Tight reservoir bag: the bag itself may be the culprit

  • Published:
Journal of Clinical Monitoring and Computing Aims and scope Submit manuscript

Abstract

Numerous possibilities exist which may cause obstruction to ventilation under anesthesia resulting in a tight reservoir bag with low compliance. We report an interesting case where a reservoir bag twisted around its own neck and resulted in a tight bag situation. The neck portion of the reservoir bag would be hidden from the view of anesthesiologists in head and neck surgery and hence it is easier to miss early recognition of the twist. We caution all anesthesiologists using the disposable modified Jackson-Rees breathing system to be aware of such an eventuality. We also urge the manufacturer to consider strengthening the neck of the reservoir bag by improving the quality of the material used for its construction.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Schlager A, Furtner B, Mitterschiffthaler G. Acute obstruction during manual ventilation caused by an end-tidal rubber cap in the reservoir bag. Anesth Analg. 1999;89:804–5.

    Article  CAS  PubMed  Google Scholar 

  2. Foreman MJ, Moyes DG. Anaesthetic breathing circuit obstruction due to blockage of tracheal tube connector by a foreign body: two cases. Anaesth Intensive Care. 1999;27:73–5.

    CAS  PubMed  Google Scholar 

  3. Gallacher BP, Kelly M, Mora RR. Failure to ventilate due to glass ampule fragment occlusion of the breathing circuit. Anesthesiology. 1997;87:180.

    Article  CAS  PubMed  Google Scholar 

  4. Krensavage TJ, Richards E. Sudden development of anesthesia circuit obstruction by an end-tidal CO2 cap in the gas sampling elbow. Anesth Analg. 1995;81:207.

    Article  CAS  PubMed  Google Scholar 

  5. Mehta M, Garg R, Rath GP, Prabhakar H. Beware kinking of disposable plastic circuit tubing during anaesthesia. Anaesth Intensive Care. 2008;36:624.

    CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Goneppanavar Umesh MD.

Additional information

Umesh G, Jasvinder K. Tight reservoir bag: the bag itself may be the culprit.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Umesh, G., Jasvinder, K. Tight reservoir bag: the bag itself may be the culprit. J Clin Monit Comput 24, 189–190 (2010). https://doi.org/10.1007/s10877-010-9232-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10877-010-9232-6

Key Words

Navigation