The possibility of using mobile phases with ion-pairing reagents for HPLC determination of benzethonium chloride in eye drops was studied. The most suitable mobile phases among those studied were acetonitrile:aqueous tetrabutylammonium hydrophosphate (5 mM) (65:35) and acetonitrile:aqueous tetrabutylammonium hydrophosphate (2.5 mM) and sodium heptanesulfonate (2.5 mM) (65:35). The most efficient chromatography columns for the benzethonium peak were Zorbax SB C8 (150 × 4.6 mm, 3.5 μm) and Diaspher-110-Phenyl (150 × 4.6 mm, 5 μm).
Similar content being viewed by others
References
European Pharmacopoeia, Ed. 6.8, Monograph: Benzethonium Chloride.
United States Pharmacopeia, XXX, Monograph: Benzethonium Chloride.
P. Sadek, Solvents for HPLC [in Russian], BINOM, Moscow (2006), pp. 20 – 21, 26 – 27.
M. D. Mashkovskii, Drugs [in Russian], Novaya Volna, Moscow (2010), pp. 203, 277.
E. B. Nechaeva, A. S. Osipov, and V. L. Bagirova, in: Abstracts of Papers of the XIIIth Russian Scientific Congress “Man and Medicine” [in Russian], Moscow (2006), p. 643.
Author information
Authors and Affiliations
Additional information
Translated from Khimiko-Farmatsevticheskii Zhurnal, Vol. 46, No. 11, pp. 41 – 43, November, 2012.
Rights and permissions
About this article
Cite this article
Osipov, A.S., Orlov, E.N. Application of HPLC to the Analysis of Benzethonium Chloride in Eye Drops. Pharm Chem J 46, 684–686 (2013). https://doi.org/10.1007/s11094-013-0870-0