Introduction

Dry eye disease (DED) is a multifactorial syndrome related to impaired tear film homeostasis and ocular surface inflammation [1]. Because of the large discrepancy between a patient’s symptoms, the observed signs, and the dry eye symptoms overlapping with those of other ocular surface diseases, diagnosing DED is difficult in practice [2]. Therefore, a new diagnostic device using technologies such as interferometry and meibography is desirable.

The tear film consists of two phases, a lipid layer overlying a muco-aqueous layer [3]. Tear lipids secreted from the meibomian glands (MG) aids the tear film maintaining its solidity [4]. Meibomian gland dysfunction (MGD), the main cause of evaporative dry eye, is more frequent than aqueous-deficient dry eye [5]. The evaluation of both the function and morphology of MG is essential for the diagnosis of MGD [6]. Meibography can detect the morphological alterations of MG, whereas tear interferometry permits assessments of the lipid layer of the tear film. Such assessments of MG morphology provide clinical evidence contributing to the diagnosis of evaporative dry eye, whereas that of the lipid layer of the tear film allows the monitoring of MG function [7]. Blinking can retain a dynamic balance of ocular surface tear capacity and is important for the development and distribution of the lipid layer [8, 9].

Various instruments evaluating DED are currently being used [10]. The most used device is the LipiView® II ocular surface interferometer (TearScience Inc., Morrisville, NC, USA) (LVII), launched in 2011. It provides an absolute LLT, a partial blink rate (PBR), and an imaging of the MG. Similar to the LVII, IDRA® Ocular surface analyzer (SBM SISTEMI, Inc., Torino, Italy) (IDRA), which was released in 2018, provides LLT, MG dropouts, non-invasive tear break-up time (NITBUT), tear meniscus height (TMH), and blinking pattern [11].

There was no significant difference between LVII and IDRA in LLT measurements, PBR was measured higher in IDRA than LVII [11]. A paper has been reported using LVII to investigate the relationship between partial blinking and ocular surface parameters related with DED [12], but no research has identified the correlation of PBR to other dry eye parameters measured using IDRA.

This study aimed to elucidate the relationship between blinking pattern and ocular surface parameters by IDRA in DED patients.

Materials and methods

We conducted a retrospective cross‐sectional study of the patients who visited the Kim’s Eye Hospital between March 2020 and July 2020 when the coronavirus disease 2019 (COVID-19) pandemic was raging. By May 2020, there were more than 4 million confirmed COVID-19 cases worldwide, with nearly 300,000 deaths [13]. There is a high risk of transmission of COVID-19 because ophthalmological practice during both slit lamp examination and direct ophthalmoscopy are conducted close to the patient's face. Therefore, our hospital made thorough efforts to prevent infection. All patients filled out questionnaires on fever, respiratory symptoms, and whether they visited other countries within 14 days. In addition, a mandatory non-contact temperature check was conducted at the hospital entrance. All patients and hospital staff were required to wear masks throughout their stay in the hospital. Hospital staff including doctors, nurses, and examiners washed their hands using waterless hand sanitizer before and after contact with patients. Participants in the study, who visited the hospital as an outpatient, did not take the COVID-19 test because only hospitalized patients were tested.

This study received ethical approval from the Institutional Review Board at Kim’s Eye Hospital, Seoul, Republic of Korea (2020–08-001) and adhered to the tenets of the Declaration of Helsinki. Considering the retrospective nature of the study and the use of deidentified patient data, the written informed consent was waived by the Institutional Review Board of Kim’s Eye Hospital, Seoul, Republic of Korea. Data of blinking dynamics, tear film LLT, NITBUT, and MG dropout rates were assessed using IDRA.

Inclusion criteria were adult patients from 20 to 80 years old who met the following two criteria: Standard Patient Evaluation of Eye Dryness (SPEED) ≥ 6 points and NITBUT < 10 s. Patients were excluded from the study if they were diagnosed with supra-nuclear motor weakness influencing movements of the eyelid, ptosis, pre-existing ocular pathology, history of ocular surgery or trauma, history of use of any systemic medications, with systemic autoimmune or connective tissue disease including Sjögren syndrome, had a history of using a punctual plug, previous eye infection, and contact lens usage. A single investigator (JYJ) handled the IDRA throughout the study. Test values of only the right eye were evaluated.

Dry eye questionnaire

DED symptoms were evaluated using the SPEED validated questionnaire (0–28) [14, 15]. A single ophthalmologist (KK) diagnosed DED based on the Tear Film and Ocular Surface Society Dry Eye Workshop II criteria, with SPEED ≥ 6 points [16]. A previous study showed that the SPEED questionnaire was compatible with the Ocular Surface Disease Index (OSDI) [17].

Interferometric assessment of lipid layer thickness

The IDRA, released in 2018, is a novel device for the specific evaluation of tear film permitting a rapid detailed structural analysis of the tear components. The device can analyze all layers of the tear film and MG, allowing clinicians to determine the components to be treated, according to the type of insufficiency. IDRA performs a non-invasive test for approximately 5 min, and it should be incorporated between a slit lamp and biomicroscope. Its pins have been built to fit completely into the hole, found when the plate used for the tonometer is eliminated. The participants sat comfortably using a chin holder and were asked to look at the camera with natural blinking of their eyes. Videos can be recorded for a few seconds, with the adjustment of the recording time according to the convenience of the clinician. In this study, it was recorded for 20 s to proceed. The device projects white light over the cornea, and the light reflected from the tear film can be observed as a white fan-shaped area covering the lower third of the cornea. The automatic interferometry IDRA test detected the interference of colors from the lipid layer on the tear film. It determined the average, maximum, and minimum LLT using the international grade scale with the thicknesses related to each grade of the lipid layer pattern (Fig. 1) [18, 19]. Depending on the patterns, the grades were converted to nanometers and could be classified between 15 and 100 nm. Like LVII, IDRA has an upper cutoff of 100 nm [16].

Fig. 1
figure 1

a Image about lipid layer thickness (LLT) of 58 years old male patient taken with the IDRA®. b Compared to the image of LipiView® II, there are smaller areas that can be identified by interferometry. In this patient, the LLT values were measured by IDRA® as follows: average 68 nm, maximum 85 nm, and minimum 58 nm

MG dropout measurement

The LVII only produces MG dropout results on the image. In contrast, IDRA calculated the MG dropout rate of lower lid in the resulting item (Fig. 2a). The dropout rate was calculated as a percentage by dividing the non-glandular zone by the total visible area of the lower lid [20].

Fig. 2
figure 2

a IDRA® image shows meibomian gland (MG) dropout of 40 years old male patient. The lower eyelids were turned over and MGs were observed using an infrared transmitting filter, meibomian glands are apparent as areas of high reflectivity. IDRA® automatically calculates the dropout rate of MGs, and this patient came out at 45%. As IDRA® calculates the remaining part of meibomian glands, so to subtract the dropout ratio, it was written as the value subtracted from 100. b IDRA® image about blinking pattern of 72-year-old female patient. It shows a partial blink 2 times per 20 s, total blink 7 times per 20 s, and partial/total blink ratio 0.29

Blinking pattern measurement

Participants were asked to blink freely. IDRA automatically detects and analyzes blink rate and quality using recorded video (Fig. 2b). It is numerically displayed the number of complete and partial blinks and blink frequencies. Each stage in the blinking cycle was measured and recorded during the examination (Fig. 2B), and the partial blinks were defined as blinks without touching the upper and lower eyelids [12].

NITBUT and TMH

IDRA can determine NITBUT by using the projected ring patterns from a Placido’s disk onto the cornea. NITBUT evaluates the stability of the tear film by measuring the time from the full blink to the presence of the first disruption of the reflected image on the cornea in seconds. Participants were required to blink twice and then keep their eyes open as long as possible. IDRA can also measure TMH, which can be recorded non-non-contactinvasively in a flash by taking a photograph. The device can detect the upper and lower tear meniscus and evaluate the TMH along the lower lid margin in the photograph.

Statistical analysis

Data were analyzed with the IBM Statistical Package for the Social Sciences (SPSS) version 22 (Chicago, USA), and a p value of < 0.05 was considered statistically significant. Continuous variables were presented as means ± standard deviation. Statistical analyses were performed using linear regression analysis, Pearson correlation analysis, and Student’s t tests. Pearson correlation analysis and linear regression analysis were used to evaluate the correlation between the SPEED scores and ocular surface data obtained by IDRA. Correlations of each value were analyzed with the Pearson correlation coefficient (r). Student’s t tests were used to compare ocular surface data between males and females.

Results

Fifty-seven patients with dry eye were recruited. Among them, 10 patients with LLT ≥ 100 nm were excluded. This could not be precisely calculated as IDRA has an upper cutoff of 100 nm. We finally enrolled 16 men and 31 women (mean age = 56.8 ± 14.5 [21–79] years) in our study. Mean age, spherical equivalent, non-contact tonometry measurements, and central corneal thickness are listed in Table 1. Comparisons by sex revealed no significant difference in characteristics, except for the TMH value. The TMH of female patients (0.17 ± 0.07) was significantly higher than that of males (0.14 ± 0.04) (p = 0.041). Looking at the clinical dry eye parameters of the patients, the average SPEED score was 12.53 ± 5.33 points, average NITBUT measured by IDRA was 8.22 ± 1.23 s, and TMH was 0.16 ± 0.06 nm. The mean value of average LLT measured was 75.45 ± 12.19 nm, and mean value of maximum LLT was 85.53 ± 13.33 nm, and mean value of minimum LLT was 66.11 ± 13.99 nm. The mean values of blink dynamics are shown in Table 1.

