Cryoballoon ablation for pulmonary vein isolation in patients with atrial fibrillation: preliminary results using novel short-tip cryoballoon

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Purpose

In this study, we aimed to report our preliminary experience regarding the impact of the novel short-tip (ST) cryoballoon (CB) on procedural efficacy and signal quality during pulmonary vein (PV) isolation for both paroxysmal and persistent atrial fibrillation.

Methods

Between March, 2015, and August, 2015, we enrolled a total of 64 patients (47 patients male, 73 %) with a mean age of 60 ± 11 years. In the study population, 31 patients (48 %) underwent PVI using Advance (ADV) CB and 33 (52 %) patients with ST CB. In all patients, a 28-mm balloon was used.

Results

Acute procedural success rates were 100 % for the entire study population. A statistically insignificant increase in the percentage of PV signal recordings was observed with ST CB in all PVs compared to ADV CB [88 vs. 81 % for left superior PV (LSPV), 82 vs. 78 % for left inferior PV (LIPV), 85 vs. 84 % for right superior PV (RSPV), 82 vs. 71 % for right inferior PV (RIPV), p < 0.05]. Additionally, the difference in minimum temperature reached during the procedure per PV was not statistically significant between ST CB and ADV CB except the LIPV (LSPV −44.2 ± 5.9 vs. −45.6 ± 5.3 °C, p = 0.970; LIPV −38.7 ± 4.6 °C vs −44.6 ± 6.8 °C, p < 0.001; RSPV −45.6 ± 7.4 °C vs.−47.2 ± 6.1 °C, p = 0.168; RIPV −41.4 ± 5.1 °C vs.−43.7 ± 6.3 °C, p = 0.360). Time to isolation for each PV was similar between ST and ADV CB (p > 0.05).

Conclusion

Our preliminary findings indicated similar acute procedural success for the novel ST CB compared with second generation ADV CB. The increase in the ratio of PV signal recordings obtained during the PVI using the novel ST CB was not significant.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Metzner A, Reissmann B, Rausch P, Mathew S, Wohlmuth P, Tilz R, et al. One-year clinical outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon. Circ Arrhythm Electrophysiol. 2014;7:288–92.

    Article  PubMed  Google Scholar 

  2. Aytemir K, Gurses KM, Yalcin MU, Kocyigit D, Dural M, Evranos B, et al. Safety and efficacy outcomes in patients undergoing pulmonary vein isolation with second-generation cryoballoondagger. Europace. 2015;17:379–87.

    Article  PubMed  Google Scholar 

  3. Ciconte G, Ottaviano L, de Asmundis C, Baltogiannis G, Conte G, Sieira J, et al. Pulmonary vein isolation as index procedure for persistent atrial fibrillation: one-year clinical outcome after ablation using the second-generation cryoballoon. Heart Rhythm. 2015;12:60–6.

    Article  PubMed  Google Scholar 

  4. Koektuerk B, Yorgun H, Hengeoz O, Turan CH, Dahmen A, Yang A, et al. Cryoballoon ablation for pulmonary vein isolation in patients with persistent atrial fibrillation: one-year outcome using second generation cryoballoon. Circ Arrhythm Electrophysiol. 2015;8:1073–9.

    Article  PubMed  Google Scholar 

  5. Furnkranz A, Bordignon S, Dugo D, Perotta L, Gunawardene M, Schulte-Hahn B, et al. Improved 1-year clinical success rate of pulmonary vein isolation with the second-generation cryoballoon in patients with paroxysmal atrial fibrillation. J Cardiovasc Electrophysiol. 2014;25:840–4.

    Article  PubMed  Google Scholar 

  6. Chun KR, Furnkranz A, Metzner A, Schmidt B, Tilz R, Zerm T, et al. Cryoballoon pulmonary vein isolation with real-time recordings from the pulmonary veins. J Cardiovasc Electrophysiol. 2009;20:1203–10.

