in a synchrotron radiation beam by ionization free-air chamber. Proceedings of the Absorbed Dose and Air Kerma Primary Standards Workshop, Paris, France, May 9–11, 2007. http://www.nucleide.org/ADAKPS_WS/Session%20G%20-%20Special%20Standards/G4_Or-Toni.pdf. Last access April 2012

  15. Bovi M, Laitano RF, Pimpinella M et al (2009) Misure assolute di kerma in aria del fascio di luce di sincrotrone prodotto presso l’impianto Elettra di Trieste per applicazioni nella diagnostica medica ad alta risoluzione. Proceedings of the 6th Conference “Metrologia & Qualità”, Turin, Italy, 7–9 April 2009

  16. Boone JM (2002) Normalized glandular dose (DgN) coefficients for arbitrary X-ray spectra in mammography: Computer-fit values of Monte Carlo derived data. Med Phys 29:869–875

    Article  PubMed  Google Scholar 

  17. Dance DR, Skinner CL, Young KC et al (2000) Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol Phys Med Biol 45:3225–3240

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to R. Longo.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Quai, E., Longo, R., Zanconati, F. et al. First application of computed radiology to mammography with synchrotron radiation. Radiol med 118, 89–100 (2013). https://doi.org/10.1007/s11547-012-0847-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11547-012-0847-1

Keywords

Parole chiave

Navigation