Abstract
Background and Aims
Every country has standardized reprocessing guidelines for reducing the risk of microorganism transmission via reusable biopsy forceps. Sterilization is performed either by autoclaving or with the use of ethylene oxide (EO) gas. However, there are no clear standard global recommendations. The aim of this study was to determine whether EO gas or autoclaving is a safer and more effective method for the sterilization of reusable forceps.
Methods
This was a prospective study conducted at multiple tertiary referral centers. Seventy reusable biopsy forceps that had been reused at least 20 times each were collected from six endoscopy centers. In all, 61 forceps from five centers were sterilized using EO gas, and the nine forceps from the remaining center were placed in an autoclave. We performed real-time polymerase chain reaction (RT-PCR) for Mycobacterium tuberculosis and hepatitis B virus and performed bacterial cultures on the reusable forceps, which were cut into 2- to 3-cm sections. The forceps were also scanned with an electron microscope (EM) to detect surface damage and contamination.
Results
Escherichia coli bacteria were cultured from 2 of the 61 (3.3%) reusable biopsy forceps sterilized with EO gas. On EM scanning, abundant debris and tissue materials remained on the cup surfaces of the reused biopsy forceps and on their inner wires. No microorganisms were found on the autoclaved forceps.
Conclusions
Sterilization with EO gas may be inadequate because the complicated structure of the forceps may interfere with sterilization. Therefore, for optimum safety, reusable biopsy forceps should be sterilized by autoclaving.
Similar content being viewed by others
References
Kimmey MB, Burnett DA, Carr-Locke DL, et al. Transmission of infection by gastrointestinal endoscopy. Gastrointest Endosc. 1993;39:885–888.
Spach DH, Silverstein FE, Stamm WE. Transmission of infection by gastrointestinal endoscopy and bronchoscopy. Ann Intern Med. 1993;118:117–128.
Beilenhoff U, Neumann CS, Rey JF, et al. ESGE-ESGENA Guideline: cleaning and disinfection in gastrointestinal endoscopy. Endoscopy. 2008;40:939–957. doi:10.1055/s-2008-1077722.
Jung M, Beilenhoff U, Pietsch M, Kraft B, Rippin G. Standardized reprocessing of reusable colonoscopy biopsy forceps is effective: results of a German multicenter study. Endoscopy. 2003;35:197–202. doi:10.1055/s-2003-37270.
Favero MS, Bond WW. Disinfection of medical and surgical materials. In: Block SS, ed. Disinfection, Sterilization, and Preservation. 5th ed. Philadelphia, PA: Lippincott Williams & Wilkins; 2001:881–917.
Kim Y-T. Introduction of proposed guideline for endoscopy disinfection. Korean J Gastrointest Endosc. 2010;40:49.
Wilkinson M, Simmons N, Bramble M, et al. Report of the Working Party of the Endoscopy Committee of the British Society of Gastroenterology on the reuse of endoscopic accessories. Gut. 1998;42:304–306.
Tandon RK. Disinfection of gastrointestinal endoscopes and accessories. J Gastroenterol Hepatol. 2000;15:G69–G72.
Systchenko R, Marchetti B, Canard JN, et al. Guidelines of the French Society of Digestive Endoscopy: recommendations for setting up cleaning and disinfection procedures in gastrointestinal endoscopy. Endoscopy. 2000;32:807–818. doi:10.1055/s-2000-7710.
Alvarado CJ, Reichelderfer M. APIC guideline for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control. Am J Infect Control. 2000;28:138–155. doi:S0196655300780515.
Nelson DB, Jarvis WR, Rutala WA, et al. Multi-society guideline for reprocessing flexible gastrointestinal endoscopes Society for Healthcare Epidemiology of America. Infect Control Hosp Epidemiol. 2003;24:532–537. doi:10.1086/502237.
Banerjee S, Shen B, Nelson DB, et al. Infection control during GI endoscopy. Gastrointest Endosc. 2008;67:781–790. doi:10.1016/j.gie.2008.01.027.
Ogoshi K. Reprocessing of gastrointestinal endoscopic accessories. J Gastroenterol Hepatol. 2000;15:G82–G85.
Lee RM, Kozarek RA, Sumida SE, Raltz SL. Risk of contamination of sterile biopsy forceps in disinfected endoscopes. Gastrointest Endosc. 1998;47:377–381. doi:S001651079800100X.
Kozarek RA, Attia FM, Sumida SE, et al. Reusable biopsy forceps: a prospective evaluation of cleaning, function, adequacy of tissue specimen, and durability. Gastrointest Endosc. 2001;53:747–750. doi:10.1067/mge.2001.114055.
Shim TS, Chi HS, Lee SD, et al. Adequately washed bronchoscope does not induce false-positive amplification tests on bronchial aspirates in the diagnosis of pulmonary tuberculosis. Chest. 2002;121:774–781.
Abe A, Inoue K, Tanaka T, et al. Quantitation of hepatitis B virus genomic DNA by real-time detection PCR. J Clin Microbiol. 1999;37:2899–2903.
O’Connor HJ, Axon AT. Gastrointestinal endoscopy: infection and disinfection. Gut. 1983;24:1067–1077.
Langenberg W, Rauws EA, Oudbier JH, Tytgat GN. Patient-to-patient transmission of Campylobacter pylori infection by fiberoptic gastroduodenoscopy and biopsy. J Infect Dis. 1990;161:507–511.
Dwyer DM, Klein EG, Istre GR, Robinson MG, Neumann DA, McCoy GA. Salmonella newport infections transmitted by fiberoptic colonoscopy. Gastrointest Endosc. 1987;33:84–87.
Bronowicki JP, Venard V, Botte C, et al. Patient-to-patient transmission of hepatitis C virus during colonoscopy. N Engl J Med. 1997;337:237–240. doi:10.1056/NEJM199707243370404.
Alfa MJ, Castillo J. Impact of FDA policy change on the reuse of single-use medical devices in Michigan hospitals. Am J Infect Control. 2004;32:337–341. doi:10.1016/j.ajic.2004.03.003.
Fireman Z. Biopsy forceps: reusable or disposable? J Gastroenterol Hepatol. 2006;21:1089–1092. doi:10.1111/j.1440-1746.2006.04429.x.
Hambrick D 3rd. Reprocessing of single-use endoscopic biopsy forceps and snares. One hospital’s study. Gastroenterol Nurs. 2001;24:112–115.
Merritt K, Hitchins VM, Brown SA. Safety and cleaning of medical materials and devices. J Biomed Mater Res. 2000;53:131–136. doi:10.1002/(SICI)1097-4636(2000)53:2<131:AID-JBM1>3.0.CO;2-I.
Deprez PH, Horsmans Y, Van Hassel M, Hoang P, Piessevaux H, Geubel A. Disposable versus reusable biopsy forceps: a prospective cost evaluation. Gastrointest Endosc. 2000;51:262–265. doi:S0016510700262603.
Rizzo J, Bernstein D, Gress F. A performance, safety and cost comparison of reusable and disposable endoscopic biopsy forceps: a prospective, randomized trial. Gastrointest Endosc. 2000;51:257–261. doi:S0016510700514291.
Wolfsen HC. Advantages of reusable accessories. Gastrointest Endosc Clin N Am. 2000;10:349–359.
Yang R, Ng S, Nichol M, Laine L. A cost and performance evaluation of disposable and reusable biopsy forceps in GI endoscopy. Gastrointest Endosc. 2000;51:266–270. doi:S0016510700499414.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Yoon, J.H., Yoon, B.C., Lee, H.L. et al. Comparison of Sterilization of Reusable Endoscopic Biopsy Forceps by Autoclaving and Ethylene Oxide Gas. Dig Dis Sci 57, 405–412 (2012). https://doi.org/10.1007/s10620-011-1884-7
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10620-011-1884-7