Abstract

Purpose

To evaluate in an experimental model, the mobility of a novel peritoneal dialysis catheter with a metallic tip (Mag-Cath) using an extra-corporeal magnet.

Materials and methods

The Mag-Cath was installed in a porcine model under general anesthesia using laparoscopic access. A senior urologist, a urology resident, and a 3rd-year medical student were assigned to complete 2 tasks guided by fluoroscopy: (1) Move the catheter tip from the right flank to the pelvis and (2) move the catheter tip from the pelvis to the left flank. Accomplishment and time to complete the tasks were recorded. Participants were asked independently to grade the difficulty of the tasks using an analog scale from 1 (extremely easy) to 10 (extremely difficult).

Results

All participants completed the tasks. The difficulty of the tasks was graded equal to 1 by the senior urologist and urology resident and equal to 3 by the medical student. The time to move the Mag-Cath from the right flank to the target was 14, 10, and 55 s for the senior urologist, urology resident, and medical student, respectively. The time to move the Mag-Cath from the target to the left flank was 17, 18, and 43 s for the senior urologist, urology resident, and medical student, respectively.

Conclusion

In this preliminary analysis, the use of an extra-corporeal magnet to move the Mag-Cath was feasible, with a low grade of difficulty, and individuals with different levels of expertise could perform it. Further studies in human subjects are needed to evaluate the clinical applicability of the Mag-Cath properly.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Jain AK, Blake P, Cordy P, Garg AX (2012) Global trends in rates of peritoneal dialysis. J Am Soc Nephrol 23:533–544

    Article  Google Scholar 

  2. Mehrotra R, Devuyst O, Davies SJ, Johnson DW (2016) The current state of peritoneal dialysis. J Am Soc Nephrol 27:3238–3252

    Article  CAS  Google Scholar 

  3. Hole B, Caskey F, Evans K et al (2017) UK Renal Registry 19th Annual Report: chapter 12 multisite dialysis access audit in England, Northern Ireland and Wales in 2015 and 2014 PD one year follow-up: national and centre-specific analyses. Nephron 137(suppl 1):269–296

    Article  Google Scholar 

  4. Miller M, McCormick B, Lavoie S, Biyani M, Zimmerman D (2012) Fluoroscopic manipulation of peritoneal dialysis catheters: outcomes and factors associated with successful manipulation. Clin J Am Soc Nephrol 7:795–800

    Article  Google Scholar 

  5. Saka Y, Ito Y, Iida Y, Maruyama S, Matsuo S (2015) Efficacy and safety of fluoroscopic manipulation using the alpha-replacer for peritoneal catheter malposition. Clin Exp Nephrol 19:521–526

    Article  Google Scholar 

  6. Ögünç G (2002) Malfunctioning peritoneal dialysis catheter and accompanying surgical pathology repaired by laparoscopic surgery. Perit Dial Int 22:454–462

    Article  Google Scholar 

  7. Santarelli S, Zeiler M, Marinelli R, Monteburini T, Federico A, Ceraudo E (2006) Videolaparoscopy as rescue therapy and placement of peritoneal dialysis catheters: a thirty-two case single centre experience. Nephrol Dial Transpl 21:1348–1354

    Article  Google Scholar 

  8. Yilmazlar T, Kirdak T, Bilgin S, Yavuz M, Yurtkuran M (2005) Laparoscopic findings of peritoneal dialysis catheter. Perit Dial Int 26:374–379

    Article  Google Scholar 

  9. Haskins IN, Strong AT, Allemang MT, Bencsath KP, Rodriguez JH, Kroh MD (2018) Magnetic surgery: first U.S. experience with a novel device. Surg Endosc 32:895–899

    Article  Google Scholar 

  10. Steinberg RL, Johnson BA, Meskawi M, Gettman MT, Cadeddu JA (2019) Magnet assisted robotic prostatectomy using the da Vinci SP robot: an initial case series. J Endourol. https://doi.org/10.1089/end.2019.0263

    Article  PubMed  Google Scholar 

  11. Rassweiler MC, Michel MS, Ritter M, Honeck P (2017) Magnetic ureteral stent removal without cystoscopy: a randomized controlled trial. J Endourol 31:762–766

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Victor Srougi.

Ethics declarations

Conflict of interest

None.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Srougi, V., Fonseca, L.M., Zampolli, L.J. et al. Mobilization of a peritoneal dialysis catheter using an extra-corporeal magnet: preliminary experimental phase study. Int Urol Nephrol 53, 185–188 (2021). https://doi.org/10.1007/s11255-020-02594-9

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11255-020-02594-9

Keywords

Navigation