Abstract
Purpose
Diagnostic electrophysiology catheters are single-use and their expense can act as a barrier to their use. A decapolar catheter with a re-sterilisable handle and inner core was developed, along with a disposable outer sheath. This initial experience evaluated the performance of the Khelix partially recyclable steerable decapolar electrophysiology catheter.
Methods
Sequential patients requiring a decapolar steerable catheter were included in the study. Parameters such as time to positioning from right femoral vein, fluoroscopy time and dose, pacing and sensing thresholds, and the requirement for reposition and/or support sheath were recorded. A retrospective analysis from cases using a standard disposable decapolar catheter was used to compare CS cannulation success, support sheath requirement and cost.
Results
11 re-sterilisable catheter handles were used in 64 patients, each time with a disposable sheath, at two New Zealand centres from December 2017 to June 2018. Pre-defined safety measures were met in all patients, with no catheter-related adverse events. The catheter was successfully positioned in the coronary sinus for 59 cases (92%). Based on initial manufacturer recommendations, the inner recyclable portion of the catheter was able to be re-used successfully up to 12 times with re-sterilisation between cases.
Conclusions
This initial study of an innovative, partially reusable, steerable decapolar electrophysiology catheter demonstrates that it may be used safely and successfully. Use of such a catheter may decrease the cost required for electrophysiology studies.
Similar content being viewed by others
References
Jackman WM, Friday KJ, Yeung-Lai-Wah JA, Fitzgerald DM, Beck B, Bowman AJ, et al. New catheter technique for recording left free-wall accessory atrioventricular pathway activation. Identification of pathway fiber orientation. Circulation. 1988;78(3):598–611.
Joseph JP, Rajappan K. Radiofrequency ablation of cardiac arrhythmias: past, present and future. QJM-INT J MED. 2012;105(4):303–14.
Daoud EG, Niebauer M, Bakr O, Jentzer J, Man KC, Williamson BD, et al. Placement of electrode catheters into the coronary sinus during electrophysiology procedures using a femoral vein approach. Am J Cardiol. 1994;74:194–5. https://doi.org/10.1016/0002-9149(94)90102-3.
Haines DE, Beheiry S, Akar JG, Baker JL, Beinborn D, Beshai JF, et al. Heart Rhythm Society expert Consensus statement on electrophysiology laboratory standards: process, protocols, equipment, personnel, and safety. Heart Rhythm. 2014;11(8):e9–e51.
Mont L, Brugada J. Electrophysiology: it is time to simplify! Europace. 2009;11(8):985–6.
Liew R, Baker V, Richmond L, Rajappan K, Gupta D, Finlay M, et al. A randomized-controlled trial comparing conventional with minimal catheter approaches for the mapping and ablation of regular supraventricular tachycardias. Europace. 2009;11:1057–64.
Avitall B, Khan M, Krum D, Jazayeri M, Hare J. Repeated use of ablation catheters: a prospective study. J Am Coll Cardiol. 1993;22:1367–72.
Shandling AH, Rieders D, Edwards M. Use of a unique long pre-curved sheath to facilitate femoral placement of coronary sinus catheters. IPEJ. 2009;9(5):247–50.
Skala T, Taborsky M. Electromechanical mapping in electrophysiology and beyond. Czech Society of Cardiology. 2015;57(6):e470–82.
Kuck KH, Schluter M. Single-catheter approach to radiofrequency current ablation of left-sided accessory pathways in patients with Wolff–Parkinson–White syndrome. Circulation. 1991;84:2366–75.
Kleman JM, Pinski SL, Sterba R, et al. A technique for rapid transfemoral catheterization of the coronary sinus with multielectrode catheters. Pacing Clin Electrophysiol. 1994;17(11 Pt1):1737–40.
Kuniyoshi RR, Sternick EB, Nadalin E, Hachul DT. Reprocessing of medical products in electrophysiology. Arq Bras Cardiol. 2017;108(2):169–72.
Bracklo M, Marczak U. Reprocessing cardiology devices - reviewing 15 years of experience in Germany. ECR. 2012;8(2):82–3.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
This observational study was approved by the local Institutional Review Board (WaikatoDHB Quality and Patient Safety).
Conflict of interest
A. Wilson has received funding from CathRx to attend a conference.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards. The audit was approved by the local research governance body and participants were included in the study after obtaining standard informed consent for the clinical procedure.
Khelix catheter inners and sheaths were provided by CathRx Ltd. None of the authors have financial relationships with CathRx.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Wilson, A.L., Timmins, K.A., Allen, R.F. et al. Initial experience with a novel re-sterilisable decapolar electrophysiology catheter. J Interv Card Electrophysiol 58, 177–183 (2020). https://doi.org/10.1007/s10840-019-00583-2
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-019-00583-2