Abstract
This study compares the two analytical methods for uranium concentration determination with high accuracy in uranium pellet: K-edge densitometer (KED) and the isotope dilution with thermal ionisation mass spectrometry measurements (ID-TIMS). Both techniques are compared in terms of time, generated radioactive effluent, simplicity, uncertainty estimation and detection limit. ID-TIMS shows lower detection limit and uncertainties than KED. However, the KED analysis time is shorter and generates less effluent. Both techniques were used for metrological analysis of uranium concentration in nuclear materials. The optimization of sample spike mixture isotope ratio for ID-TIMS to decrease uncertainties is also discussed.
Similar content being viewed by others
References
Godoy MLDP, Godoy JM, Roldão LA, Tauhata L (2009) Determination of total content and isotopic compositions of plutonium and uranium in environmental samples for safeguards purposes by ICP-QMS. J Environ Radioact 100:613–625. https://doi.org/10.1016/j.jenvrad.2009.04.012
Metzger SC, Ticknor BW, Rogers KT, Bostick DA, McBay EH, Hexel CR (2018) Automated separation of uranium and plutonium from environmental swipe samples for multiple collector inductively coupled plasma mass spectrometry. Anal Chem 90:9441–9448. https://doi.org/10.1021/acs.analchem.8b02095
Boulyga S, Konegger-Kappel S, Richter S, Sangély L (2015) Mass spectrometric analysis for nuclear safeguards. J Anal At Spectrom 30:1469–1489. https://doi.org/10.1039/c4ja00491d
Esbelin E, Rivier C, Duhamel G (2018) High concentration measurements of U and Pu with nondestructive and standardless K-edge densitometer device. X Ray Spectrom 47:22–33. https://doi.org/10.1002/xrs.2802
Ottmar H, Eberle H (1991) The hybrid K-edge/K-XRF densitometer: principles-design-performance. KFK Report 4590, Karlsruhe
Quemet A, Maillard C, Ruas A (2015) Determination of zirconium isotope composition and concentration for nuclear sample analysis using thermal ionization mass spectrometry. Int J Mass Spectrom 392:34–40. https://doi.org/10.1016/j.ijms.2015.08.023
Aggarwal SK (2016) Thermal ionisation mass spectrometry (TIMS) in nuclear science and technology—a review. Anal Methods 8:942–957. https://doi.org/10.1039/c5ay02816g
Quemet A, Maloubier M, Dalier V, Ruas A (2014) Development of an analysis method of minor uranium isotope ratio measurements using electron multipliers in thermal ionization mass spectrometry. Int J Mass Spectrom 374:26–32. https://doi.org/10.1016/j.ijms.2014.10.008
Quemet A, Maloubier M, Ruas A (2016) Contribution of the Faraday cup coupled to 1012 ohms current amplifier to uranium 235/238 and 234/238 isotope ratio measurements by thermal ionization mass spectrometry. Int J Mass Spectrom 404:35–39. https://doi.org/10.1016/j.ijms.2016.04.005
Quemet A, Ruas A, Dalier V, Rivier C (2019) Development and comparison of high accuracy thermal ionization methods for uranium isotope ratios determination in nuclear fuel. Int J Mass Spectrom 438:166–174. https://doi.org/10.1016/j.ijms.2019.01.008
Richter S, Kühn H, Aregbe Y, Hedberg M, Horta-Domenech J, Mayer K, Zuleger E, Bürger S, Boulyga S, Köpf A, Poths J, Mathew K (2011) Improvements in routine uranium isotope ratio measurements using the modified total evaporation method for multi-collector thermal ionization mass spectrometry. J Anal At Spectrom 26:550–564. https://doi.org/10.1039/C0JA00173B
Wegener MR, Mathew KJ, Hasozbek A (2013) The direct total evaporation (DTE) method for TIMS analysis. J Radioanal Nucl Chem 296:441–445. https://doi.org/10.1007/s10967-012-2182-3
Mathew KJ, O’Connor G, Hasozbek A, Kraiem M (2013) Total evaporation method for uranium isotope-amount ratio measurements. J Anal At Spectrom 28:866–876
Quemet A, Ruas A, Dalier V, Rivier C (2018) Americium isotope analysis by thermal ionization mass spectrometry using the total evaporation method. Int J Mass Spectrom 431:8–14. https://doi.org/10.1016/j.ijms.2018.05.017
AFNOR (2009) NF T90-210 norm: water quality—protocol for the initial method performance assessment in a laboratory
Désenfant M, Priel M, Rivier C (2005) Evaluation des incertitudes des résultats d’analyse. Ref P105 V1. Les Tech l’Ingénieur 1–17
International Atomic Energy Agency (2010) International Target Values 2010 for Measurement Uncertainties in Safeguarding Nuclear Materials - STR368. Vienna, Austria
Irrgeher J, Vogl J, Santner J, Prohaska T (2015) Chapter 8 measurement strategies. In: Sect. F. Mass Spectrom. Elem. Isot. Anal. The Royal Society of Chemistry, pp 126–151
Acknowledgements
We are grateful to Dr. S. Baghdadi (IRSN/PSN-EXP/SSRD/BTE) for her precious advice on the present paper. We would like to thank all of the IAEA staff (Seibersdorf-Austria) who organized the interlaboratory comparison and all the KED staff for the KED measurements.
Author information
Authors and Affiliations
Corresponding author
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Quemet, A., Ruas, A., Esbelin, E. et al. Development and comparison of two high accuracy methods for uranium concentration in nuclear fuel: ID-TIMS and K-edge densitometry. J Radioanal Nucl Chem 321, 997–1004 (2019). https://doi.org/10.1007/s10967-019-06670-y
Received:
Published:
Issue Date: