Abstract
Objective
Processed EEG monitoring during cardiopulmonary bypass (CPB) may help determine loss of consciousness and depth of anesthesia. This study compared the SNAP™ II and BIS Vista monitors in patients undergoing isoflurane anesthesia with normothermic CPB.
Methods
40 subjects undergoing CPB with isoflurane anesthesia were enrolled. Subjects were premedicated with 1–2 mg midazolam approximately 5 min prior to acquisition of baseline index values and anesthesia induced with midazolam and fentanyl. Anesthesia was maintained with isoflurane, midazolam, and fentanyl and a cis-atracurium infusion. SNAP™ II (version 1.2.9 algorithm 1.88) and BIS Vista (application version 3.00 platform version 2.03) indices were recorded at baseline, pre-induction, post-intubation, incision, start of CPB, every 15 min during CPB, end of CPB, and end of case. Agreement between methods was determined using Pearson correlation and the Bland–Altman method with repeated observa- tions.
Results
Twenty-four male and 12 female subjects completed the analysis. The correlation between SNAP™ II and BIS Vista index values was 0.61 (P < 0.005). A linear relationship between the difference in the indices and the average index values was observed following the induction of anesthesia. In awake subjects, the bias between the SNAP™ II and BIS Vista was 5 (95% CI 3–7). The limits of agreement were 23 (95% CI 19–26) and −13 (95% CI −9–−16). During anesthesia, the mean difference on a log scale was 0.11 (95% CI 0.09–0.12). The limits of agreement were 0.43 (95% CI 0.40–0.45) and −0.21 (95% CI −0.18–−0.24). The antilog of the mean difference demonstrated that the SNAP™ II value was 28% (95% CI 24–33%) higher than the BIS Vista value following induction of anesthesia.
Conclusions
The SNAP™ II monitor demonstrates a consistently positive bias during cardiopulmonary bypass under isoflurane anesthesia compared with the BIS Vista.
Similar content being viewed by others
References
Wong CA, Fragen RJ, Fitzgerald P, McCarthy RJ. A comparison of the SNAP™ II and BIS XP indices during sevoflurane and nitrous oxide anesthesia at 1 and 1.5 MAC and at awakening. Br J Anaesth. 2006;97:181–186.
Wong CA, Fragen R, Fitzgerald P, McCarthy RJ. The association between propofol-induced loss of consciousness and the SNAP index. Anesth Analg. 2005;100:141–148.
Messner M, Beese U, Romstöck J, Dinkel M, Tsaichowsky K. The bispectral index declines during neuromuscular block in fully awake persons. Anesth Analg. 2003;97:488–491.
Dahaba A, Mattweber M, Fuchs A, et al. The effect of different stages of neuromuscular block on the bispectral index and the bispectral index-XP under remifentanil/propofol anesthesia. Anesth Analg. 2004;99:781–787.
Greif R, Greenwald S, Schweitzer E, et al. Muscle relaxation does not alter hypnotic level during propofol anesthesia. Anesth Analg. 2002;94:604–608.
Miller A, Sleigh JW, Barnard J, Steyn-Ross DA. Does bispectral analysis of the electroencephalogram add anything but complexity? Br J Anaesth. 2004;92:8–13.
Hrelec C, Puente E, Bergese S, Dzwonczyk R. SNAP II versus BIS vista monitor comparison during general anesthesia. J Clin Monit Comput. 2010;24:283–288.
Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17:571–582.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;8476:307–310.
Ruiz-Gimeno P, Soro M, Pérez-Solaz A, et al. Comparison of the EEG-based SNAP index and the bispectral index during sevoflurane-nitrous oxide anesthesia. J Clin Monit Comput. 2005;19:383–389.
Schmidt GN, Bischoff P, Standl T, et al. SNAP index and bispectral index during different states of propofol/remifentanil anesthesia. Anaesth. 2005;60:228–234.
Salova EM, Lubnin AIu, Israelian LA, Shimanskiĭ VN. Bilateral bispectral index monitoring during surgery for posterior cranial fossa tumors. Anesteziol Reanimatol. 2011;2:52–8.
Kodaka M, Nishikawa Y, Suzuki T, et al. Does bilateral bispectral index monitoring (BIS) detect the discrepancy of cerebral reperfusion during carotid endarterectomy? J Clin Anesth. 2009;21:431–434.
Heller H, Hatami R, Mullin P, et al. Bilateral bispectral index monitoring during suppression of unilateral hemispheric function. Anesth Analg. 2005;101:235–241.
Financial support for this study was provided by the Department of Anesthesiology, Northwestern University. Stryker Inc. supplied the SNAP™ II devices and disposables used in this study. This study was presented in part at the American Society of Anesthesiologist Annual Meeting Orlando Florida, October 20, 2008.
Author information
Authors and Affiliations
Corresponding author
Additional information
Patel KM, Sherwani SS, Fitzgerald PC, McCarthy RJ. Comparison of SNAP™II and BIS Vista indices during normothermic cardiopulmonary bypass under isoflurane anesthesia.
Rights and permissions
About this article
Cite this article
Patel, K.M., Sherwani, S.S., Fitzgerald, P.C. et al. Comparison of SNAP™ II and BIS Vista indices during normothermic cardiopulmonary bypass under isoflurane anesthesia. J Clin Monit Comput 25, 365–370 (2011). https://doi.org/10.1007/s10877-011-9313-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10877-011-9313-1