Abstract
A method of efficiency calibration for the measurement of 88Kr and 138Xe by HPGe γ-spectrometer is proposed in the present paper. The question for the efficient calibration is, how to achieve homogeneous sources of 88Kr-88Rb and 138Xe-138Cs. The fission product gases were obtained by irradiating a precisely measured amount of U3O8 (90% 235U) filled in a quartz glass ampoule. Source cell was first filled up with stearic acid, and then the fission product gases were charged into it. Xenon and krypton are not adsorbed on stearic acid, therefore, homogeneous sources of 88Kr-88Rb and 138Xe-138Cs can be prepared. The results of the experiment demonstrate that the method is feasible and successful.
Similar content being viewed by others
References
Dangchang Gu, Detecting Method and Technology for Radioactive Nuclides, Science Press, Beijing, 1994, p. 349.
Qi Li, Ph. D. Dissertation, Study on the 4πβ+4πγ Counting Standard Device, Northwest Institute of Nuclear Technology, Xi’an, 2004.
Hongbin Gao, Applied Organic Chemistry Thesaurus, Higher Education Press, Beijing, 1997, p. 767.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Chen, Z.Y., Xie, F., Wang, X.H. et al. Preparation of calibration sources of 88Kr and 138Xe for HPGe detector. J Radioanal Nucl Chem 273, 257–261 (2007). https://doi.org/10.1007/s10967-007-0745-5
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10967-007-0745-5