Existing techniques for sterilizing equipment using pressurized steam are discussed. Methods for confirming the sterility of equipment are considered. A method for determining sterility based on lethality criterion is proposed, along with a structural scheme of an information and measurement system for sterilization–in–place.
Similar content being viewed by others
References
State Standard GOST R ISO 17665–1–2016 (ISO/TS 17665–2:2016): Sterilization of Health Care Products. Moist Heat. Part 1. Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices [in Russian].
Agalloco, J., “Steam sterilization–in–place technology.” J. Parent. Sci. Technol., 44, No. 5, 253 (1990).
State Standard GOST R ISO 13408–5–2011: Aseptic Processing of Health Care Products — Part 5: Sterilization in Place [in Russian].
State Standard GOST R 56893–2016 (ISO/TS 17665–2:2009): Sterilization of Health Care Products. Moist Heat. Part 2. Guidance on the Application of ISO 17665–1 [in Russian].
Dion, M. and Parker, W., “Steam sterilization principles,” Pharmac. Eng., 33, No. 6, 3–5 (2013).
Flickinger, M. C., Downstream Industrial Biotechnology: Recovery and Purification, Wiley, North Carolina (2013).
Nordhauser, F. M., and Olson, W. P., Sterilization of Drugs and Devices. Technologies for the 21st Century, Taylor & Francis Group, London (2000), pp. 55–59.
Paul, R. S. and Heldman, D. R., Introduction to Food Engineering, Elsevier, Amsterdam (2009), pp. 435–440.
Slovesnov, N. A., Mirsaitov, S. F., and Svetlichnyi, V. V., “Development of an automated system for cleaning and sterilizing mobile pharmaceutical reactors,” in: Proc. Conf. Priorities in the XXI Century: Interdisciplinary Research in Our Times [in Russian] (2019), pp. 95–98.
Author information
Authors and Affiliations
Corresponding author
Additional information
Translated from Meditsinskaya Tekhnika, Vol. 55, No. 2, Mar.–Apr., 2021, pp. 49–52.
Rights and permissions
About this article
Cite this article
Slovesnov, N.A., Mirsaitov, S.F. & Svetlichnyi, V.V. Monitoring the Sterility of Aseptic Manufacturing Equipment. Biomed Eng 55, 145–148 (2021). https://doi.org/10.1007/s10527-021-10089-9
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10527-021-10089-9