Abstract
Purpose
The aim of the present study is to evaluate if remote monitoring with the CareLink Network may improve clinical management of tachyarrhythmias and heart failure episodes in patients treated with biventricular defibrillators (CRT-D).
Methods
Patients implanted with CRT-D for more than 6 months received the CareLink monitor and were trained to perform device interrogation. At-home transmissions were scheduled at 2 weeks, 1 and 2 months after training, with a final in-office visit after 3 months.
Results
Sixty-seven patients performed 264 data transmissions. Twenty-three unscheduled data transmissions were requested by the centers after patient contact. Ventricular tachyarrhythmias were reported in nine patients during 16 data reviews. Thirteen data reviews (81%) were performed remotely via CareLink transmissions (nine scheduled and four unscheduled), in seven patients. Of these events, in two cases (15%) in-hospital visits were requested, while in 11 (85%) no action was needed and no additional in-clinic visits were scheduled. During the study period, in 20/28 (71%) intra-thoracic impedance alerts, the patients remotely transmitted their device data. After remote data review, in ten cases drug therapy was adjusted by phone and in four cases no action was needed and the patient reassured. In six episodes an in-hospital extra visit was scheduled. On the whole, in 14 cases (70%), the patient could be managed remotely avoiding a visit to the hospital.
Conclusions
Our study showed that remote follow-up is an efficient method to manage tachyarrhythmias and heart failure episodes in CRT-D patients. Early reaction to clinical events may improve overall patient care.
Similar content being viewed by others
References
Moss, A. J., Hall, W. J., Cannom, D. S., Daubert, J. P., Higgins, S. L., Klein, H., et al. (1996). Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. New England Journal of Medicine, 335, 1933–1940.
Buxton, A. E., Lee, K. L., Fisher, J. D., Josephson, M. E., Prystowsky, E. N., & Hafley, G. (1999). A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. New England Journal of Medicine, 341, 1882–1890.
Moss, A. J., Zareba, W., Hall, W. J., Klein, H., Wilber, D. J., Cannom, D. S., et al. (2002). Multicenter automatic defibrillator implantation trial II investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. New England Journal of Medicine, 346, 877–883.
Bardy, G. H., Lee, K. L., Mark, D. B., Poole, J. E., Packer, D. L., Boineau, R., Domanski, M., et al. (2005). Sudden Cardiac Death in Heart Failure Trial (SCD-HeFT) Investigators. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. New England Journal of Medicine, 352, 225–237.
Cleland, J. G., Daubert, J. C., Erdmann, E., Freemantle, N., Gras, D., Kappenberger, L., et al. (2005). Cardiac Resynchronization-Heart Failure (CARE-HF) Study investigators. The effect of cardiac resynchronization on morbidity and mortality in heart failure. New England Journal of Medicine, 352, 1539–1549.
Bristow, M. R., Saxon, L. A., Boehmer, J., Krueger, S., Kass, D. A., De Marco, T., et al. (2004). Comparison of Medical Therapy, Pacing, and Defibrillation in Heart Failure (COMPANION) Investigators. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. New England Journal of Medicine, 350, 2140–2150.
Hunt, S. A., Abraham, W. T., Chin, M. H., Feldman, A. M., Francis, G. S., Ganiats, T. G., et al. (2005). American College of Cardiology; American Heart Association Task Force on Practice Guidelines; American College of Chest Physicians; International Society for Heart and Lung Transplantation; Heart Rhythm Society. ACC/AHA 2005 guideline update for the diagnosis and management of chronic heart failure in the adult: summary article: a report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Writing Committee to Update the 2001 Guidelines for the Evaluation and Management of Heart Failure). Circulation, 112, 1116–1143.
Swedberg, K., Cleland, J., Dargie, H., Drexler, H., Follath, F., Komajda, M., et al. (2005). Task force for the diagnosis and treatment of chronic heart failure of the European society of cardiology. Guidelines for the diagnosis and treatment of chronic heart failure: executive summary (update 2005): The task force for the diagnosis and treatment of chronic heart failure of the European society of cardiology. European Heart Journal, 26, 1115–1140.
Schoenfeld, M. H., Compton, S. J., Mead, R. H., Weiss, D. N., Sherfesee, L., Englund, J., et al. (2004). Remote monitoring of implantable cardioverter defibrillators: a prospective analysis. Pacing and Clinical Electrophysiology, 27(6 Pt 1), 757–763.
