Abstract
Purpose
The objective of this study was to evaluate whether a new auto-adjusting bi-level algorithm was comparable to a standard method for prescribing bi-level therapy.
Methods
This study was a prospective randomized, double-blinded crossover evaluation of the equivalency of the auto-adjusting bi-level mode (VAuto™) compared to standard bi-level mode, using a pre-determined difference in Apnea–Hypopnea Index (AHI) of five events per hour. Data were obtained during sleep studies performed on two separate nights. Twenty-two subjects met the entry criteria and were enrolled in the study at four investigational sites in the USA.
Results
Mean AHI for the auto-adjusting bi-level mode was 6.2 ± 5.4 events per hour and for the standard bi-level mode 8.3 ± 5.8 events per hour. The AHI for the two modes were clinically equivalent. The difference in median pressure between these two modes was −3.8 cm H2O ± 3.6 (p = 0.0008) in favor of the auto-adjusting bi-level mode. In addition, the maximum pressure was significantly higher in the auto-adjusting bi-level mode (16.0 cm H2O vs. 14.1 cm H2O, p = 0.02).
Conclusions
Our results demonstrated that the auto-adjusting bi-level mode normalized AHI comparable to the standard bi-level mode. The results of this study have several significant implications for the clinical management of sleep apnea. Obstructive sleep apnea (OSA) is a common condition and is associated with untoward complications. Non-compliance with positive airway pressure (PAP) limits the efficacy of the PAP therapy. The auto-adjusting bi-level mode provides a potentially reliable alternative for sleep clinicians faced with prescribing bi-level PAP for non-compliant patients. This study documents that this type of auto-adjusting device provides effective treatment of OSA.
Similar content being viewed by others
References
Pepperell JC, Ramdassingh-Dow S, Crosthwaite N, Mullins R, Jenkinson C, Stradling JR, Davies RJ (2002) Ambulatory blood pressure after therapeutic and subtherapeutic nasal continuous positive airway pressure for obstructive sleep apnoea: a randomised parallel trial. Lancet 359:204–210
White DP (2006) Sleep apnea. Proc Am Thorac Soc 3:124–128
Caples SM, Gami AS, Somers VK (2005) Obstructive sleep apnea. Ann Intern Med 142:187–197
Shahar E, Whitney CW, Redline S, Lee ET, Newman AB, Javier Nieto F, O'Connor GT, Boland LL, Schwartz JE, Samet JM (2001) Sleep-disordered breathing and cardiovascular disease: cross-sectional results of the Sleep Heart Health Study. Am J Respir Crit Care Med 163:19–25
Punjabi NM, Caffo BS, Goodwin JL, Gottlieb DJ, Newman AB, O'Connor GT, Rapoport DM, Redline S, Resnick HE, Robbins JA, Shahar E, Unruh ML, Samet JM (2009) Sleep-disordered breathing and mortality: a prospective study. PloS Med 6(8):e1000132. (http://www.plosmedicine.org/article/info%3Adoi%2F10.1371%2Fjournal.pmed.1000132) Accessed 03/26/2010
Becker HF, Jerrentrup A, Ploch T, Grote L, Penzel T, Sullivan CE, Peter JH (2003) Effect of nasal continuous positive airway pressure treatment on blood pressure in patients with obstructive sleep apnea. Circulation 107:68–73
Doherty LS, Kiely JL, Swan V, McNicholas WT (2005) Long-term effects of nasal continuous positive airway pressure therapy on cardiovascular outcomes in sleep apnea syndrome. Chest 127:2076–2084
Engleman HM, Douglas NJ (2004) Sleep 4: sleepiness, congestive function, and quality of life in obstructive sleep apnoea/hypopnoea syndrome. Thorax 59:618–622
Gay P, Weaver T, Loube D, Iber C, Positive Airway Pressure Task Force, Standards of Practice Committee, American Academy of Sleep Medicine (2006) Evaluation of positive airway pressure treatment for sleep related breathing disorders in adults. Sleep 29:381–401
Weaver TE, Grunstein RR (2008) Adherence to continuous positive airway pressure therapy: the challenge to effective treatment. Proc Am Thorac Soc 5:173–178
Stepnowsky CJ, Marler MR, Palau J, Brooks JA (2006) Social-coginitive correlates of CPAP adherence in experienced users. Sleep Med 4:350–356
Weaver TE, Maislin G, Dinges DF, Bloxham T, George CF, Greenberg H, Kader G, Mahowald M, Younger J, Pack AL (2007) Relationship between hours of CPAP use and achieving normals levels of sleepiness and daily functioning. Sleep 30:711–719
Reeves-Hoche MK, Hudgel DW, Meck R, Witteman R, Ross A, Zwillich CW (1995) Continuous versus bilevel positive airway pressure for obstructive sleep apnea. Am J Respir Crit Care Med 151:443–449
Kushida CA, Littner MR, Hirshkowitz M, Morgenthaler TI, Alessi CA, Bailey D, Boehlecke B, Brown TM, Coleman JJr, Friedman L, Kapen S, Kapur VK, Kramer M, Lee-Chiong T, Owens J, Pancer JP, Swick TJ, Wise MS, American Academy for Sleep Medicine (2006) Practice parameters for the use of continuous and bilevel positive airway pressure devices to treat adult patients with sleep-related breathing disorders. Sleep 29:375–380
Gay PC, Herold DL, Olson EJ (2003) A randomized, double-blind clinical trial comparing continuouspositive airway pressure with a novel bilevel pressure system for treatment of obstructive sleep apnea syndrome. Sleep 26:864–869
Ballard RD, Gay PC, Strollo PJ (2007) Interventions to improve compliance in sleep apnea patients previously non-compliant with continuous positive airway pressure. J Clin Sleep Med 3:706–712
Chan J, Chan C, Hansford A, Yen J, Chan CS (2006) The clinical efficacy and comfort of modified AutoSet™ compared with standard AutoSet™. Sleep Med 7:S75–S76
Kushida CA, Littner MR, Morgenthaler T, Alessi CA, Bailey D, Coleman JJr, Friedman L, Hirshkowitz M, Kapen S, Kramer M, Lee-Chiong T, Loube D, Owens J, Pancer JP, Wise M (2005) Practice parameters for the indications for polysomnography and related procedures: an update for 2005. Sleep 28:499–519
The Report of an American Academy of Sleep Medicine Task Force (1999) Sleep-related breathing disorders in adults: recommendations for syndrome definition and measurement techniques in clinical research. Sleep 22:667–689
Parish J, Miller B, Hentz J (2008) Autotitration positive airway pressure therapy in patients with obstructive sleep apnea who are intolerant of fixed continuous positive airway pressure. Sleep Breath 12:235–241
Vennelle M, White S, Riha RL, Mackay TW, Engleman HM, Douglas NJ (2010) Randomized controlled trial of variable-pressure versus fixed-pressure Continous Positive Airway Pressure (CPAP) treatment for patients with Obstructive Sleep Apnea/Hypopnea Syndrome (OSAHS). Sleep 33:267–271
Bakker J, Campbell A, Neill A (2010) Randomized controlled trial comparing flexible and continuous positive airway pressure delivery: effects on compliance, objective, and subjective sleepiness and vigilance. Sleep 33:523–529
Acknowledgments
All work was supported by a grant from ResMed Corp. The authors gratefully acknowledge the contributions of all the investigators in the sleep centers who contributed to this study (Swedish Medical Center, Darius Rhodes-Zoroufy, MD; Tristate Sleep Disorder Center, David Berkowitz, MD, Martin B Sharf PhD; St John’s Plaza Sleep Disorders Center, Paul Haberman, MD).
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Ball, N., Gordon, N., Casal, E. et al. Evaluation of auto bi-level algorithm to treat pressure intolerance in obstructive sleep apnea. Sleep Breath 15, 301–309 (2011). https://doi.org/10.1007/s11325-010-0381-0
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11325-010-0381-0