We demonstrate resolution improvement in an optical coherence tomography (OCT) system with multiple scans of segmented spectral bands and software signal processing. With an effective spectral full-width at half-maximum (FWHM) of 45 nm, centered at 950 nm, we compress the interference fringe FWHM from 31 μm down to 7 μm, which is smaller than the theoretical limit of 8.9 μm if a Gaussian spectral shape is assumed, through the combined effect of dispersion compensation and spectral shaping. With the amplitude adjustments of the OCT images of the segmented spectral bands, side-lobes can be significantly suppressed with a slight increase of the fringe envelope FWHM. With other amplitude adjustments, the envelope FWHM can be reduced to 5.6 μm at the expense of significant side-lobes.
Similar content being viewed by others
References
J.F. Boer Particlede C.E. Saxer J.S. Nelson (2001) Appl.Opt 40 5787 Occurrence Handle2001ApOpt..40.5787D
J.F. Boer Particlede B. Cense B.H. Park M.C. Pierce G.J. Tearney B.E. Bouma (2003) Opt. Lett. 28 2067 Occurrence Handle2003OptL...28.2067D
A.F. Fercher C.K. Hitzenberger M. Sticker R. Zawadzki B. Karamata T. Lasser (2001) Opt. Exp. 9 610 Occurrence Handle2001OExpr...9..610F
A.F. Fercher C.K. Hitzenberger M. Sticker R. Zawadzki B. Karamata T. Lasser (2002) Opt. Comm. 204 67 Occurrence Handle2002OptCo.204...67F
R. Grote H. Fouckhardt (1999) Opt. Exp. 8 328 Occurrence Handle1999OExpr...4..328G
I.J. Hsu C.W. Sun C.W. Lu C.C. Yang C.P. Chiang C.W. Lin (2003) Appl. Opt. 42 227 Occurrence Handle2003ApOpt..42..227H
A.M. Kowalevicz T. Ko I. Hartl J.G. Fujimoto (2002) Opt. Exp. 10 349 Occurrence Handle2002OExpr..10..349K
R. Leitgeb C.K. Hitzenberger A.F. Fercher (2003) Opt. Exp. 11 889 Occurrence Handle2003OExpr..11..889L
D.L. Marks A.L. Oldenburg J.J. Reynolds (2002) Opt. Lett 27 2010 Occurrence Handle2002OptL...27.2010M
D.L. Marks A.L. Oldenburg J.J. Reynolds S.A. Boppart (2003) Appl. Opt 42 204 Occurrence Handle2003ApOpt..42..204M
W.K. Niblack J.O. Schenk B. Liu M.E. Brezinski (2003) Appl. Opt. 42 4115 Occurrence Handle2003ApOpt..42.4115N
E.D.J. Smith A.V. Zvyagin D.D. Sampson (2002) Opt. Lett. 27 1998 Occurrence Handle2002OptL...27.1998S
M. Takeoka D. Fujishima F. Kannari (2001) Opt. Lett. 26 1592 Occurrence Handle2001OptL...26.1592T
R. Tripathi N. Nassif J.S. Nelson B.H. Park J.F. Boer Particlede (2002) Opt. Lett. 27 406 Occurrence Handle2002OptL...27..406T
A. Unterhuber B. Povazay B. Hermann H. Sattmann W. Drexler (2003) Opt. Lett. 28 905 Occurrence Handle2003OptL...28..905U
Y. Wang Y. Zhao JS. Nelson Z. Chen (2003) Opt. Lett. 28 182 Occurrence Handle2003OptL...28..182W
A.M. Weiner A.M. Kan’an (1998) IEEE J. Select. Top. Quan. Electron. 4 317
M. Wojtkowski V.J. Srinivasan T.H. Ko J.G. Fujimoto A. Kowalczyk J.S. Duker (2004) Opt. Exp. 12 2404 Occurrence Handle2004OExpr..12.2404W
A.V. Zvyagin E.D.J. Smith D.D. Sampson (2003) J. Opt. Soc. Am. A 20 333 Occurrence Handle2003OSAJ...20..333Z
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Lu, Cw., Tsai, Mt., Wang, Ym. et al. Resolution Improvement in Optical Coherence Tomography with Segmented Spectrum Management. Opt Quant Electron 37, 1165–1173 (2005). https://doi.org/10.1007/s11082-005-4188-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11082-005-4188-3