Rationale and design of the SPICE study—septal positioning of ventricular ICD electrodes

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Aim

Detrimental effects of right ventricular apical pacing have prompted the search for alternative pacing sites. Data from pacemaker studies suggest that the mid-septum may be a hemodynamically favorable pacing site within the right ventricle. However, data on the safety of implantable cardioverter defibrillator (ICD) lead placement in this location are limited. The aim of the SPICE study is to ascertain that implantation of a ventricular ICD lead at the mid-septum is not inferior to the traditional apical site in terms of feasibility and safety.

Methods

In total, 286 patients with the indication for ICD implantation will be included in the study and stratified in a 1:1 ratio to have the lead placed in the right ventricular apex or the mid-septum. Primary outcome measure of this randomized, prospective study is the event-free survival of lead-related complications at 3 months. The combined end point includes the feasibility to reach the randomized pacing site in combination with the achievement of adequate pacing, sensing, and defibrillation thresholds. Data analysis and sample size calculation are based on a non-inferiority design.

Conclusion

This trial will primarily assess the feasibility, safety, and efficacy of right ventricular ICD lead positioning in the mid-septum compared to the traditional apical position. Documentation of non-inferiority of the novel lead position in terms of applicability and safety will be the prerequisite for potential further trials that may evaluate long-term clinical effects associated with mid-septal lead positions.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1

Similar content being viewed by others

References

  1. Tantengco, M. V., Thomas, R. L., & Karpawich, P. P. (2001). Left ventricular dysfunction after long-term right ventricular apical pacing in the young. Journal of the American College of Cardiology, 37, 2093–2100.

    Article  PubMed  CAS  Google Scholar 

  2. Thambo, J. B., Bordachar, P., Garrigue, S., Lafitte, S., Sanders, P., Reuter, S., et al. (2004). Detrimental ventricular remodeling in patients with congenital complete heart block and chronic right ventricular apical pacing. Circulation, 110, 3766–3772.

    Article  PubMed  Google Scholar 

  3. Tse, H. F., & Lau, C. P. (1997). Long-term effect of right ventricular pacing on myocardial perfusion and function. Journal of the American College of Cardiology, 29, 744–749.

    Article  PubMed  CAS  Google Scholar 

  4. Nielsen, J. C., Bøttcher, M., Nielsen, T. T., Pedersen, A. K., & Andersen, H. R. (2000). Regional myocardial blood flow in patients with sick sinus syndrome randomized to long-term single chamber atrial or dual chamber pacing—effect of pacing mode and rate. Journal of the American College of Cardiology, 35, 1453–1461.

    Article  PubMed  CAS  Google Scholar 

  5. Karpawich, P. P., Rabah, R., & Haas, J. E. (1999). Altered cardiac histology following apical right ventricular pacing in patients with congenital atrioventricular block. Pacing and Clinical Electrophysiology, 22, 1372–1377.

    Article  PubMed  CAS  Google Scholar 

  6. Manolis, A. S. (2006). The deleterious consequences of right ventricular apical pacing: time to seek alternate site pacing. Pacing and Clinical Electrophysiology, 29, 298–315.

    Article  PubMed  Google Scholar 

  7. Wilkoff, B. L., Cook, J. R., Epstein, A. E., Greene, H. L., Hallstrom, A. P., Hsia, H., et al. (2002). Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the dual chamber and VVI implantable defibrillator (DAVID) trial. JAMA, 288, 3115–3123.

    Article  PubMed  Google Scholar 

  8. Olshansky, B., Day, J. D., Moore, S., Gering, L., Rosenbaum, M., McGuire, M., et al. (2007). Is dual-chamber programming inferior to single-chamber programming in an implantable cardioverter-defibrillator? Results of the INTRINSIC RV (Inhibition of unnecessary RV pacing with AVSH in ICDs) study. Circulation, 115, 9–16.

    Article  PubMed  Google Scholar 

  9. Victor, F., Mabo, P., Mansour, H., Pavin, D., Kabalu, G., de Place, C., et al. (2006). A randomized comparison of permanent septal versus apical right ventricular pacing: short-term results. Journal of Cardiovascular Electrophysiology, 17, 238–242.

    Article  PubMed  Google Scholar 

  10. Muto, C., Ottaviano, L., Canciello, M., Carreras, G., Calvanese, R., Ascione, L., et al. (2007). Effect of pacing the right ventricular mid-septum tract in patients with permanent atrial fibrillation and low ejection fraction. Journal of Cardiovascular Electrophysiology, 18, 1032–1036.

    Article  PubMed  Google Scholar 

  11. Kypta, A., Steinwender, C., Kammler, J., Leisch, F., & Hofmann, R. (2008). Long-term outcomes in patients with atrioventricular block undergoing septal ventricular lead implantation compared with standard apical pacing. Europace, 10, 574–579.

    Article  PubMed  Google Scholar 

  12. Zipes, D. P., Camm, A. J., Borggrefe, M., Buxton, A. E., Chaitman, B., Fromer, M., et al. (2006). ACC/AHA/ESC 2006 guidelines for management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Europace, 8, 746–837.

