Laboratory evaluation of performance of pulse oximeters from six different manufacturers during motion artifacts produced by Fluke 2XL SpO2 simulator

  • Original Research
  • Published:
Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

We performed laboratory evaluation of six pulse oximeters from different manufacturers using the Fluke 2XL SpO2 simulator. The pulse oximeter probes were labeled 1 through 6 and tested using the two pre-programed preset functions of the Fluke 2XL SpO2 simulator, level 01 and level 02, for their performance in the presence of motion artifacts. The pulse oximeters were also tested at low perfusion index (PI) settings. The accuracy of the individual probes was ranked as good, mediocre, or poor based on the degree of deviation of the measured SpO2 and pulse rate (PR) by the probes from that generated by the simulator in both the presence and absence of motion artifacts and at lower PI. In preset level 01, probes numbered 1, 2, and 6 performed flawlessly, probes numbered 3, 4, and 5 performed poorly. In preset level 02, all probes performed well in the absence of motion artifacts. When probes were attached in the absence of motion, but readings were recorded after the emergence of motion artifacts, probes 1, 2, and 6 continued to perform well. However, the performance of probes 3, 4, and 5 deteriorated. When probes were attached directly in the presence of motion artifacts, probes 2 and 6 performed well, whereas all other probes performed poorly. Successively lowering the PI degraded performance of probes 3, 4, and 5 at extremely low PI. It is observed that during motion and/or low PI conditions, multiple probes see deterioration in performance.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

Data availability

All the data will be available on request.

Code availability

Not applicable.

References

  1. Tierney LM Jr, Whooley MA, Saint S. Oxygen saturation: a fifth vital sign? West J Med. 1997;166(4):285–6.

    PubMed  PubMed Central  Google Scholar 

  2. Aoyagi T, Miyasaka K. Pulse oximetry: its invention, contribution to medicine, and future tasks. Anesth Analg. 2002;94(1):s1-3.

    Article  Google Scholar 

  3. Hanning CD, Alexander-Williams JM. Pulse oximetry: a practical review. Br Med J. 1985;311:367–70.

    Article  Google Scholar 

  4. Petterson MT, Begnoche VL, Graybeal JM. The effect of motion on pulse oximetry and its clinical significance. Anesth Analg. 2007;105(6):S78-84.

    Article  Google Scholar 

  5. Lawless S. Crying wolf: false alarms in a pediatric intensive care unit. Crit Care Med. 1994;22:981–5.

    Article  CAS  Google Scholar 

  6. Meredith C, Edworthy J. Are there too many alarms in the intensive care unit? An overview of the problems. J Adv Nurs. 1995;21:15–20.

    Article  CAS  Google Scholar 

  7. Wiklund L, Hok B, Stahl K, Jordeby-Jonsson A. Postanesthesia monitoring revisited: frequency of true and false alarms from different monitoring devices. J Clin Anesth. 1994;6:182–8.

    Article  CAS  Google Scholar 

  8. Karen KG, Thomas LH. New-generation pulse oximetery in the care of critically Ill patients. Am J Crit Care. 2005;14:7–24.

    Google Scholar 

  9. Stabile N, Reynolds K. Technological review of pulse oximeter simulators. J Clin Eng. 2002;27(4):287–97.

    Article  Google Scholar 

  10. Cooke JE. When pulse oximeters fail: motion and low perfusion. Anesthesiology. 2000;93(3A):A554.

    Article  Google Scholar 

  11. Shah N, Ragaswamy H, Govindugari K, Estanol L. Performance of three new-generation pulse oximeters during motion and low perfusion in volunteers. J Clin Anesth. 2012;24:385–91.

    Article  Google Scholar 

  12. Aaron L, Feiner JR, Bickler PE, Rhodes L, Bernstein M, Lucero J. Four types of pulse oximeters accurately detect hypoxia during low perfusion and motion. Anesthesiology. 2018;128:520–30.

    Article  Google Scholar 

Download references

Funding

None.

Author information

Authors and Affiliations

Authors

Contributions

GKM: Helped in conducting the research, analyzing the data and writing the manuscript. R and SK: Helped in conducting the research and analyzing the data.

Corresponding author

Correspondence to Munirathinam Ganesh Kumar.

Ethics declarations

Conflict of interest

All Authors declare that they have no competing interest.

Ethical approval

This is an observational study. The Postgraduate Institute of medical education & research, Chandigarh Research Ethics Committee has confirmed that no ethical approval is required.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ganesh Kumar, M., Kaur, S. & Kumar, R. Laboratory evaluation of performance of pulse oximeters from six different manufacturers during motion artifacts produced by Fluke 2XL SpO2 simulator. J Clin Monit Comput 36, 1181–1191 (2022). https://doi.org/10.1007/s10877-021-00747-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10877-021-00747-4

Keywords

Navigation