Abstract
Large diameter metal-on-metal bearings (MOM) are becoming increasingly popular, addressing the needs of young and more active patients. Clinical data has shown excellent short-to-mid-term results, though incidences of transient squeaking have been noted between implantation and up to 2 years post-operative. Geometric design features, such as clearance, have been significant in influencing the performance of the bearings. Sets of MOM bearings with different clearances were investigated in this study using a hip friction simulator to examine the influence of clearance on friction, lubrication and squeaking. The friction factor was found to be highest in the largest clearance bearings under all test conditions. The incidence of squeaking was also highest in the large clearance bearings, with all bearings in this group squeaking throughout the study. A very low incidence of squeaking was observed in the other two clearance groups. The measured lubricating film was found to be lowest in the large clearance bearings. This study suggests that increasing the bearing clearance results in reduced lubricant film thickness, increased friction and an increased incidence of squeaking.
Similar content being viewed by others
References
J. CHARNLEY, Hip (1982) 42
S. KOBAYASHI, N. S. EFTEKHAR, K. TERAYAMA and R. P. JOSHI, Clin. Orthop. Relat. Res. 339 (1997) 140
H. WAGNER, Clin. Orthop. Relat. Res. 134 (1978) 102
R.S. BELL, et al., J. Bone Joint Surg. Am. 67(8) (1985) 1165
S. TANAKA, Clin. Orthop. Relat. Res. 134 (1978) 75
D. MCMINN, R. TREACY, K. LIN and P. PYNSENT, Clin. Orthop. Relat. Res. (329 Suppl) (1996) S89
R. B. C. TREACY, C. W. MCBRYDE and P. B. PYNSENT, J. Bone Joint Surg. Br. 87B(2) (2005) 167
J. DANIEL, P. PYNSENT and D. MCMINN, J. Bone Joint Surg. Br. 86B(2) (2004) 177
Z. M. JIN, in Proceedings of the Institution of Mechanical Engineers. Part H: J. Eng. Med. 216(1) (2002) 85
S. C. SCHOLES and A. UNSWORTH, in Proceedings of the Institution of Mechanical Engineers, Part H: J. Eng. Med. V220(2) (2006) 183
F. W. CHAN, et al., Clin. Orthop. Relat. Res. 369 (1999) 10
A. EBIED, S. F. JOURNEAUX and J.A. POPE, in International Conference for Engineers and Surgeons—Joined at the Hip (IMechE, London, 2002)
D. L. BACK, et al., J. Bone Joint Surg. Br. 87B(3) (2005)
C. Brockett, et al., J. Biomed. Mater. Res. B: Appl. Biomater. 81B(2) (2007) 508
R. DWYER-JOYCE, P. HARPER and B. DRINKWATER, Tribol. Lett. (2004) 337
C. B. RIEKER, et al., Orthop. Clin. North Am. 36(2) (2005) 135
D. DOWSON, C. M. MCNIE and A. A. GOLDSMITH, J. Mech. Eng. Sci. 214(1) (2000) 175
M. T. MAI, et al., J. Bone Joint Surg. Am. 78(4) (1996) 505
S. MA, J. KABO and H. AMSTUTZ, J. Bone Joint Surg. Am. 65(3) (1983) 366
A. KUSABA, et al., J. Long-Term Eff. Med. Implants 12(1) (2002) 53
W. L. WALTER, G. C. O’TOOLE, W. K. WALTER, A. ELLIS and B. A. ZICAT, Arthroplasty 22(4) (2007) 496
A. EBIED and S. JOURNEAUX, Curr. Orthop. 16(6) (2002) 420
P. HARPER, R. S. DWYER-JOYCE, U. SJODIN and U. OLOFSSON, in Proceedings of the 31st Leeds-Lyon Symposium (Elsevier, 2005)
R. DWYER-JOYCE, B. W. DRINKWATER and C. DONOHOE, in Proceedings of the Math. Phys. Eng. Sci. (2003) 957
Acknowledgement
The implants tested during this study were provided by DePuy International, Leeds, UK.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Brockett, C.L., Harper, P., Williams, S. et al. The influence of clearance on friction, lubrication and squeaking in large diameter metal-on-metal hip replacements. J Mater Sci: Mater Med 19, 1575–1579 (2008). https://doi.org/10.1007/s10856-007-3298-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10856-007-3298-9