Arm circumference: its importance for dialysis patients in the obesity era

  • Nephrology - Original Paper
  • Published:
International Urology and Nephrology Aims and scope Submit manuscript

Abstract

Purpose

The purposes of this study were to investigate the association between arm circumference and body mass index (BMI) and to discuss problems, mainly arm circumference and cuff size mismatch, that could affect the reliability of home blood pressure monitoring (HBPM) among peritoneal dialysis (PD) and hemodialysis (HD) patients.

Methods

525 PD and 502 HD patients from 16 centers were included in the study. A two-part questionnaire was used to gather information from the participants. Arm circumferences were categorized into four groups according to the British Hypertension Society cuff size recommendations.

Results

Mean BMI and arm circumference of all participants were 25.0 kg/m2 and 27.6 cm, respectively. There was a significant correlation between BMI and arm circumference. The mean BMI and arm circumference values were higher in PD patients than in HD patients. Requirement of a large-sized adult cuff was more common among PD patients compared to HD patients (14 % vs 8 %, p = 0.002).

Conclusions

Since HBPM is a useful tool for clinicians to improve BP control, nephrologists should be aware of the problems related to HBPM in dialysis patients and take an active role to increase the reliability of HBPM.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Goldsmith D, Covic A (2009) Blood pressure control in CKD stage 5D patients-are we more or less certain what to do in 2009? Nephrol Dial Transpl 24:3597–3601

    Article  Google Scholar 

  2. Agarwal R, Nissenson AR, Batlle D, Coyne DW, Trout JR, Warnock DG (2003) Prevalence, treatment, and control of hypertension in chronic hemodialysis patients in the United States. Am J Med 115:291–297

    Article  PubMed  Google Scholar 

  3. Agarwal R, Peixoto AJ, Santos SF, Zoccali C (2009) Out-of-office blood pressure monitoring in chronic kidney disease. Blood Press Monit 14:2–11

    Article  PubMed  Google Scholar 

  4. Pickering TG, Miller NH, Ogedegbe G, Krakoff LR, Artinian NT, Goff D et al (2008) Call to action on use and reimbursement for home blood pressure monitoring: a joint scientific statement from the American Heart Association, American Society of Hypertension, and Preventive Cardiovascular Nurses Association. J Cardiovasc Nurs 23:299–323

    Article  PubMed  Google Scholar 

  5. Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN et al (2005) Recommendations for blood pressure measurement in human and experimental animals. Part: 1: blood pressure measurement in humans. A statement for professionals from the subcommittee of professional and public education of the American Heart Association Council on high blood pressure research. Hypertension 45:142–161

    PubMed  CAS  Google Scholar 

  6. Parati G, Stergiou GS, Asmar R, Bilo G, de Leeuw P, Imai Y et al (2010) European Society of Hypertension practice guidelines for home blood pressure monitoring. J Hum Hypertens 24:779–785

    Article  PubMed  CAS  Google Scholar 

  7. Goldsmith D, Covic A (2011) Ambulatory blood pressure measurements in chronic kidney disease: ready to (rock and) roll? Arch Intern Med 171:1098–1099

    Article  PubMed  Google Scholar 

  8. Agarwal R (2010) Managing hypertension using home blood pressure monitoring among haemodialysis patients–a call to action. Nephrol Dial Transpl 25:1766–1771

    Article  Google Scholar 

  9. British Hypertension Society (2012) http://www.bhsoc.org/how_to_measure_blood_pressure.stm. Accessed 12 Jan 2012

  10. Maxwell MH, Schroth PC, Waks AU, Karam M, Dornfeld LP (1982) Error in blood-pressure measurement due to incorrect cuff size in obese patients. Lancet 2:33–36

    Article  PubMed  CAS  Google Scholar 

  11. Kalaitzidis RG, Siamopoulos KC (2011) The role of obesity in kidney disease: recent findings and potential mechanisms. Int Urol Nephrol 43:771–784

