Abstract
Purpose
The high density mesh ablator (HDMA), a mesh electrode catheter using radiofrequency delivery for pulmonary vein isolation (PVI) in patients with atrial fibrillation (AF), has been shown to have a high acute ablation success rate. The primary endpoint of our study was the efficacy of HDMA to maintain sinus rhythm (SR), and the secondary endpoint is its efficacy to reduce AF burden in an 18-month follow-up.
Methods
Forty-two patients with AF [25 patients with paroxysmal AF (PAF, 59.5%), 17 patients with persistent AF (persAF, 40.5%)], were followed up with 7-day Holter monitoring and anamnestic survey at 6 and 12 months after PVI. To evaluate symptomatic AF episodes after 12 months, we used data from further hospital stays or information given by phone.
Results
In 42 patients at intervals of 6 and 12 months, stable SR was documented in 7-day Holter monitoring in 18 patients (42.9%, 10/25 and 9/25 for PAF and 8/17 and 9/17 for persAF, respectively). Symptomatic episodes of AF were reported in 31 patients (73.8%, 18/25 for PAF and 13/17 for persAF). Re-ablation was performed in seven patients (3/25 for PAF and 4/17 for persAF) not yielding to stable SR. At an 18-month follow-up (12–29 months), AF recurrence persisted in all 42 patients; in four patients, a permanent pacemaker was implanted (9.5%). No complications attributable to the HDMA occurred in this series.
Conclusions
Our findings suggest that the PVI using the HDMA is a safe method but yields to a high AF recurrence rate at long-term follow-up even after a second procedure with the same tool.
Similar content being viewed by others
Abbreviations
- AF:
-
Atrial fibrillation
- AAD:
-
Antiarrhythmic drugs
- HDMA:
-
High density mesh ablator
- LA:
-
Left atrium
- LIPV:
-
Left inferior pulmonary vein
- LSPV:
-
Left superior pulmonary vein
- PAF:
-
Paroxysmal atrial fibrillation
- PersAF:
-
Persistent atrial fibrillation
- PV:
-
Pulmonary vein
- RF:
-
Radio frequency
- RIPV:
-
Right inferior pulmonary vein
- RSPV:
-
Right superior pulmonary vein
- SR:
-
Sinus rhythm
References
Cox, J. L., Boineau, J. P., Schuessler, R. B., Krater, K. M., & Lappas, D. G. (1993). Five year experience with the Maze procedure for atrial fibrillation. The Annals of Thoracic Surgery, 56, 814–824.
Haïssaguerre, M., Jaïs, P., Shah, D., Takahashi, A., Hocini, M., Quiniou, G., et al. (1998). Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. The New England Journal of Medicine, 339, 659–666.
Meissner, A., Christ, M., Maagh, P., Borchard, R., van Bracht, M., Wickenbrock, I., et al. (2007). Quality of life and occurrence of atrial fibrillation in long term follow up of common type atrial flutter: ablation with irrigated 5 mm tip- and conventional 8 mm tip electrode. Clinical Research in Cardiology, 96, 794–802.
Oral, H., Knight, B. P., Tada, H., Özaydin, M., Chugh, A., Hassan, S., et al. (2002). Pulmonary vein isolation for paroxysmal and persistent atrial fibrillation. Circulation, 105, 1077–1081.
Pappone, C., Oreto, G., Rosanio, S., Vicedomini, G., Tocchi, M., Gugliotta, F., et al. (2001). Atrial electroanatomical remodeling after circumferential radiofrequency pulmonary vein isolation: efficacy of an anatomic approach in a large cohort of patients with atrial fibrillation. Circulation, 104, 2539–2544.
Haïssaguerre, M., Shah, D., Jaïs, P., Hocini, M., Yamane, T., Deisenhofer, I., et al. (2000). Electrophysiological breakthroughs from the left atrium to the pulmonary veins. Circulation, 102, 2463–2465.
Geha, A. S., & Abdelhady, K. (2008). Current status of the surgical treatment of atrial fibrillation. World Journal of Surgery, 32, 346–349.
Marrouche, N. F., Dresing, N. F., Cole, C., Bash, D., Saad, E., Balaban, K., et al. (2002). Circular mapping and ablation of the pulmonary veins for treatment of atrial fibrillation: impact of different catheter techniques. Journal of the American College of Cardiology, 40, 464–474.
Takahashi, A., Iesaka, Y., Takahashi, Y., Takahashi, R., Kobayashi, K., Takagi, K., et al. (2002). Electrical connections between pulmonary veins: implication for ostial ablation of pulmonary veins in patients with atrial fibrillation. Circulation, 105, 2998–3003.
Karch, M. R., Zrenner, B., Deisenhofer, I., Schreieck, J., Ndrepepa, G., Dong, J., et al. (2005). Freedom of atrial tachyarrhythmias after catheter ablation of atrial fibrillation: a randomized comparison between 2 current ablation strategies. Circulation, 111, 2875–2880.
Oral, H., Scharf, C., Chugh, A., Hall, B., Cheung, P., Good, E., et al. (2003). Catheter ablation for paroxysmal atrial fibrillation. Segmental pulmonary vein ostial ablation versus left atrial ablation. Circulation, 108, 2355–2360.
Wright, M., Haïssaguerre, M., Knecht, S., et al. (2008). State of the art: catheter ablation of atrial fibrillation. Journal of Cardiovascular Electrophysiology, 19, 583–592.
Ouyang, F., Bänsch, D., Ernst, S., Schaumann, A., Hachiya, H., Chen, M., et al. (2004). Complete isolation of left atrium surrounding the pulmonary veins. New insights from the double-Lasso technique in paroxysmal atrial fibrillation. Circulation, 110, 2090–2096.
Meissner, A., Plehn, G., van Bracht, M., Schrage, M.O., Christ, M., Maagh, P., Butz, T., Borchard, R., Trappe, H.J. (2009). First experiences for pulmonary vein isolation with the high-density mesh ablator (HDMA): a novel mesh electrode catheter for both mapping and radiofrequency delivery in a single unit. Journal of Cardiovascular Electrophysiology, 20, 359–366
Gage, B. F., Watermann, A. D., Shannon, W., Boechler, M., Rich, M. W., & Radford, M. J. (2001). Validation of clinical classification schemes for predicting stroke: results from the National Registry of Atrial Fibrillation. JAMA, 285, 2864–2870.
Steinwender, C., Hoenig, S., Leisch, F., & Hofmann, R. (2009). Acute results of pulmonary vein isolation in patients with paroxysmal atrial fibrillation using a single mesh catheter. Journal of Cardiovascular Electrophysiology, 20, 147–152.
Lickfett, L., Schwab, J. O., & Lewalter, T. (2008). Advanced mapping techniques in atrial fibrillation. Journal of Interventional Cardiac Electrophysiology, 22, 155–159.
Wijffels, M. C., Van Oosterhout, M., Boersma, L. V., et al. (2009). Characterization of in vitro and in vivo lesions made by a novel multichannel ablation generator and a circumlinear decapolar ablation catheter. Journal of Cardiovascular Electrophysiology, 20, 1142–1148.
Kanj, M., Wazni, O., & Natale, A. (2007). Pulmonary vein antrum isolation. Heart Rhythm, 4(3 supplement), p73–p79.
De Filoppo, P., Sheng He, D., Brambilla, R., Gavazzi, A., & Cantu, F. (2009). Clinical experience with a single catheter for mapping and ablation of pulmonary vein ostium. Journal of Cardiovascular Electrophysiology, 20, 367–373.
Ouyang, F., Antz, M., Ernst, S., Hachiya, H., Mavrakis, H., Deger, F. T., et al. (2005). Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation, 11, 127–135.
Mansour, M., Forleo, G. B., Pappalardo, A., Heist, K., Avella, A., Laurenzi, F., et al. (2008). Initial experience with the mesh catheter for pulmonary vein isolation in patients with paroxysmal atrial fibrillation. Heart Rhythm, 5, 1510–1516.
Page, R. L., Wilkinson, W. E., Clair, W. K., McCarthy, E. A., & Pritchett, E. L. (1994). Asymptomatic arrhythmias in patients with symptomatic paroxysmal atrial fibrillation and paroxysmal supraventricular tachycardia. Circulation, 89, 224–227.
Patten, M., Maas, R., Karim, A., Mu¨ller, H. W., Simonossky, R., & Meinertz, T. (2006). Event-recorder monitoring in the diagnosis of atrial fibrillation in symptomatic patients: subanalysis of the SOPAT trial. Journal of Cardiovascular Electrophysiology, 17, 1216–1220.
Hindricks, G., Piorkowski, C., Tanner, H., Kobza, R., Gerds-Li, J. H., Carbucicchio, C., et al. (2005). Perception of atrial fibrillation before and after radiofrequency catheter ablation: relevance of asymptomatic arrhythmia recurrence. Circulation, 112, 307–313.
Franck, H., Van Erven, L., on behalf of the EHRA Scientific Initiatives Committee (SIC). (2009). Validation of success following atrial fibrillation ablation: a European survey. Europace, 11, 1002–1003.
Piorkowski, C., Kottkamp, H., Tanner, H., Kobza, R., Nielsen, J. C., Arya, A., et al. (2005). Value of different follow-up strategies to assess the efficacy of circumferential pulmonary vein ablation for the curative treatment of atrial fibrillation. Journal of Cardiovascular Electrophysiology, 16(12), 1286–1292.
Steven, D., Rostock, T., Lutomsky, B., Klemm, H., Servatius, H., Drewitz, I., et al. (2008). What is the real atrial fibrillation burden after catheter ablation of atrial fibrillation? A prospective rhythm analysis in pacemaker patients with continuous atrial monitoring. European Heart Journal, 29, 1037–1042.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Maagh, P., van Bracht, M., Butz, T. et al. Eighteen months follow-up of the clinical efficacy of the high density mesh ablator (HDMA) in patients with atrial fibrillation after pulmonary vein isolation. J Interv Card Electrophysiol 29, 43–52 (2010). https://doi.org/10.1007/s10840-010-9498-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-010-9498-0