Evaluation of auto bi-level algorithm to treat pressure intolerance in obstructive sleep apnea

  • Original Article
  • Published:
Sleep and Breathing Aims and scope Submit manuscript

Abstract

Purpose

The objective of this study was to evaluate whether a new auto-adjusting bi-level algorithm was comparable to a standard method for prescribing bi-level therapy.

Methods

This study was a prospective randomized, double-blinded crossover evaluation of the equivalency of the auto-adjusting bi-level mode (VAuto™) compared to standard bi-level mode, using a pre-determined difference in Apnea–Hypopnea Index (AHI) of five events per hour. Data were obtained during sleep studies performed on two separate nights. Twenty-two subjects met the entry criteria and were enrolled in the study at four investigational sites in the USA.

Results

Mean AHI for the auto-adjusting bi-level mode was 6.2 ± 5.4 events per hour and for the standard bi-level mode 8.3 ± 5.8 events per hour. The AHI for the two modes were clinically equivalent. The difference in median pressure between these two modes was −3.8 cm H2O ± 3.6 (p = 0.0008) in favor of the auto-adjusting bi-level mode. In addition, the maximum pressure was significantly higher in the auto-adjusting bi-level mode (16.0 cm H2O vs. 14.1 cm H2O, p = 0.02).

Conclusions

Our results demonstrated that the auto-adjusting bi-level mode normalized AHI comparable to the standard bi-level mode. The results of this study have several significant implications for the clinical management of sleep apnea. Obstructive sleep apnea (OSA) is a common condition and is associated with untoward complications. Non-compliance with positive airway pressure (PAP) limits the efficacy of the PAP therapy. The auto-adjusting bi-level mode provides a potentially reliable alternative for sleep clinicians faced with prescribing bi-level PAP for non-compliant patients. This study documents that this type of auto-adjusting device provides effective treatment of OSA.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Pepperell JC, Ramdassingh-Dow S, Crosthwaite N, Mullins R, Jenkinson C, Stradling JR, Davies RJ (2002) Ambulatory blood pressure after therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: a randomised parallel trial. Lancet 359:204–210

    Article  PubMed  Google Scholar 

  2. White DP (2006) Sleep apnea. Proc Am Thorac Soc 3:124–128

    Article  PubMed  CAS  Google Scholar 

  3. Caples SM, Gami AS, Somers VK (2005) Obstructive sleep apnea. Ann Intern Med 142:187–197

    PubMed  Google Scholar 

  4. Shahar E, Whitney CW, Redline S, Lee ET, Newman AB, Javier Nieto F, O'Connor GT, Boland LL, Schwartz JE, Samet JM (2001) Sleep-disordered breathing and cardiovascular disease: cross-sectional results of the Sleep Heart Health Study. Am J Respir Crit Care Med 163:19–25

    PubMed  CAS  Google Scholar 

  5. Punjabi NM, Caffo BS, Goodwin JL, Gottlieb DJ, Newman AB, O'Connor GT, Rapoport DM, Redline S, Resnick HE, Robbins JA, Shahar E, Unruh ML, Samet JM (2009) Sleep-disordered breathing and mortality: a prospective study. PloS Med 6(8):e1000132. (http://www.plosmedicine.org/article/info%3Adoi%2F10.1371%2Fjournal.pmed.1000132) Accessed 03/26/2010

    Google Scholar 

  6. Becker HF, Jerrentrup A, Ploch T, Grote L, Penzel T, Sullivan CE, Peter JH (2003) Effect of nasal continuous positive airway pressure treatment on blood pressure in patients with obstructive sleep apnea. Circulation 107:68–73

    Article  PubMed  Google Scholar 

  7. Doherty LS, Kiely JL, Swan V, McNicholas WT (2005) Long-term effects of nasal continuous positive airway pressure therapy on cardiovascular outcomes in sleep apnea syndrome. Chest 127:2076–2084

    Article  PubMed  Google Scholar 

  8. Engleman HM, Douglas NJ (2004) Sleep 4: sleepiness, congestive function, and quality of life in obstructive sleep apnoea/hypopnoea syndrome. Thorax 59:618–622

    Article  PubMed  CAS  Google Scholar 

  9. Gay P, Weaver T, Loube D, Iber C, Positive Airway Pressure Task Force, Standards of Practice Committee, American Academy of Sleep Medicine (2006) Evaluation of positive airway pressure treatment for sleep related breathing disorders in adults. Sleep 29:381–401