    Article  PubMed  Google Scholar 

  12. Graves JW, Bailey KR, Sheps SG (2003) The changing distribution of arm circumferences in NHANES III and NHANES 2000 and its impact on the utility of the ‘standard adult’ blood pressure cuff. Blood Press Monit 8:223–227

    Article  PubMed  Google Scholar 

  13. Bross R, Chandramohan G, Kovesdy CP, Oreopoulos A, Noori N, Golden S et al (2010) Am J Kidney Dis 55:885–896

    Article  PubMed  Google Scholar 

  14. Noori N, Kovesdy CP, Dukkipati R, Kim Y, Duong U, Bross R et al (2010) Survival predictability of lean and fat mass in men and women undergoing maintenance hemodialysis. Am J Clin Nutr 92:1060–1070

    Article  PubMed  CAS  Google Scholar 

  15. Noori N, Kovesdy CP, Bross R, Lee M, Oreopoulos A, Benner D et al (2011) Novel equations to estimate lean body mass in maintenance hemodialysis patients. Am J Kidney Dis 57:130–139

    Article  PubMed  Google Scholar 

  16. Akpolat T (2010) Obesity, hypertension and home sphygmomanometer cuffs. Eur J Intern Med 21:338–340

    Article  PubMed  Google Scholar 

  17. Akpolat T, Aydogdu T, Erdem E (2011) Ideal bladder length. Int J Cardiol 148:372–373

    Article  PubMed  Google Scholar 

  18. Finucane MM, Stevens GA, Cowan MJ, Danaei G, Lin JK, Paciorek CJ (2011) National, regional, and global trends in body-mass index since 1980: systematic analysis of health examination surveys and epidemiological studies with 960 country-years and 9·1 million participants. Lancet 377:557–567

    Article  PubMed  Google Scholar 

  19. Ricks J, Molnar MZ, Kovesdy CP, Kopple JD, Norris KC, Mehrotra R (2011) Racial and ethnic differences in the association of body mass index and survival in maintenance hemodialysis patients. Am J Kidney Dis 58:574–582

    Article  PubMed  Google Scholar 

  20. Peixoto AJ, Gray TA, Crowley ST (1999) Validation of the SpaceLabs 90207 ambulatory blood pressure device for hemodialysis patients. Blood Press Monit 4:217–221

    PubMed  CAS  Google Scholar 

  21. Association for the Advancement of Medical Instrumentation (1993) American National Standard for Electronic or Automated Sphygmomanometers: ANSI/AAMI SP10—1993. AAMI, Arlington

  22. O’Brien E, Petrie J, Littler W, DeSweet M, Padfield PL, Altman DG et al (1993) The British Hypertension Society Protocol for the evaluation of blood pressure measuring devices. J Hypertens 11(suppl 2):S43–S63

    Google Scholar 

  23. O’Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R (2010) European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit 15:23–38

    Article  PubMed  Google Scholar 

  24. Omboni S, Riva I, Giglio I, Caldara G, Groppelli A, Parati G (2007) Validation of the Omron M5-I, R5-I and HEM-907 automated blood pressure monitors in elderly individuals according to the International Protocol of the European Society of Hypertension. Blood Press Monit 12:233–242

    Article  PubMed  Google Scholar 

  25. Kanbay M, Afsar B, Gusbeth-Tatomir P, Covic A (2010) Arterial stiffness in dialysis patients: where are we now? Int Urol Nephrol 42:741–752

    Article  PubMed  Google Scholar 

  26. Thompson AM, Eguchi K, Reznik ME, Shah SS, Pickering TG (2007) Validation of an oscillometric home blood pressure monitor in an end-stage renal disease population and the effect of arterial stiffness on its accuracy. Blood Press Monit 12:227–232

