Abstract
A method of calibrating a resistive voltage divider for voltage ratios of 1000 V/10 V and 100 V/10 V with an uncertainty at the level of 10−7 is developed. The method enables a calibration to be carried out at the working voltage of the divider and can be used to calibrate a transportable voltage ratio standard when carrying out CCEM-K8 key comparisons.
Similar content being viewed by others
REFERENCES
H. Slinde and K. Lind, IEEE Trans. Instrum. Meas., 52, No.2, 461 (2003).
Y. Sakamoto and H. Fujiki, ibid, p. 465.
The Datron 4902 and 4902S Voltage Dividers, 4901 Calibrator Bridge/Lead Compensator. User Manual, Datron Instrum. Ltd. (1996).
G. Marullo-Reedtz et al., IEEE Trans. Instrum. Meas., 52, No.2, 419 (2003).
Guide to the Expression of Uncertainty in Measurement. ISO, Geneva (1995).