    PubMed  Google Scholar 

  10. Weaver TE, Grunstein RR (2008) Adherence to continuous positive airway pressure therapy: the challenge to effective treatment. Proc Am Thorac Soc 5:173–178

    Article  PubMed  Google Scholar 

  11. Stepnowsky CJ, Marler MR, Palau J, Brooks JA (2006) Social-coginitive correlates of CPAP adherence in experienced users. Sleep Med 4:350–356

    Article  Google Scholar 

  12. Weaver TE, Maislin G, Dinges DF, Bloxham T, George CF, Greenberg H, Kader G, Mahowald M, Younger J, Pack AL (2007) Relationship between hours of CPAP use and achieving normals levels of sleepiness and daily functioning. Sleep 30:711–719

    PubMed  Google Scholar 

  13. Reeves-Hoche MK, Hudgel DW, Meck R, Witteman R, Ross A, Zwillich CW (1995) Continuous versus bilevel positive airway pressure for obstructive sleep apnea. Am J Respir Crit Care Med 151:443–449

    PubMed  CAS  Google Scholar 

  14. Kushida CA, Littner MR, Hirshkowitz M, Morgenthaler TI, Alessi CA, Bailey D, Boehlecke B, Brown TM, Coleman JJr, Friedman L, Kapen S, Kapur VK, Kramer M, Lee-Chiong T, Owens J, Pancer JP, Swick TJ, Wise MS, American Academy for Sleep Medicine (2006) Practice parameters for the use of continuous and bilevel positive airway pressure devices to treat adult patients with sleep-related breathing disorders. Sleep 29:375–380

    PubMed  Google Scholar 

  15. Gay PC, Herold DL, Olson EJ (2003) A randomized, double-blind clinical trial comparing continuouspositive airway pressure with a novel bilevel pressure system for treatment of obstructive sleep apnea syndrome. Sleep 26:864–869

    PubMed  Google Scholar 

  16. Ballard RD, Gay PC, Strollo PJ (2007) Interventions to improve compliance in sleep apnea patients previously non-compliant with continuous positive airway pressure. J Clin Sleep Med 3:706–712

    PubMed  Google Scholar 

  17. Chan J, Chan C, Hansford A, Yen J, Chan CS (2006) The clinical efficacy and comfort of modified AutoSet™ compared with standard AutoSet™. Sleep Med 7:S75–S76

    Article  Google Scholar 

  18. Kushida CA, Littner MR, Morgenthaler T, Alessi CA, Bailey D, Coleman JJr, Friedman L, Hirshkowitz M, Kapen S, Kramer M, Lee-Chiong T, Loube D, Owens J, Pancer JP, Wise M (2005) Practice parameters for the indications for polysomnography and related procedures: an update for 2005. Sleep 28:499–519

    PubMed  Google Scholar 

  19. The Report of an American Academy of Sleep Medicine Task Force (1999) Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. Sleep 22:667–689

    Google Scholar 

  20. Parish J, Miller B, Hentz J (2008) Autotitration positive airway pressure therapy in patients with obstructive sleep apnea who are intolerant of fixed continuous positive airway pressure. Sleep Breath 12:235–241

    Article  PubMed  Google Scholar 

  21. Vennelle M, White S, Riha RL, Mackay TW, Engleman HM, Douglas NJ (2010) Randomized controlled trial of variable-pressure versus fixed-pressure Continous Positive Airway Pressure (CPAP) treatment for patients with Obstructive Sleep Apnea/Hypopnea Syndrome (OSAHS). Sleep 33:267–271

    PubMed  Google Scholar 

  22. Bakker J, Campbell A, Neill A (2010) Randomized controlled trial comparing flexible and continuous positive airway pressure delivery: effects on compliance, objective, and subjective sleepiness and vigilance. Sleep 33:523–529

    PubMed  Google Scholar 

Download references

Acknowledgments

All work was supported by a grant from ResMed Corp. The authors gratefully acknowledge the contributions of all the investigators in the sleep centers who contributed to this study (Swedish Medical Center, Darius Rhodes-Zoroufy, MD; Tristate Sleep Disorder Center, David Berkowitz, MD, Martin B Sharf PhD; St John’s Plaza Sleep Disorders Center, Paul Haberman, MD).

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to James Parish.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ball, N., Gordon, N., Casal, E. et al. Evaluation of auto bi-level algorithm to treat pressure intolerance in obstructive sleep apnea. Sleep Breath 15, 301–309 (2011). https://doi.org/10.1007/s11325-010-0381-0

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11325-010-0381-0