Lazarus, A. (2007). Remote, wireless, ambulatory monitoring of implantable pacemakers, cardioverter defibrillators, and cardiac resynchronization therapy systems: analysis of a worldwide database. Pacing and Clinical Electrophysiology, 30(Suppl 1), S2–S12.
Saxon, L. A., Boehmer, J. P., Neuman, S., & Mullin, C. M. (2007). Remote Active Monitoring in Patients with Heart Failure (RAPID-RF): design and rationale. Journal of Cardiac Failure, 13, 241–246.
Rosenqvist, M., Beyer, T., Block, M., den Dulk, K., Minten, J., & Lindemans, F. (1998). Adverse events with transvenous implantable cardioverter-defibrillators: a prospective multicenter study. European 7219 Jewel ICD investigators. Circulation, 98, 663–670.
Theuns, D. A., Klootwijk, A. P., Goedhart, D. M., & Jordaens, L. J. (2004). Prevention of inappropriate therapy in implantable cardioverter-defibrillators: results of a prospective, randomized study of tachyarrhythmia detection algorithms. Journal of the American College of Cardiology, 44, 2362–2367.
Perings, C., Klein, G., Toft, E., Moro, C., Klug, D., Bocker, D., et al. (2006). RIONI Investigators. The RIONI study rationale and design: validation of the first stored electrograms transmitted via home monitoring in patients with implantable defibrillators. Europace, 8, 288–292.
Koos, R., Sinha, A. M., & Stellbrink, C. (2005). Home Monitoring in an ICD patient with incessant ventricular tachycardia. Zeitschrift fuÉr Kardiologie, 94, 461–464.
Lunati, M., Gasparini, M., Santini, M., Landolina, M., Perego, G. B., Pappone, C., et al. (2008). on behalf of the InSync ICD Italian Registry Investigators. Follow-Up of CRT-ICD: Implications for the Use of Remote Follow-Up Systems. Data from the InSync ICD Italian Registry. Pacing and Clinical Electrophysiology, 31, 38–46.
Varma, N., Stambler, B., & Chun, S. (2005). Detection of atrial fibrillation by implanted devices with wireless data transmission capability. Pacing and Clinical Electrophysiology, 28(Suppl 1), S133–S136.
Ricci, R. P., Morichelli, L., & Santini, M. (2008). Home monitoring remote control of pacemaker and ICD patients in clinical practice. Impact on medical management and health care resource utilization. Europace, 10, 164–170.
Yu, C. M., Wang, L., Chau, E., Chan, R. H., Kong, S. L., Tang, M. O., et al. (2005). Intrathoracic impedance monitoring in patients with heart failure: correlation with fluid status and feasibility of early warning preceding hospitalization. Circulation, 112, 841–848.
Vollmann, D., Nagele, H., Schauerte, P., Wiegand, U., Butter, C., Zanotto, G., et al. (2007). European InSync Sentry Observational Study Investigators. Clinical utility of intrathoracic impedance monitoring to alert patients with an implanted device of deteriorating chronic heart failure. European Heart Journal, 28, 1835–1840.
Knackstedt, C., Mischke, K., Schimpf, T., Waringer, J., Fache, K., Frechen, D., et al. (2008). Integration of automatic intrathoracic fluid content measurement into clinical decision making in patients with congestive heart failure. Pacing and Clinical Electrophysiology, 31, 961–967.
Scholten, M. F., Thornton, A. S., Theuns, D. A., Res, J., & Jordaens, L. J. (2004). Twiddler’s syndrome detected by home monitoring device. Pacing and Clinical Electrophysiology, 27, 1151–1152.
Schoenfeld, M. H., & Markowitz, H. T. (2000). Device follow-up in the age of automaticity. Pacing and Clinical Electrophysiology, 23, 803–806.
Acknowledgements
The authors would like to thank Walter Bernardini and Marco Vimercati (clinical service team, Medtronic Italy) for the data management, and Jane Moore for her assistance in the editing of this manuscript.
Conflict of Interest Disclosure
Serena Gilardi and Sergio Valsecchi are employees of Medtronic, Inc.
No other conflicts exist.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Santini, M., Ricci, R.P., Lunati, M. et al. Remote monitoring of patients with biventricular defibrillators through the CareLink system improves clinical management of arrhythmias and heart failure episodes. J Interv Card Electrophysiol 24, 53–61 (2009). https://doi.org/10.1007/s10840-008-9321-3
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10840-008-9321-3