DQE and eDQE were evaluated for a slot scan X-ray system using four standard radiation qualities: RQA3, RQA5, RQA7, and RQA9. The results showed that the system under consideration, KARS-BKS2 fluorograph, provided maximal DQE(u,v) and eDQE(u′,v′) in the case of radiation quality RQA5 and a small focal point. At null spatial frequencies DQE(0) was greater than 40% and eDQE(0) was 35%.
Similar content being viewed by others
References
Samei, E., Ranger, N. T., MacKenzie, A., Honey, I. D., Dobbins III, J. T., and Ravin, C. E., “Detector or system? Extending the concept of detective quantum efficiency to characterize the performance of digital radiographic imaging systems,” Radiology, 249, No. 3, 926-937 (2008).
“Assessment of digital radiographic detectors and systems using eDQE”, Carl E. Ravin Advanced Imaging Laboratories website: http:deckard.duhs.duke.edu/railabs/radiographicassessment.html (accessed May 25, 2019).
Zelikman, M. I., Kabanov, S. P., and Kruchinin, S. A., “Assessment of effect of radiation scattered in a patient’s body on parameters of digital X-ray image formation channel,” Biomed. Eng., 47, 228-234 (2014).
Samei, E., Ranger, N. T., MacKenzie, A., Honey, I. D., Dobbins III, J. T., and Ravin, C. E., “Effective DQE (eDQE) and speed of digital radiographic systems: An experimental methodology,” Med. Phys., 36, No. 8, 3806-3817 (2009).
Bertolini, M., Nitrosi, A., Rivetti, S., Lanconelli, N., Pattacini, P., Ginocchi, V., and Iori, M., “A comparison of digital radiography systems in terms of effective detective quantum efficiency,” Med. Phys., 39, No. 5, 2617-2627 (2012).
Ranger, N. T., Mackenzie, A., Honey, I. D., Dobbins III, J. T., Ravin, C. E., and Samei, E., “Extension of DQE to include scatter, grid, magnification, and focal spot blur: A new experimental technique and metric,” Proc. SPIE, 7258, 1-12 (2009).
Samei, E., Ranger, N. T., Dobbins, III, J. T., and Ravin, C. E., “Effective dose efficiency: An application-specific metric of quality and dose for digital radiography,” Phys. Med. Biol., 56, No. 16, 5099-5118 (2011).
Yalcin, A. and Olgar, T., “Characterizing the digital radiography system in terms of effective detective quantum efficiency and CDRAD measurement,” Nuclear Instr. Meth. Phys. Res., A 896, 113-121 (2018).
Alves, A. F. F., Alvarez, M., Ribeiro, S. M., Duarte, S. B., Miranda, J. R. A., and Pina, D. R., “Association between subjective evaluation and physical parameters for radiographic images optimization,” Phys. Med., 32, No. 1, 123-132 (2016).
Clavel, A. H., Monnin, P., Létang, J. M., Verdun, F. R., and Darbon, A., “Characterising the EOS slot-scanning system with the effective detective quantum efficiency,” Rad. Prot. Dos., 169, 319-324 (2016).
Kabanov, S. P., Studies and Development of Methods for Programmable Systems for Monitoring the Recording Quantum Efficiency of Digital X-Ray Image Receivers [in Russian], Dissertation for Master’s Degree in Technical Sciences (2011).
Bor, D., Guven, A., Yusuf, A. R., Birgul, O., Yuksel, S., Yalcin, A., Marshall, N., and Olgar, T., “A modified formulation of eDQE for digital radiographic imaging,” Rad. Phys. Chem., 156, 6-14 (2019).
Author information
Authors and Affiliations
Corresponding author
Additional information
Translated from Meditsinskaya Tekhnika, Vol. 55, No. 3, May-Jun., 2021, pp. 39-43.
Rights and permissions
About this article
Cite this article
Bekhterev, A.V., Kotkov, R.V., Labusov, V.A. et al. DQE and eDQE of the KARS-BKS2 Digital Slot Scan X-Ray System. Biomed Eng 55, 204–209 (2021). https://doi.org/10.1007/s10527-021-10103-0
Received:
Published:
Issue Date: