The aim of this study was to assess the feasibility, safety, and performance of a novel diagnostic catheter specifically designed for engaging both coronary arteries via the right transradial artery approach. A total of 160 patients were randomized between the standard Judkins (5F R4, L4; Cordis Corporation, Miami, FL) and the 5F Tiger II (Terumo Corporation, Tokyo, Japan) catheters. End points included the duration of various procedures and the assessments of angiographic image quality and catheter performance. The Tiger II was associated with a significantly shorter (40%) total procedure time (199.6±50.2 vs. 331.5±72.9 s, p=0.001) and a 33% shorter total fluoroscopic time (93.1±33.8 vs. 138.2±47.6 s, p=0.001) for diagnostic coronary angiography, compared with those with the Judkins catheter. There was no significant difference between the Tiger II and Judkins catheters for left coronary angiographic quality (left anterior descending, 2.82±0.48 vs. 2.94±0.29, p=0.084; left circumflex, 2.90±0.38 vs. 2.87±0.44, p=0.629). The Tiger II provided superior right coronary angiograms, compared with the Judkins catheter (2.99± 0.11 vs. 2.82±0.48, p=0.003). For the left coronary angiograms, the initial randomized catheter completed the procedure in 91% of the patients with the Tiger II and in 98% with the Judkins (p=0.167) catheters. For the right coronary angiograms, 100% were completed with the Tiger II and 95% with the Judkins (p=0.120) catheters. There were no angiographic or clinical complications in either group, so the procedural success rate was 100%. The potential of the Tiger II catheter for use as a multipurpose catheter for right transradial coronary angiography to reduce procedural and X-ray times to the level of classic transfemoral coronary angiography has to be confirmed in a randomized study.
Similar content being viewed by others
References
1. Campeau L, (1989) Percutaneous radial artery approach for coronary angiography. Cathet Cardiovasc Diagn 16:3–7
2. Kiemeneij F, Laarman GJ, (1993) Percutaneous transradial artery approach for coronary stent implantation. Cathet Cardiovasc Diagn 30:173–178
3. Kiemeneij F, Laarman GJ, (1995) Transradial artery Palmaz-Schatz coronary stent implantation: result of a single-center feasibility study. Am Heart J 130:14–21
4. Lotan C, Hasin Y, Mosseri M, Rozenman Y, Admon D, Nassar H, Gotsman MS, (1995) Transradial approach for coronary angiography and angioplasty. Am J Cardiol 76:164–167
5. Kiemeneij F, Hofland J, Laarman GJ, Hupkens D, Lubbe H, (1995) Cost comparison between two modes of Palmaz-Schatz coronary stent implantation: transradial bare stent technique vs transfemoral sheath-protected stent technique. Cathet Cardiovasc Diagn 35:301–308
6. D’Urbano M, Cafiero F, (1996) Percutaneous radial approach for coronary angiography. G Ital Cardiol 26:1149–1155
7. Kiemeneij F, Laarman GJ, Odekerken D, Slagboom T, Wieken R, (1997) A randomized comparison of percutaneous transluminal coronary angioplasty by the radial, brachial and femoral approaches: the access study. J Am Coll Cardiol 29:1269–1275
8. Hildick-Smith DJR, Lowe MD, Walsh JT, Ludman PF, Stephens NG, Schofield PM, Stone DL, Shapiro LM, Petch MC, (1998) Coronary angiography from the radial artery- experience, complications and limitations. Int J Cardiol 64:231–239
9. Saito S, Miyake S, Hosokawa G, Tanaka S, Kawamitsu K, Kaneda H, Ikei H, Shiono T, (1999) Transradial coronary intervention in Japanese patients. Cathet Cardiovasc Intervent 46:37–41
10. Louvard Y, Lefevre T, Allain A, Morice MC, (2001) Coronary Angiography Through the Radial or the Femoral Approach: the CARAFE Study. Cathet Cardiovasc Intervent 52:181–187
11. Ochiai M, Ikari Y, Yamaguchi T, Isshiki T, Koyama Y, Eto K, Takeshita S, Tamura T, Sato T (2000) New long-tip guiding catheters designed for right transradial coronary intervention. Cathet Cardiovasc Intervent 49:218–224
12. Ikari Y, Ochiai M, Hangaishi M, Ohno M, Taguchi J, Hara K, Isshiki T, Tamura T, Yamaguchi T, (1998) Novel guide catheter for left coronary intervention via a right upper limb approach. Cathet Cardiovasc Diagn 44:244–247
13. Goldberg SL, Renslo R, Sinow R, French WJ, (1998) Learning curve in the use of the radial artery as vascular access in the performance of percutaneous transluminal coronary angioplasty. Cathet Cardiovasc Diagn 44:147–152
14. Todd DM, Hubner PJ, Hudson N, Sarma J, McCance AJ, Caplin J, (2001) Multicentre, prospective, randomized trial of 4 vs 6 French catheters in 410 patients undergoing coronary angiography. Cathet Cardiovasc Intervent 54:269–275
15. Ootomo T, Meguro T, Endoh N, Terashima M, Ito Y, Abe S, Ogata K, Fujiwara S, Honda H, Kuhara R, Miyazaki Y, Kawashima O, Isoyama S, (2002) A new miniature catheter with side-holes for percutaneous transradial or transbrachial coronary angiography. J Invas Cardiol 14:379–84
16. Louvard Y, Benamer H, Garot P, Hildick-Smith D, Loubeyre C, Rigattieri S, Monchi M, Lefèvre T, Hamon M, (2004) Comparison of Transradial and Transfemoral Approaches for Coronary Angiography and Angioplasty in Octogenarians (the OCTOPLUS Study). Am J Cardiol 94:1177–1180
Acknowledgments
We acknowledge the help and support of the cardiac room nurses and technicians for the completion of this project. Particular thanks are due to Mr Fukamizu Junichi (Terumo Japan, Hong Kong branch) and Dae-Jae Im (Terumo Korea, Busan branch) for catheter preparation.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Kim, SM., Kim, DK., Kim, DI. et al. Novel diagnostic catheter specifically designed for both coronary arteries via the right transradial approach. Int J Cardiovasc Imaging 22, 295–303 (2006). https://doi.org/10.1007/s10554-005-9029-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10554-005-9029-8