Table 1 Clinical evaluations and measurements of the ocular surface in patients with non-Sjögren dry eye disease

There was no significant relationship between LLT and SPEED or the MG dropout rate. Among the ocular surface parameters, PBR had strong correlations with the MG dropout rate (r = 0.811; p < 0.01) and moderate correlations with SPEED scores (r = 0.596; p < 0.01) (Fig. 3a and 3b). The MG dropout rate had moderate correlations with SPEED scores (r = 0.416; p = 0.04) (Fig. 3c). The results of correlations between ocular surface parameters, SPEED questionnaires, and blinking parameters are summarized in Table 2.

Fig. 3
figure 3

a Pearson correlation scatter plot for partial blinking rate and meibomian gland dropout rate. It showed significant correlations (Spearman’s r = 0.811, p < 0.01). b Pearson correlation scatter plot for partial blinking rate and Standard Patient Evaluation of Eye Dryness (SPEED) questionnaire scores. It showed significant correlations (Spearman’s r = 0.596, p < 0.01). c Pearson correlation scatter plot for SPEED scores and meibomian gland dropout rate. It showed significant correlations (Spearman’s r = 0.416, p = 0.04)

Table 2 Correlation between SPEED questionnaire scores and ocular surface parameters measured by IDRA® Ocular surface analyzer (SBM SISTEMI, Inc., Torino, Italy)

Discussion

In our study, PBR was correlated with the MG dropout rate and SPEED score. The MG dropout rate correlated with the SPEED score. It is the first study to assess the relationship between blinking patterns and ocular surface parameters measured by IDRA. Many aspects, such as ocular surface disorders, psychological status, and systemic diseases, may affect the blinking rate [21]. Blinking serves a vital function in preserving moisture and unity of the lipid layer’s ocular surface and extending the tear lipids [22]. Inadequate lipid distribution may occur, which may increase evaporation with a rise in partial blinking [23]. This is associated with TBUT and MGD, possibly through its contribution to MG obstruction and subsequent loss of tear film homeostasis [12]. Right blinking and stability of tear film are critical elements in the protection of the ocular surface [23].

Previous studies have shown that the volume of the lipid layer is correlated to the number and function of MGs [24]. LLT and PBR are significantly related to DED symptoms [23]. When comparing for LLT, the field of the white light projected to cause interference of colors is directed at the lower third of the cornea, approximately 1 mm above the inferior tear meniscus, in LVII [16], while it is directed slightly higher, approximately 2 mm above the inferior tear meniscus, in IDRA (Fig. 1).

A study in DED patients showed that increased age was positively associated with LLT [25]. In a retrospective analysis of 153 patients, SPEED values showed a significant correlation with age and LLT [15]. Like this study, SPEED scores were correlated with age. However, there was no statistically significant correlation between SPEED values and LLT in our study. This might be because of the small number of participants in our study or because of the characteristics of IDRA.

Figure 2 shows the MG dropout display of an MGD patient. The lower eyelids were turned over and MGs were observed using an infrared transmitting filter, while MGs were apparent as areas of high reflectivity. IDRA automatically calculates the dropout rate of MGs. As IDRA calculates the remaining part of the MGs to subtract the dropout ratio, it was written as the value subtracted from 100. IDRA recognizes alterations of the light intensity across the surface and compensates for the variations in lid thickness between participants through the illumination attached to the lid everter, known as the adaptive transillumination. This transillumination also forms shadows on areas where MGs are located; thus, any glands located below the visible exterior or at the suboptimal positions may be seen with the illumination.

IDRA has the advantage of being able to conduct tests for various ocular surface parameters with one equipment. The measurement of LLT through auto-interferometry along with that of TMH, auto-NIBUT, blinking quality, meibography, and bulbar redness classification is feasible. In addition, white-to-white measurement and pupillometry are possible with this device.

There are several limitations in interpreting the results of our study. First, the relatively small number of participants included in the study may have influenced the results. Second, the study subjects were homogenous, consisting only of the Korean population. Future studies needed to verify whether other races have the same results in a larger number of participants. Despite the limitations, we believe that our results showed a clear correlation between ocular surface parameters of the IDRA.

In conclusion, PBR was correlated with the MG dropout rate and SPEED score. The MG dropout rate correlated with the SPEED score. IDRA has the advantage of being able to conduct tests for various ocular surface parameters with one equipment.