Skip to main content
Log in

The applicability of the ITS-90 method to platinum resistance thermometers of different purity

  • Thermal Measurements
  • Published:
Measurement Techniques Aims and scope

Abstract

The static characteristics of platinum resistance thermometers with different platinum purities were measured. It is shown that in the 0–420°C range the ITS-90 method gives an error of less than 0.01°C, while in the 0–230°C range the error is less than 0.006°C.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. N. P. Moiseeva, Izmer. Tekh., No. 5, 39 (2001).

  2. N. P. Moiseeva, Izmer. Tekh., No. 11, 39 (2004).

Download references

Author information

Authors and Affiliations

Authors

Additional information

__________

Translated from Izmeritel’naya Tekhnika, No. 12, pp. 33–36, December, 2006.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kryukov, A.V. The applicability of the ITS-90 method to platinum resistance thermometers of different purity. Meas Tech 49, 1218–1223 (2006). https://doi.org/10.1007/s11018-006-0264-8