Randomised short-term trial of high-span versus low-span APAP for treating sleep apnoea

  • Original Article
  • Published:
Sleep and Breathing Aims and scope Submit manuscript

Abstract

Purpose

Auto-titrating continuous positive airway pressure (APAP) devices were developed to improve treatment efficacy and compliance in patients with obstructive sleep apnoea syndrome (OSAS). Since there are insufficient data on the optimal pressure range setting, we aimed to compare the adherence, efficacy and tolerability of treatment with high-span versus low-span APAP.

Methods

Seventy-six newly diagnosed OSAS patients fulfilling the treatment criteria were randomised to receive high-span (HS, range 4–15cmH2O, n = 38) or low-span (LS, range 8–12cmH2O, n = 38) APAP. Patients were assessed at 1 and 3 months.

Results

Median Epworth sleepiness scale (ESS) was 13 (IQR, 6–16) and median apnoea-hypopnoea index (AHI) was 35.9 (IQR, 27.6–56.3). There were no significant differences in baseline demographic and clinical characteristics between groups. Overall, no significant differences were found at the first month assessment. After 3 months of therapy, we found again no differences in residual AHI or ESS. However, the group HS proved less adherent than group LS, respectively, with median 87 % (IQR, 60.5–97.5) versus 94 % (IQR, 80.0–98.3) of the nights using ≥4 h (P = 0.014) and mean (±SD) usage 5.7 ± 1.6 versus 6.4 ± 1.2 h/night (P = 0.049). The group HS reported more frequently nasal congestion, excessive oronasal dryness and nocturnal awakenings of at least moderate intensity, the latter with statistical significance (P = 0.005).

Conclusions

Both pressure ranges appear to be equally effective to correct AHI and to improve symptoms. Though, patients with high-span APAP were less compliant to treatment, raising issues about the tolerability of wide pressure range settings of these devices.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Young T, Palta M, Dempsey J, Skatrud J, Weber S, Badr S (1993) The occurrence of sleep-disordered breathing among middle-aged adults. N Engl J Med 328(17):1230–1235

    Article  CAS  PubMed  Google Scholar 

  2. McNamara SG, Grunstein RR, Sullivan CE (1993) Obstructive sleep apnea. Thorax 48(7):754–764

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  3. Dong JY, Zhang YH, Qin LQ (2013) Obstructive sleep apnea and cardiovascular risk: meta-analysis of prospective cohort studies. Atherosclerosis 229(2):489–495

    Article  CAS  PubMed  Google Scholar 

  4. Castronovo V, Scifo P, Castellano A, Aloia MS, Iadanza A, Marelli S, Cappa SF, Strambi LF, Falini A (2014) White matter integrity in obstructive sleep apnea before and after treatment. Sleep 37(9):1465–1475

    PubMed  PubMed Central  Google Scholar 

  5. Montserrat JM, Ferrer M, Hernandez L, Farre R, Vilagut G, Navajas D, Badia JR, Carrasco E, de Pablo J, Ballester E (2001) Effectiveness of CPAP treatment in daytime function in sleep apnea syndrome: a randomized controlled study with an optimized placebo. Am J Respir Crit Care Med 164(4):608–613

    Article  CAS  PubMed  Google Scholar 

  6. McNicholas WT (2007) Cardiovascular outcomes of CPAP therapy in obstructive sleep apnea syndrome. Am J Physiol Regul Integr Comp Physiol 293(4):R1666–1670

    Article  CAS  PubMed  Google Scholar 

  7. Engleman HM, Martin SE, Kingshott RN, Mackay TW, Deary IJ, Douglas NJ (1998) Randomized placebo controlled trial of daytime function after continuous positive airway pressure (CPAP) therapy for the sleep apnea/hypopnea syndrome. Thorax 53(5):341–345

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  8. Karimi M, Hedner J, Häbel H, Nerman O, Grote L (2014) A Sleep Apnea-Related Risk of Motor Vehicle Accidents is Reduced by Continuous Positive Airway Pressure: Swedish Traffic Accident Registry Data. Sleep

