Optimal level of the reference transducer for central venous pressure and pulmonary artery occlusion pressure monitoring in supine, prone, and sitting position

  • Original Research
  • Published:
Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

To guarantee accurate measurement of central venous pressure (CVP) or pulmonary artery occlusion pressure (PAOP), proper positioning of a reference transducer is a prerequisite. We investigated ideal transducer levels in supine, prone, and sitting position in adults. Chest computed tomography images of 113 patients, taken in supine or prone position were reviewed. For supine position, distances between the back and the uppermost blood level of both atria and their ratios to the largest anteroposterior (AP) diameter of thorax were calculated. For prone position, same distances and ratios were calculated from the anterior chest. For sitting position, distances between the mid-sternoclavicular joint and the most cephalad blood level of both atria and their ratios to the sternal length were calculated. The ratio of the uppermost blood level of right atrium (RA) and left atrium (LA) to the largest AP diameter of thorax was 0.81 ± 0.04 and 0.59 ± 0.03 from the back in supine position. That calculated from the anterior chest in prone position was 0.54 ± 0.03 and 0.46 ± 0.03. The ratio of the most cephalad blood level of RA and LA to the sternal length was 0.70 ± 0.10 and 0.68 ± 0.09 from the mid-sternoclavicular joint in sitting position, which corresponded to the upper border of 4th rib. Optimal CVP transducer levels are at four-fifths of the AP diameter of thorax in supine position, at a half of that in prone position, and at upper border of the 4th sternochondral joint in sitting position. PAOP transducer levels are similar in prone and sitting position, except for supine position which is at three-fifths of the AP diameter of thorax.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Courtois M, Fattal PG, Kovacs SJ Jr, Tiefenbrunn AJ, Ludbrook PA. Anatomically and physiologically based reference level for measurement of intracardiac pressures. Circulation. 1995;92:1994–2000.

    Article  CAS  PubMed  Google Scholar 

  2. Mueller HS, Chatterjee K, Davis KB, Fifer MA, Franklin C, Greenberg MA, et al. ACC expert consensus document. Present use of bedside right heart catheterization in patients with cardiac disease. J Am Coll Cardiol. 1998;32:840–64.

    Article  CAS  PubMed  Google Scholar 

  3. Pittman JA, Ping JS, Mark JB. Arterial and central venous pressure monitoring. Int Anesthesiol Clin. 2004;42:13–30.

    Article  PubMed  Google Scholar 

  4. Seo JH, Jung CW, Bahk JH. Uppermost blood levels of the right and left atria in the supine position: implication for measuring central venous pressure and pulmonary artery wedge pressure. Anesthesiology. 2007;107:260–3.

    Article  PubMed  Google Scholar 

  5. Kee LL, Simonson JS, Stotts NA, Skov P, Schiller NB. Echocardiographic determination of valid zero reference levels in supine and lateral positions. Am J Crit Care. 1993;2:72–80.

    CAS  PubMed  Google Scholar 

  6. Rauen CA, Makic MB, Bridges E. Evidence-based practice habits: transforming research into bedside practice. Crit Care Nurse. 2009;29:46–59 (quiz 60–61).

    Article  PubMed  Google Scholar 

  7. Winsor T, Burch GE. Phlebostatic Axis and Phlebostatic Level, Reference Levels for Venous Pressure Measurements in Man. P Soc Exp Biol Med. 1945;58:165–9.

    Article  Google Scholar 

  8. Figg KK, Nemergut EC. Error in central venous pressure measurement. Anesth Analg. 2009;108:1209–11.

    Article  PubMed  Google Scholar 

  9. Sondergaard S, Parkin G, Aneman A. Central venous pressure: we need to bring clinical use into physiological context. Acta Anaesthesiol Scand. 2015;59:552–60.

    Article  CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jin-Tae Kim.

Ethics declarations

Conflict of interest

No conflicts of interest declared by the authors.

Ethical statement

This study was approved by the Seoul National University Hospital Institutional Review Board (02, February, 2015/No. 1501-091-642). All procedures performed in this study were in accordance with the ethical standards of the Seoul National University Hospital Institutional Review Board with the 1964 Helsinki declaration. This study also complied with the current laws of Republic of Korea.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Song, IK., Park, HS., Lee, JH. et al. Optimal level of the reference transducer for central venous pressure and pulmonary artery occlusion pressure monitoring in supine, prone, and sitting position. J Clin Monit Comput 31, 381–386 (2017). https://doi.org/10.1007/s10877-016-9864-2

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10877-016-9864-2

Keywords

Navigation