Abstract

Background or purpose

Real-life comparative data of implantable cardioverter-defibrillator (ICD) longevity are needed to identify the best possible device longevity for optimal patient comfort and to minimize risk for multiple replacements.

Methods

We retrospectively studied ICD longevity in a large single-center population. At our institution, 1272 consecutive patients were implanted with 1665 ICD or cardiac resynchronization therapy with defibrillator (CRT-D) devices between 1998 and 2010. Device data and follow-up were retrieved from the continuous ICD clinic documentation, and generator replacements were verified from the hospital charts. Kaplan-Meier event probabilities for the time to generator replacement were calculated according to device type, pacing percentage (sum of atrial, right and left ventricular pacing, if applicable) and right ventricular pacing threshold, incidence of ICD shocks, and manufacturer. Multivariate Cox proportional hazards regression was performed in addition.

Results

A total of 470 devices were replaced for elective replacement indicator. These occurred after 5.4 years (95 % confidence interval, 5.1 to 5.6 years) in 175 of 625 implanted Boston Scientific or Guidant devices, after 5.7 years (5.5 to 5.9 years) in 266 of 883 Medtronic devices, and 5.2 years (5.0 to 5.4 years) in 29 of 157 Biotronik devices, respectively. Differences between manufacturers remained significant upon multivariate analysis (Medtronic vs. Boston Scientific: P = 0.01; both vs. Biotronik: P < 0.01) as did differences of device type (single- and dual-chamber ICD vs. CRT-D, P < 0.01) and pacing percentage (P < 0.01). Pacing output and ICD shocks did not influence battery longevity.

Conclusion

ICD longevity differed significantly between manufacturers, independent of device type or pacing parameters.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3

Similar content being viewed by others

Abbreviations

ATP:

Antitachycardia pacing

CI:

Confidence interval

CRT-D:

Cardiac resynchronization therapy with defibrillator

DDD:

Dual-chamber device

ERI:

Elective replacement indicator

HR:

Hazard ratio

ICD:

Implantable cardioverter-defibrillator

RV:

Right ventricular

SCD:

Sudden cardiac death

VVI:

Single-chamber device

VF:

Ventricular fibrillation

VT:

Ventricular tachycardia

References

  1. The Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. (1997). A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. New England Journal of Medicine, 337(22), 1576–1583.

    Article  Google Scholar 

  2. Moss, A. J., Hall, W. J., Cannom, D. S., et al. (1996). Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. New England Journal of Medicine, 335(26), 1933–1940.

    Article  CAS  PubMed  Google Scholar 

  3. Buxton, A. E., Lee, K. L., Fisher, J. D., et al. (1999). A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. New England Journal of Medicine, 341(25), 1882–1890.

    Article  CAS  PubMed  Google Scholar 

  4. Moss, A. J., Zareba, W., Hall, W. J., et al. (2002). Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. New England Journal of Medicine, 346(12), 877–883.

    Article  PubMed  Google Scholar 

  5. Bardy, G. H., Lee, K. L., Mark, D. B., et al. (2005). Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. New England Journal of Medicine, 352(3), 225–237.

    Article  CAS  PubMed  Google Scholar 

  6. Tracy, C. M., Epstein, A. E., Darbar, D., et al. (2012). 2012 ACCF/AHA/HRS focused update of the 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Journal of the American College of Cardiology, 60(14), 1297–1313.

    Article  PubMed  Google Scholar 

  7. Hauser, R. G. (2005). The growing mismatch between patient longevity and the service life of implantable cardioverter-defibrillators. Journal of the American College of Cardiology, 45(12), 2022–2025.

    Article  PubMed  Google Scholar 

  8. Cowie, M. R., Marshall, D., Drummond, M., et al. (2009). Lifetime cost-effectiveness of prophylactic implantation of a cardioverter defibrillator in patients with reduced left ventricular systolic function: results of Markov modelling in a European population. Europace (European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology), 11(6), 716–726.

    Google Scholar 

  9. Krahn, A. D., Lee, D. S., Birnie, D., et al. (2011). Predictors of short-term complications after implantable cardioverter-defibrillator replacement: results from the Ontario ICD Database. Circulation. Arrhythmia and Electrophysiology, 4(2), 136–142.

    Article  PubMed  Google Scholar 

  10. Schaer, B. A., Koller, M. T., Sticherling, C., et al. (2009). Longevity of implantable cardioverter-defibrillators, influencing factors, and comparison to industry-projected longevity. Journal of Heart Rhythm Society, 6(12), 1737–1743.

    Article  Google Scholar 

  11. Thijssen, J., Borleffs, C. J. W., van Rees, J. B., et al. (2012). Implantable cardioverter-defibrillator longevity under clinical circumstances: an analysis according to device type, generation, and manufacturer. Journal of Heart Rhythm Society, 9(4), 513–519.

