Abstract
Background/aims To study the observer-related variability of optic nerve head (ONH) measurements using confocal laser scanning tomography (HRT I) in a screening setting.
Methods Six experienced glaucoma specialists independently evaluated 50 ONH topographies from 25 adults using HRT software ver. 2.01 in a masked fashion. ONH topographies were obtained from a cohort study of 882 healthy adults and additionally included one patient with glaucomatous eyes. A glaucoma-screening-like setting was intended. The mean interobserver difference was defined as the mean percentual difference between an observer’s analysis and the mean of all six observers for all eyes and all observers. The interobserver range was calculated for each eye as the percentual difference between the lowest and highest measurement, with the highest measurement as denominator. Additionally, Kendall’s coefficient of rank concordance was assessed for the main HRT parameters.
Results Mean disc area ranged from 1.83 ± 0.49 to 2.21 ± 0.40 mm² (mean interobserver difference: 8.3%; interobserver range: 5–50%; rank concordance: 0.86). The lowest mean interobserver differences were found for mean retinal nerve fibre layer thickness (RNFLT; 6.5%), maximum cup depth (2.9%) and cup shape (6.8%). An increased interobserver range was significantly correlated to a low cup to disc area ratio (r = 0.64, P < 0.0001).
Conclusions The observer-dependent diagnostic variability of HRT measurements can lead to divergent diagnostic evaluation of the ONH in a screening setting. Any HRT software relying on a reference database is exposed to relevant observer-related variability of the disc area. For screening purposes, HRT measurements should be completed by other diagnostic methods to compensate for possible diagnostic uncertainty.
Similar content being viewed by others
References
Rohrschneider K, Burk RO, Kruse FE, Volcker HE (1994) Reproducibility of the optic nerve head topography with a new laser tomographic scanning device. Ophthalmology 101(6):1044–1049
Asawaphureekorn S, Zangwill L, Weinreb RN (1996) Ranked-segment distribution curve for interpretation of optic nerve topography. J Glaucoma 5(2):79–90
Bartz-Schmidt KU, Sengersdorf A, Esser P, Walter P, Hilgers RD, Krieglstein GK (1996) The cumulative normalised rim/disc area ratio curve. Graefes Arch Clin Exp Ophthalmol 234(4):227–231
Wollstein G, Garway-Heath D, Hitchings RA (1998) Identification of early glaucoma cases with the scanning laser ophthalmoscope. Ophthalmology 105(8):1557–1563
Hermann MM, Theofylaktopoulos I, Bangard N, Jonescu-Cuypers C, Coburger S, Diestelhorst M (2004) Optic nerve head morphometry in healthy adults using confocal laser scanning tomography. Br J Ophthalmol 88(6):761–765
Arenas-Archila E, Caycedo-Yunis F, Rodriguez MR (2001) Evaluation and definition of physiologic macro cups with confocal optic nerve analysis (HRT). Int Ophthalmol 23(4–6):239–244
Bartz-Schmidt KU, Jonescu-Cuypers CP, Thumann G, Frucht J, Krieglstein GK (1996) Effect of the contourline on the cup area using the Heidelberg retina tomograph. Klin Monatsbl Augenheilkd 209(5):292–297 German
Burk RO, Vihanninjoki K, Bartke T, Tuulonen A, Airaksinen PJ, Volcker HE, Konig JM (2000) Development of the standard reference plane for the Heidelberg retina tomograph. Graefes Arch Clin Exp Ophthalmol 238(5):375–384
Jonas JB, Schmidt AM, Muller-Bergh JA, Schlotzer-Schrehardt UM, Naumann GO (1992) Human optic nerve fiber count and optic disc size. Invest Ophthalmol Vis Sci 33(8):2012–2018
Papastathopoulos KI, Jonas JB, Panda-Jonas S (1995) Large optic discs in large eyes, small optic discs in small eyes. Exp Eye Res 60(4):459–461
Janknecht P, Funk J (1994) Optic nerve head analyser and Heidelberg retina tomograph: accuracy and reproducibility of topographic measurements in a model eye and in volunteers. Br J Ophthalmol 78(10):760–768
Kruse FE, Burk RO, Volcker HE, Zinser G, Harbarth U (1989) Reproducibility of topographic measurements of the optic nerve head with laser tomographic scanning. Ophthalmology 96(9):1320–1324
Orgul S, Cioffi GA, Bacon DR, van Buskirk EM (1996) Sources of variability of topometric data with a scanning laser ophthalmoscope. Arch Ophthalmol 114(2):161–164
Garway-Heath DF, Rudnicka AR, Lowe T, Foster PJ, Fitzke FW, Hitchings RA (1998) Measurement of optic disc size: equivalence of methods to correct for ocular magnification. Br J Ophthalmol 82(6):643–649
Wollstein G, Garway-Heath DF, Fontana L, Hitchings RA (2000) Identifying early glaucomatous changes. Comparison between expert clinical assessment of optic disc photographs and confocal scanning ophthalmoscopy. Ophthalmology 107(12):2272–2277
Boros AS, Jonescu-Cuypers CP, Bartz-Schmidt KU (2001) Variability of the normalised rim/disc area quotient estimated by laser scanning tomography. A comparison with conventional planimetry. Int Ophthalmol 24(5):263–267
Iester M, Mikelberg FS, Courtright P, Burk RO, Caprioli J, Jonas JB et al (2001) Interobserver variability of optic disk variables measured by confocal scanning laser tomography. Am J Ophthalmol 132(1):57–62
Jonescu-Cuypers CP, Thumann G, Hilgers RD, Bartz-Schmidt KU, Krott R, Krieglstein GK (1999) Long-term fluctuations of the normalised rim/disc area ratio quotient in normal eyes. Graefes Arch Clin Exp Ophthalmol 237(3):181–186
Chauhan BC, McCormick TA, Nicolela MT, LeBlanc RP (2001) Optic disc and visual field changes in a prospective longitudinal study of patients with glaucoma: comparison of scanning laser tomography with conventional perimetry and optic disc photography. Arch Ophthalmol 119(10):1492–1499
Garway-Heath DF, Poinoosawmy D, Wollstein G, Viswanathan A, Kamal D, Fontana L et al (1999) Inter- and intraobserver variation in the analysis of optic disc images: comparison of the Heidelberg retina tomograph and computer assisted planimetry. Br J Ophthalmol 83(6):664–669
Swindale NV, Stjepanovic G, Chin A, Mikelberg FS (2000) Automated analysis of normal and glaucomatous optic nerve head topography images. Invest Ophthalmol Vis Sci 41(7):1730–1742
Chrastek R, Wolf M, Donath K, Niemann H, Paulus D, Hothorn T et al (2005) Automated segmentation of the optic nerve head for diagnosis of glaucoma. Med Image Anal 9(4):297–314
Acknowledgements
This study has been presented in part as poster at the Association for Research in Vision and Ophthalmology (ARVO) 2002 in Fort Lauderdale, Fla., USA.
Author information
Authors and Affiliations
Corresponding author
Additional information
The study was presented in part at the ARVO meeting 2002.
Rights and permissions
About this article
Cite this article
Hermann, M.M., Garway-Heath, D.F., Jonescu-Cuypers, C.P. et al. Interobserver variability in confocal optic nerve analysis (HRT). Int Ophthalmol 26, 143–149 (2005). https://doi.org/10.1007/s10792-006-9022-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10792-006-9022-9