  9. Berthon-Jones M (1993) Feasibility of a self-setting CPAP machine. Sleep 16(8):S120–123

    CAS  PubMed  Google Scholar 

  10. Behbehani K, Yen F-C, Burk JR, Lucas EA, Axe JR (1995) Automatic control of airway pressure for treatment of obstructive sleep apnea. IEEE Trans Biomed Eng 42(10):1007–1015

    Article  CAS  PubMed  Google Scholar 

  11. Konermann M, Sanner B, Vyleta M, Laschewski F, Groetz J, Sturm A, Zidek W (1998) Use of conventional and self-adjusting nasal continuous positive airway pressure for treatment of severe obstructive sleep apnea syndrome. Chest 113(3):714–718

    Article  CAS  PubMed  Google Scholar 

  12. Massie CA, McArdle N, Hart RW, Schmidt-Nowara WW, Lankford A, Hudgel DW, Gordon N, Douglas NJ (2003) Comparison between automatic and fixed positive airway pressure therapy in the home. Am J Respir Crit Care Med 167(1):20–23

    Article  PubMed  Google Scholar 

  13. Senn O, Brack T, Matthews F, Russi EW, Bloch KE (2003) Randomized short-term trial of two autoCPAP devices versus fixed continuous positive airway pressure for the treatment of sleep apnea. Am J Respir Crit Care Med 168(12):1506–1511

    Article  PubMed  Google Scholar 

  14. Lofaso F, Desmarais G, Leroux K, Zalc V, Fodil R, Isabey D, Louis B (2006) Bench evaluation of flow limitation detection by automated continuous positive airway pressure devices. Chest 130(2):343–349

    Article  PubMed  Google Scholar 

  15. Abdenbi F, Chambille B, Escourrou P (2004) Bench testing of auto-adjusting positive airway pressure devices. Eur Respir J 24(4):649–658

    Article  CAS  PubMed  Google Scholar 

  16. Smith I, Lasserson TJ (2009) Pressure modification for improving usage of continuous positive airway pressure machines in adults with obstructive sleep apnoea. Cochrane Database Syst Rev (4). doi:10.1002/14651858.CD003531.pub3

  17. Ip S, D’Ambrosio C, Patel K, Obadan N, Kitsios GD, Chung M, Balk EM (2012) Auto-titrating versus fixed continuous positive airway pressure for the treatment of obstructive sleep apnea: a systematic review with meta-analyses. Syst Rev 1 (20). doi:10.1186/2046-4053-1-20

  18. Morgenthaler TI, Aurora RN, Brown T, Zak R, Alessi C, Boehlecke B, Chesson AL Jr, Friedman L, Kapur V, Maganti R, Owens J, Pancer J, Swick TJ, Standards of Practice Committee of the AASM (2008) Practice parameters for the use of autotitrating continuous positive airway pressure devices for titrating pressures and treating adult patients with obstructive sleep apnea syndrome: an update for 2007. Sleep 31(1):141–147

    PubMed  PubMed Central  Google Scholar 

  19. American Academy of Sleep Medicine (2005) International classification of sleep disorders. Diagnostic and coding manual, 2nd edn. American Academy of Sleep Medicine, Westchester

    Google Scholar 

  20. Epstein LJ, Kristo D, Strollo PJ Jr, Friedman N, Malhotra A, Patil SP, Ramar K, Rogers R, Schwab RJ, Weaver EM, Weinstein MD, Adult Obstructive Sleep Apnea Task Force of the American Academy of Sleep Medicine (2009) Clinical guideline for the evaluation, management and long-term care of obstructive sleep apnea in adults. J Clin Sleep Med 5(3):263–276

    Google Scholar 

  21. Johns MW (1991) A new method for measuring daytime sleepiness: the Epworth sleepiness scale. Sleep 14(6):540–545

    CAS  PubMed  Google Scholar 

  22. Soares Pires F, Drummond M, Marinho A, Sampaio R, Pinto T, Gonçalves M, Neves I, Pinto C, Sucena M, Winck JC, Almeida J (2013) Effectiveness of a group education session on adherence with APAP in obstructive sleep apnea–a randomized controlled study. Sleep Breath 17(3):993–1001

    Article  PubMed  Google Scholar 

  23. Sampaio RS, Pereira MG, Winck JC (2012) Adaptation of the sleep apnea quality of life index (SAQLI) to Portuguese obstructive sleep apnea syndrome patients. Rev Port Pneumol 18(4):166–174

    Article  CAS  PubMed  Google Scholar 

  24. Kribbs NB, Pack AI, Kline LR, Smith PL, Schwartz AR, Schubert NM, Redline S, Henry JN, Getsy JE, Dinges DF (1993) Objective measurement of patterns of nasal CPAP use by patients with obstructive sleep apnea. Am Rev Respir Dis 147(4):887–895

    Article  CAS  PubMed  Google Scholar 

  25. Ryan S, Doherty LS, Nolan GM, McNicholas WT (2009) Effects of heated humidification and topical steroids on compliance, nasal symptoms, and quality of life in patients with obstructive sleep apnea syndrome using nasal continuous positive airway pressure. J Clin Sleep Med 5(5):422–427

    PubMed  PubMed Central  Google Scholar 

  26. Sullivan CE, Issa FG, Berthon-Jones M, Eves L (1981) Reversal of obstructive sleep apnea by continuous airway pressure applied through the nares. Lancet 1(8225):862–865

    Article  CAS  PubMed  Google Scholar 

  27. Patel SR, White DP, Malhotra A, Stanchina ML, Ayas NT (2003) Continuous positive airway pressure therapy for treating sleepiness in a diverse population with obstructive sleep apnea: results of a meta-analysis. Arch Intern Med 163(5):565–571

    Article  PubMed  Google Scholar 

  28. Zimmerman ME, Arnedt JT, Stanchina M, Millman RP, Aloia MS (2006) Normalization of memory performance and positive airway pressure adherence in memory-impaired patients with obstructive sleep apnea. Chest 130(6):1772–1778

    Article  PubMed  Google Scholar 

  29. Dantas AP, Winck JC, Figueiredo-Braga M (2014) Adherence to APAP in obstructive sleep apnea syndrome: effectiveness of a motivational intervention. Sleep Breath. doi:10.1007/s11325-014-1024-7

    PubMed  Google Scholar 

  30. Valentin A, Subramanian S, Quan SF, Berry RB, Parthasarathy S (2011) Air leak is associated with poor adherence to autoPAP therapy. Sleep 34(6):801–806

    Article  PubMed  PubMed Central  Google Scholar 

Download references

Acknowledgments

We thank Cristina Carrondo, Elisabete Santa Clara, Ermelinda Eusébio, Joana Pipa, Patrícia Dantas, Paulo Viana and Teresa Santiago for selecting the patients, performing the sleep studies and manually evaluating the records.

Conflict of interest

T Pinto has received financial support from Linde and Vitalaire (Healthcare Providers) for attending symposia and honoraria for speaking at symposia from Philips. After the conclusion of the study, JC Winck has started working in a global position for Linde. The remaining authors declare that they have no conflict of interest.

Ethical approval

The study protocol was approved by our institution’s Ethics Committee. All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Helder Novais Bastos.

Additional information

Comment

In the era of increasing long-term auto-titrating positive airway pressure (APAP) device use for OSA treatment, this study helps sleep physicians improve adherence without substantive effects on OSA control. This randomised controlled trial evaluates the effect of high-span (4–15 cm water) and low-span (8–12 cm water) APAP initial settings on efficacy and adherence over a 3-month time span. The authors find that there is no difference in efficacy between the treatment arms and improved adherence with the low-span APAP settings. These findings indicate that there may be significant differences in long-term adherence based on initial PAP therapy settings. Long-term studies are needed to further clarify the issue; however, it may be prudent in selected patients at the start of therapy to do a trial of low-span APAP with short-term follow-up.

Anna May

Cleveland, OH

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Bastos, H.N., Cardoso, A.V., Castro, A.S. et al. Randomised short-term trial of high-span versus low-span APAP for treating sleep apnoea. Sleep Breath 20, 183–190 (2016). https://doi.org/10.1007/s11325-015-1203-1

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11325-015-1203-1

Keywords

Navigation