    Article  Google Scholar 

  12. Gregoratos, G., Cheitlin, M. D., Conill, A., et al. (1998). ACC/AHA Guidelines for Implantation of Cardiac Pacemakers and Antiarrhythmia Devices: executive summary—a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee on Pacemaker Implantation). Circulation, 97(13), 1325–1335.

    Article  CAS  PubMed  Google Scholar 

  13. Gregoratos, G., Abrams, J., Epstein, A. E., et al. (2002). ACC/AHA/NASPE 2002 guideline update for implantation of cardiac pacemakers and antiarrhythmia devices: summary article. A report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (ACC/AHA/NASPE Committee to update the 1998 Pacemaker Guidelines). Journal of Cardiovascular Electrophysiology, 13(11), 1183–1199.

    Article  PubMed  Google Scholar 

  14. Zipes, D. P., Camm, A. J., Borggrefe, M., et al. (2006). ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death: a report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death). Journal of the American College of Cardiology, 48(5), e247–346.

    Article  PubMed  Google Scholar 

  15. Horlbeck, F. W., Mellert, F., Kreuz, J., Nickenig, G., & Schwab, J. O. (2012). Real-world data on the lifespan of implantable cardioverter-defibrillators depending on manufacturers and the amount of ventricular pacing. Journal of Cardiovascular Electrophysiology, 23(12), 1336–1342.

    Article  PubMed  Google Scholar 

  16. Biffi, M., Ziacchi, M., Bertini, M., et al. (2008). Longevity of implantable cardioverter-defibrillators: implications for clinical practice and health care systems. Europace (European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology), 10(11), 1288–1295.

    Google Scholar 

  17. Alam, M. B., Munir, M. B., Rattan, R., et al. (2014). Battery longevity in cardiac resynchronization therapy implantable cardioverter defibrillators. Europace (European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology), 16(2), 246–251.

    Google Scholar 

  18. Shafat, T., Baumfeld, Y., Novack, V., Konstantino, Y., & Amit, G. (2013). Significant differences in the expected versus observed longevity of implantable cardioverter defibrillators (ICDs). Clinical Research Cardiology Off. J. Ger. Card. Soc., 102(1), 43–49.

    Article  Google Scholar 

  19. Gadler, F., Valzania, C., & Linde, C. (2015). Current use of implantable electrical devices in Sweden: data from the Swedish pacemaker and implantable cardioverter-defibrillator registry. Europace (European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology), 17(1), 69–77.

    Google Scholar 

  20. Van Welsenes, G. H., Van Rees, J. B., Thijssen, J., et al. (2011). Primary prevention implantable cardioverter defibrillator recipients: the need for defibrillator back-up after an event-free first battery service-life. Journal of Cardiovascular Electrophysiology, 22(12), 1346–1350.

    Article  PubMed  Google Scholar 

  21. Yap, S.-C., Schaer, B. A., Bhagwandien, R. E., et al. (2014). Evaluation of the need of elective implantable cardioverter-defibrillator generator replacement in primary prevention patients without prior appropriate ICD therapy. Heart British Cardiac Society, 100(15), 1188–1192.

    PubMed  Google Scholar 

  22. Borleffs, C. J. W., Thijssen, J., de Bie, M. K., et al. (2010). Recurrent implantable cardioverter-defibrillator replacement is associated with an increasing risk of pocket-related complications. Pacing Clinical Electrophysiology PACE, 33(8), 1013–1019.

    PubMed  Google Scholar 

  23. Poole, J. E., Gleva, M. J., Mela, T., et al. (2010). Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation, 122(16), 1553–1561.

    Article  PubMed  Google Scholar 

  24. Boriani, G., Braunschweig, F., Deharo, J. C., et al. (2013). Impact of extending device longevity on the long-term costs of implantable cardioverter-defibrillator therapy: a modelling study with a 15-year time horizon. Europace (European Journal of Pacing, Arrhythmias and Cardiac Electrophysiology of the European Heart Rhythm Association of the European Society of Cardiology), 15(10), 1453–1462.

    Google Scholar 

Download references

Acknowledgments

The research leading to the results has received funding from the European Community’s Seventh Framework Programme FP7/2007–2013 under grant agreement no. 602299, EU-CERT-ICD.

Conflict of interest

J. S., L. L., S. T. S., and M. Z. received travel grant support from Boston Scientific, Biotronik, and Medtronic. M. Z. received lecture fees and consultant fees from Medtronic, Biotronik, and Boston Scientific. P. M., T. F., M. L., H. W., and G. H. have no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Markus Zabel.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Seegers, J., Expósito, P.M., Lüthje, L. et al. Longevity of implantable cardioverter-defibrillators in a single-center population. J Interv Card Electrophysiol 44, 179–186 (2015). https://doi.org/10.1007/s10840-015-0044-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-015-0044-y

Keywords

Navigation