Abstract
Objective
(1) To investigate if there exist any discrepancies between the values of vital signs charted by nurses and those recorded by bedside monitors for a group of patients admitted for neurocritical care. (2) To investigate possible interpretations of discrepancies by exploring information in the alarm messages and the raw waveform data from monitors.
Methods
Each charted vital sign value was paired with a corresponding value from data collected by an archival program of bedside monitors such that the automatically archived data preceded the charted data and had minimal time lag to the charted value. Next, the absolute differences between the paired values were taken as the discrepancy between charted and automatically-collected data. Archived alarm messages were searched for technical alarms of sensor/lead failure types. Additionally, 7-min waveform data around the place of large discrepancy were analyzed using signal abnormality indices (SAI) for quantifying the quality of recorded signals.
Results
About 31,145 pairs of systolic blood pressure (BP-S) and 67,097 pairs of SpO2 were investigated. Seven and a half percent of systolic blood pressure pairs had a discrepancy greater than 20 mmHg and less than one percent of the SpO2 pairs had a discrepancy greater than 10. We could not find any technical alarms from the monitors that could explain the large difference. However, SAI calculated for the waveforms associated with this group of cases was significantly larger than the SAI values calculated for the control waveform data of the same patients with small discrepancies.
Conclusion
Charted vital signs reflect in large the raw data as reported by bedside monitors. Poor signal quality could partially explain the existence of cases of large discrepancies.
Article PDF
Similar content being viewed by others
![](https://media.springernature.com/w215h120/springer-static/image/art%3A10.1186%2Fs12912-016-0149-3/MediaObjects/12912_2016_149_Fig1_HTML.gif)
Avoid common mistakes on your manuscript.
References
Ward NS. The accuracy of clinical information systems. J Crit Care. 2004;19:221–5.
Vawdrey DK, Gardner RM, Evans RS, Orme JF Jr, Clemmer TP, Greenway L, et al. Assessing data quality in manual entry of ventilator settings. J Am Med Inform Assoc. 2007;14:295–303.
Friesdorf W, Konichezky S, Gross-Alltag F, Fattroth A, Schwilk B. Data quality of bedside monitoring in an intensive care unit. Int J Clin Monit Comput. 1994;11:123–8.
Gregorini P. Comparison of four methods of automated recording of physiologic data at one minute intervals. J Clin Monit. 1996;12:299–303.
Levy MM. Computers in the intensive care unit. J Crit Care. 2004;19:199–200.
Gardner RM, Hawley WL, East TD, Oniki TA, Young HF. Real time data acquisition: recommendations for the Medical Information Bus (MIB). Int J Clin Monit Comput. 1991;8:251–8.
Pringle M, Ward P, Chilvers C. Assessment of the completeness and accuracy of computer medical records in four practices committed to recording data on computer. Br J Gen Pract. 1995;45:537–41.
Marasovic C, Kenney C, Elliott D, Sindhusake D. A comparison of nursing activities associated with manual and automated documentation in an Australian intensive care unit. Comput Nurs. 1997;15:205–11.
Reich DL, Wood RK Jr, Mattar R, Krol M, Adams DC, Hossain S, et al. Arterial blood pressure and heart rate discrepancies between handwritten and computerized anesthesia records. Anesth Analg. 2000;91:612–6.
Moret-Bonillo V, Mosqueira-Rey E, Alonso-Betanzos A. Information analysis and validation of intelligent monitoring systems in intensive care units. IEEE Trans Inf Technol Biomed. 1997;1:87–99.
Mora FA, Passariello G, Carrault G, Le Pichon JP. Intelligent patient monitoring and management systems: a review. IEEE Eng Med Biol Mag. 1993;12:23–33.
Sun JX, Reisner AT, Mark RG. A signal abnormality index for arterial blood pressure waveforms. Comput Cardiol. 2006;33:13–6.
Hu X, Xu P, Lee DJ, Vespa P, Baldwin K, Bergsneider M. An algorithm for extracting intracranial pressure latency relative to electrocardiogram R wave. Physiol Meas. 2008;29:459–71.
Asgari S, Xu P, Bergsneider M, Hu X. A subspace decomposition approach toward recognizing valid pulsatile signals. Physiol Meas 2009 (submitted).
Hu X, Xu P, Scalzo F, Vespa P, Bergsneider M. Morphological clustering and analysis of continuous intracranial pressure. IEEE Trans Biomed Eng. 2009;56:696–705.
Asgari S, Bergsneider M, Hu X. Robust approach towards recognizing valid arterial blood pressure pulses. IEEE Trans Inf Technol Biomed 2009 (submitted).
Block FE Jr. Normal fluctuation of physiologic cardiovascular variables during anesthesia and the phenomenon of “smoothing”. J Clin Monit. 1991;7:141–5.
East TD, Young WH, Gardner RM. Digital electronic communication between ICU ventilators and computers and printers. Respir Care. 1992;37:1113–23.
Donati A, Gabbanelli V, Pantanetti S, Carletti P, Principi T, Marini B, et al. The impact of a clinical information system in an intensive care unit. J Clin Monit Comput. 2008;22:31–6.
Author information
Authors and Affiliations
Corresponding author
Additional information
Sapo M, Wu S, Asgari S, McNair N, Buxey F, Martin N, Hu X. A comparison of vital signs charted by nurses with automated acquired values using waveform quality indices.
Rights and permissions
About this article
Cite this article
Sapo, M., Wu, S., Asgari, S. et al. A comparison of vital signs charted by nurses with automated acquired values using waveform quality indices. J Clin Monit Comput 23, 263–271 (2009). https://doi.org/10.1007/s10877-009-9192-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10877-009-9192-x