    Article  PubMed  Google Scholar 

  7. Kuhne M, Knecht S, Altmann D, Ammann P, Schaer B, Osswald S, et al. Validation of a novel spiral mapping catheter for real-time recordings from the pulmonary veins during cryoballoon ablation of atrial fibrillation. Heart Rhythm. 2013;10:241–6.

    Article  PubMed  Google Scholar 

  8. Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation—developed with the special contribution of the European Heart Rhythm Association. Europace. 2012;14:1385–413.

    Article  PubMed  Google Scholar 

  9. Haissaguerre M, Jais P, Shah DC, Takahashi A, Hocini M, Quiniou G, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998;339:659–66.

    Article  CAS  PubMed  Google Scholar 

  10. Chen SA, Hsieh MH, Tai CT, Tsai CF, Prakash VS, Yu WC, et al. Initiation of atrial fibrillation by ectopic beats originating from the pulmonary veins: electrophysiological characteristics, pharmacological responses, and effects of radiofrequency ablation. Circulation. 1999;100:1879–86.

    Article  CAS  PubMed  Google Scholar 

  11. Kuck KH, Furnkranz A. Cryoballoon ablation of atrial fibrillation. J Cardiovasc Electrophysiol. 2010;21:1427–31.

    Article  PubMed  Google Scholar 

  12. Verma A, Jiang CY, Betts TR, Chen J, Deisenhofer I, Mantovan R, et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372:1812–22.

    Article  PubMed  Google Scholar 

  13. Ho SY, Cabrera JA, Tran VH, Farre J, Anderson RH, Sanchez-Quintana D. Architecture of the pulmonary veins: relevance to radiofrequency ablation. Heart. 2001;86:265–70.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  14. Saito T, Waki K, Becker AE. Left atrial myocardial extension onto pulmonary veins in humans: anatomic observations relevant for atrial arrhythmias. J Cardiovasc Electrophysiol. 2000;11:888–94.

    Article  CAS  PubMed  Google Scholar 

  15. Su W, Kowal R, Kowalski M, Metzner A, Svinarich JT, Wheelan K, et al. Best practice guide for cryoballoon ablation in atrial fibrillation: the compilation experience of more than 3000 procedures. Heart Rhythm. 2015;12:1658–66.

    Article  PubMed  Google Scholar 

  16. Furnkranz A, Koster I, Chun KR, Metzner A, Mathew S, Konstantinidou M, et al. Cryoballoon temperature predicts acute pulmonary vein isolation. Heart Rhythm. 2011;8:821–5.

    Article  PubMed  Google Scholar 

  17. Ciconte G, de Asmundis C, Sieira J, Conte G, Di Giovanni G, Mugnai G, et al. Single 3-minute freeze for second-generation cryoballoon ablation: one-year follow-up after pulmonary vein isolation. Heart Rhythm. 2015;12:673–80.

    Article  PubMed  Google Scholar 

  18. Metzner A, Rausch P, Lemes C, Reissmann B, Bardyszewski A, Tilz R, et al. The incidence of phrenic nerve injury during pulmonary vein isolation using the second-generation 28 mm cryoballoon. J Cardiovasc Electrophysiol. 2014;25:466–70.

    Article  PubMed  Google Scholar 

  19. Straube F, Dorwarth U, Schmidt M, Wankerl M, Ebersberger U, Hoffmann E. Comparison of the first and second cryoballoon: high-volume single-center safety and efficacy analysis. Circ Arrhythm Electrophysiol. 2014;7:293–9.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Buelent Koektuerk.

Ethics declarations

The study was in compliance with the principles outlined in the Declaration of Helsinki and approved by the Institutional Ethics Committee.

Conflict of Interest

Dr. Buelent Koektuerk is the consultant of Medtronic Company. Other authors did not have any conflict of interest.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Koektuerk, B., Yorgun, H., Koektuerk, O. et al. Cryoballoon ablation for pulmonary vein isolation in patients with atrial fibrillation: preliminary results using novel short-tip cryoballoon. J Interv Card Electrophysiol 47, 91–98 (2016). https://doi.org/10.1007/s10840-016-0134-5

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-016-0134-5

Key words

Navigation