Advertisement

Measurement Techniques

, Volume 61, Issue 12, pp 1196–1202 | Cite as

Definition of the Measurement Function of the Modulus of the Vibration Acceleration Vector of Three-Component Vibration Transducers

  • V. Ya. SmirnovEmail author
MECHANICAL MEASUREMENTS
  • 8 Downloads

The measurement function is calculated for the modulus of the vibration acceleration vector of three-component vibration transducers by means of a unicomponent calibrating vibration shaker. Two options for defining the measurement function of the modulus of a vector are analyzed: calibrations against one- and three-component reference vibration transducers. Formulas for finding the measurement function of the modulus of the vibration acceleration vector and calculating errors are given. Experimental results that confirmed the theoretical studies are provided.

Keywords

modulus of the vibration acceleration vector measurement function calibrating vibration shaker transverse motion 

References

  1. 1.
    G. V. Zusman, “Vibration diagnostics and current methods of measuring vibration,” Contr. Diag., No. 3 (189), 2–18 (2014).Google Scholar
  2. 2.
    A. A. Bogoyavlenskii, “Ensuring unity of measurements of vibration parameters during aircraft operation,” Mir Izmer., No. 2 (172), 6–14 (2016).Google Scholar
  3. 3.
    GOST R 8.669-2009, GSI. Vibrometers with Piezoelectric, Induction, and Eddy Current Vibration Transducers. Methods of Verification. Google Scholar
  4. 4.
    MI 2478-98, GSI. Vibration Measuring Apparatus. Methods of Verification. Google Scholar
  5. 5.
    MI 1929-2007, GSI. Vibration Verification Devices. Methods of Verification. Google Scholar
  6. 6.
    A. V. Orlov, V. I. Smirnov, Yu. A. Sholin, et al., Patent 2567987 RF, “A method of verification of three-component vibration transducers,” Izobret. Polezn. Modeli, No. 31 (2015).Google Scholar
  7. 7.
    V. Ya. Smirnov and A. V. Orlov, “Error in determination of the measurement function of the modulus of the vibration acceleration vector,” in: Measurements and Tests in Shipbuilding and Allied Industries, SUDOMETRIKA-2014: Abstr. 5th All-Russ. Sci. Techn. Conf., TsNII Elektropribor, St. Petersburg (2014), pp. 95–98.Google Scholar
  8. 8.
    V. N. Kostyukov, S. N. Kovalev, Vl. V. Yanchich, et al., “Specialized equipment for verification of three-coordinate vibration transducers,” in Current Problems in Piezoelectric Instrumentation: Proc. 2nd Int. Youth Sci. Conf., Sept. 6–10, 2015, Southern Federal University, Rostov-on-Don (2015), Vol. 2, pp. 145–149.Google Scholar
  9. 9.
    MI 2083-90, GSI. Indirect Measurements. Determination of Measurement Results and Estimation of Their Errors. Google Scholar

Copyright information

© Springer Science+Business Media, LLC, part of Springer Nature 2019

Authors and Affiliations

  1. 1.Vibro-PriborSt. PetersburgRussia