    Article  PubMed  Google Scholar 

  27. Akpolat T, Erdem E, Aydogdu T (2012) Validation of the Omron M3 Intellisense (HEM-7051-E) Upper Arm Blood Pressure Monitor, for Self-Measurement, according to the European Society of Hypertension International Protocol Revision 2010 in a Stage 3–5 Chronic Kidney Disease Population. Kidney Blood Press Res 35:82–88

    Article  PubMed  Google Scholar 

  28. Semret M, Zidehsarai M, Agarwal R (2005) Accuracy of oscillometric blood pressure monitoring with concurrent auscultatory blood pressure in hemodialysis patients. Blood Press Monit 10:249–255

    Article  PubMed  Google Scholar 

  29. Czarkowski M, Staszkow M, Kostyra K, Shebani Z, Niemczyk S, Matuszkiewicz-Rowińska J (2009) Determining the accuracy of blood pressure measurement by the Omron HEM-907 before and after hemodialysis. Blood Press Monit 14:232–238

    Article  PubMed  Google Scholar 

  30. Imai Y, Otsuka K, Kawano Y, Shimada K, Hayashi H, Tochikubo O et al (2003) Japanese Society of hypertension (JSH) guidelines for self-monitoring of blood pressure at home. Hypertens Res 26:771–782

    Article  PubMed  Google Scholar 

  31. dabl®Educational Trust (2012) http://www.dableducational.org. Accessed 12 Jan 2012

  32. British Hypertension Society (2012) http://www.bhsoc.org/bp_monitors/automatic_wrist.stm. Accessed 12 Jan 2012

  33. Ogihara T, Kikuchi K, Matsuoka H, Fujita T, Higaki J, Horiuchi M et al (2009) The Japanese Society of Hypertension Guidelines for the Management of Hypertension (JSH 2009). Hypertens Res 32:3–107

    PubMed  CAS  Google Scholar 

  34. Obara T, Ohkubo T, Fukunaga H, Kobayashi M, Satoh M, Metoki H (2010) Practice and awareness of physicians regarding home blood pressure measurement in Japan. Hypertens Res 33:428–434

    Article  PubMed  Google Scholar 

  35. Akpolat T, Erdem Y, Derici U, Erturk S, Caglar S, Hasanoglu E et al (2012) Use of home sphygmomanometers in Turkey: a nation-wide survey. Hypertens Res 35:356–361

    Article  PubMed  Google Scholar 

  36. Akpolat T, Aydogdu T, Erdem E, Karatas A (2011) Inaccuracy of home sphygmomanometers: a perspective from clinical practice. Blood Press Monit 16:168–171

    Article  PubMed  Google Scholar 

  37. Coleman A, Steel S, deGreeff A, Shennan A (2010) Validation of the Lloyds pharmacy BP11 oscillometric blood pressure monitor according to the international protocol of the European society of hypertension. Blood Press Monit 15:163–166

    Article  PubMed  Google Scholar 

  38. O’Brien E (2001) State of the market for devices for blood pressure measurement. Blood Press Monit 6:281–286

    Article  PubMed  Google Scholar 

Download references

Conflict of interest

There is no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tekin Akpolat.

Additional information

Tekin Akpolat, Cengiz Utaş, Turgay Arinsoy, Hülya Taşkapan, M. Emin Yilmaz, Rezzan Ataman, Semra Bozfakioğlu, Çetin Özener, İbrahim Karayaylali, Rümeyza Kazancioğlu, Taner Çamsari, Mahmut Yavuz, Fevzi Ersoy, Soner Duman, Kenan Ateş are the members of Turkish Multicenter Peritoneal Dialysis Study Group (TULIP).

Rights and permissions

Reprints and permissions

About this article

Cite this article

Akpolat, T., Kaya, C., Utaş, C. et al. Arm circumference: its importance for dialysis patients in the obesity era. Int Urol Nephrol 45, 1103–1110 (2013). https://doi.org/10.1007/s11255-012-0219-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11255-012-0219-8

Keywords

Navigation