    Article  PubMed  Google Scholar 

  13. Epstein, A. E., DiMarco, J. P., Ellenbogen, K. A., Estes, N. A., 3rd, Freedman, R. A., Gettes, L. S., et al. (2008). ACC/AHA/HRS 2008 Guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines developed in collaboration with the American Association for Thoracic Surgery and Society of Thoracic Surgeons. Journal of the American College of Cardiology, 51, e1–e62.

    Article  PubMed  Google Scholar 

  14. Epstein, A. E., Baker, J. H., 2nd, Beau, S. L., Deering, T. F., Greenberg, S. M., & Goldman, D. S. (2009). Performance of the St. Jude Medical Riata leads. Heart Rhythm, 6, 2004–2009.

    Google Scholar 

  15. Schuger, C., Ellenbogen, K., Faddis, M., Knight, B. P., Yong, P., Sample, R., et al. (2006). Defibrillation energy requirements in an ICD population receiving cardiac resynchronization therapy. Journal of Cardiovascular Electrophysiology, 17, 247–250.

    Article  PubMed  Google Scholar 

  16. Day, J. D., Olshansky, B., Moore, S., Brown, S., Stolen, K. Q., Lerew, D. R., et al. (2008). High defibrillation energy requirements are encountered rarely with modern dual-chamber implantable cardioverter-defibrillator systems. Europace, 10, 347–350.

    Article  PubMed  Google Scholar 

  17. Lieberman, R., Grenz, D., Mond, H. G., & Gammage, M. D. (2004). Selective site pacing: defining and reaching the selected site. Pacing and Clinical Electrophysiology, 27, 883–886.

    Article  PubMed  Google Scholar 

  18. Mond, H. G., Hillock, R. J., Stevenson, I. H., & McGavigan, A. D. (2007). The right ventricular outflow tract: the road to septal pacing. Pacing and Clinical Electrophysiology, 30, 482–491.

    Article  PubMed  Google Scholar 

  19. Van Gelder, B. M., & Bracke, F. A. (2008). The ECG lead I paradox in cardiac resynchronization therapy. Pacing and Clinical Electrophysiology, 31, 1519–1521.

    Article  PubMed  Google Scholar 

  20. Giudici, M. C., Barold, S. S., Paul, D. L., Schrumpf, P. E., Van Why, K. J., & Orias, D. W. (2004). Right ventricular outflow tract placement of defibrillation leads: five year experience. Pacing and Clinical Electrophysiology, 27, 443–446.

    Article  PubMed  Google Scholar 

  21. Mollerus, M., Lipinski, M., & Munger, T. (2008). A randomized comparison of defibrillation thresholds in the right ventricular outflow tract versus right ventricular apex. Journal of Interventional Cardiac Electrophysiology, 22, 221–225.

    Article  PubMed  Google Scholar 

  22. Crossley, G. H., Boyce, K., Roelke, M., Evans, J., Yousuf, D., Syed, Z., et al. (2009). A prospective randomized trial of defibrillation thresholds from the right ventricular outflow tract and the right ventricular apex. Pacing and Clinical Electrophysiology, 32, 166–171.

    Article  PubMed  Google Scholar 

  23. Gula, L. J., Massel, D., Krahn, A. D., Yee, R., Skanes, A. C., & Klein, G. J. (2008). Is defibrillation testing still necessary? A decision analysis and Markov model. Journal of Cardiovascular Electrophysiology, 19, 400–405.

    Article  PubMed  Google Scholar 

  24. Kolb, C., Tzeis, S., & Zrenner, B. (2009). Defibrillation threshold testing: tradition or necessity? Pacing and Clinical Electrophysiology, 32, 570–572.

    Article  PubMed  Google Scholar 

  25. Panotopoulos, P., Krum, D., Axtell, K., Dhala, A., Sra, J., Akhtar, M., et al. (2001). Ventricular fibrillation sensing and detection by implantable defibrillators: is one better than the others? A prospective, comparative study. Journal of Cardiovascular Electrophysiology, 12, 445–452.

    Article  PubMed  CAS  Google Scholar 

  26. Wathen, M. S., DeGroot, P. J., Sweeney, M. O., Stark, A. J., Otterness, M. F., Adkisson, W. O., et al. (2004). Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators: pacing fast ventricular tachycardia reduces shock therapies (PainFREE Rx II) trial results. Circulation, 110, 2591–2596.

    Article  PubMed  Google Scholar 

Download references

Conflict of interest

The study is financially supported by St. Jude Medical Germany; however, St. Jude Medical did not participate in drafting of the study design. Dr. Christof Kolb has received lecture fees from Biotronik, Boston, Medtronic, Sorin, and St. Jude Medical and is or was an advisor to Biotronik, Sorin, and St. Jude Medical. Dr. Stylianos Tzeis was supported by a training fellowship provided by the European Heart Rhythm Association. Frank Birkenhauer is an employee of St. Jude Medical. All other authors state no conflict of interest, especially no financial relationship to the sponsor of the trial.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Christof Kolb.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kolb, C., Tzeis, S., Andrikopoulos, G. et al. Rationale and design of the SPICE study—septal positioning of ventricular ICD electrodes. J Interv Card Electrophysiol 31, 247–254 (2011). https://doi.org/10.1007/s10840-011-9575-z

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10840-011-9575-z

Keywords

Navigation