Abstract
Purpose
In cases of pupillary capture after previous transscleral fixation of an intraocular lens (IOL), an across-pupil suture technique is sometimes used to stabilize the IOL. We investigated the optical effects of 10-0 polypropylene sutures placed across the pupil.
Methods
Optical performance was evaluated using the optical bench metrology system, and a single-piece IOL was used in an optical bench system consisting of a model eye, a high-resolution Hartmann–Shack wave front sensor, and an image capturing device with 2.0, 3.0, 4.0, and 4.5 mm apertures with distance focus. To simulate across-pupil sutures, two 10-0 polypropylene sutures were placed 2 mm apart across the pupil. The focus image, spherical aberration, and image quality were measured and compared with and without sutures.
Results
When pupil size increased, spherical aberration increased, irrespective of sutures. No difference was found in spherical aberration with and without sutures, and image qualities at 17.6 and 35.2 cycles per degree were not affected by the across-pupil sutures.
Conclusions
The differences in optical quality using across-pupil 10-0 polypropylene sutures were clinically negligible.
Similar content being viewed by others
References
Jin GJ, Crandall AS, Jones JJ (2005) Changing indications for and improving outcomes of intraocular lens exchange. Am J Ophthalmol 140:688–694
Fernandez-Buenaga R, Alio JL, Perez-Ardoy AL et al (2013) Late in-the-bag intraocular lens dislocation requiring explantation: risk factors and outcomes. Eye 27:795–801
Masket S, Fram NR (2013) Safety-basket suture for management of malpositioned posterior chamber intraocular lens. J Cataract Refract Surg 39:1633–1635
Jurgens I, Rey A (2015) Simple technique to treat pupillary capture after transscleral fixation of intraocular lens. J Cataract Refract Surg 41:14–17
Bading G, Hillenkamp J, Sachs HG et al (2007) Long-term safety and functional outcome of combined pars plana vitrectomy and scleral-fixated sutured posterior chamber lens implantation. Am J Ophthalmol 144:371–377
Higashide T, Shimizu F, Nishimura A, Sugiyama K (2009) Anterior segment optical coherence tomography findings of reverse pupillary block after scleral-fixated sutured posterior chamber intraocular lens implantation. J Cataract Refract Surg 35:1540–1547
Nagamoto S, Kohzuka T, Nagamoto T (1998) Pupillary block after pupillary capture of an AcrySof intraocular lens. J Cataract Refract Surg 24:1271–1274
Piers PA, Norrby NE, Mester U (2004) Eye models for the prediction of contrast vision in patients with new intraocular lens designs. Opt Lett 29:733–735
Lang AJ, Lakshminarayanan V, Portney V (1993) Phenomenological model for interpreting the clinical significance of the in vitro optical transfer function. J Opt Soc Am A: 10:1600–1610
Kawamorita T, Uozato H, Aizawa D et al (2009) Optical performance in rezoom and array multifocal intraocular lenses in vitro. J Refract Surg 25:467–469
Miller JM, Anwaruddin R, Straub J, Schwiegerling J (2002) Higher order aberrations in normal, dilated, intraocular lens, and laser in situ keratomileusis corneas. J Refract Surg 18:S579–S583
Kim MJ, Yoo YS, Joo CK, Yoon G (2015) Evaluation of optical performance of 4 aspheric toric intraocular lenses using an optical bench system: influence of pupil size, decentration, and rotation. J Cataract Refract Surg 41:2274–2282
Shiratani T, Shimizu K, Fujisawa K et al (2008) Crystalline lens changes in porcine eyes with implanted phakic IOL (ICL) with a central hole. Graefes Arch Clin Exp Ophthalmol 246:719–728
Uozato H, Shimizu K, Kawamorita T, Ohmoto F (2011) Modulation transfer function of intraocular collamer lens with a central artificial hole. Graefes Arch Clin Exp Ophthalmol 249:1081–1085
Acknowledgements
This research was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT & Future Planning (2016R1C1B2012933).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Rights and permissions
About this article
Cite this article
Rho, C.R., Kim, MJ. & Joo, CK. Influence of across-pupil sutures on the optical quality of intraocular lenses. Int Ophthalmol 38, 909–915 (2018). https://doi.org/10.1007/s10792-016-0441-y
Received:
Accepted:
Published:
Issue Date: