Experimental Determination of the Maximum Permissible Coil Misalignment in Adaptive Transcutaneous Energy Transfer Systems

  • K. O. Gurov
  • E. A. Mindubaev
  • S. V. Selishchev
  • O. A. Surkov
Article
  • 9 Downloads

The effect of the symmetrical coil couple geometry on the value of lateral coil misalignment that corresponds to zero coupling in inductive energy transfer systems was studied. It was shown that the distance between coil turns and the number of coil turns have rather small effect on the zero-coupling distance. Increasing the outer diameter of the coils was shown to be the only way to increase the zero-coupling distance at given parameters of the system. It was found that changes in the outer diameter of the coils had more significant effect on the zero-coupling distance when the outer diameter was more than twice as great as the inner diameter.

Preview

Unable to display preview. Download preview PDF.

Unable to display preview. Download preview PDF.

References

  1. 1.
    Agarwal, K., Jegadeesan, R., Guo, Y.-X, and Thakor, N. “Wireless power transfer strategies for implantable bioelectronics: Methodological review,” IEEE Rev. Biomed. Eng., 10, 1-28 (2017).CrossRefGoogle Scholar
  2. 2.
    Danilov, A. A., Itkin, G. P., and Selishchev, S. V., “Progress in methods for transcutaneous wireless energy supply to implanted ventricular assist devices,” Biomed. Eng., 44, No. 4, 125-129 (2010).CrossRefGoogle Scholar
  3. 3.
    Zeng, F.-G., Rebscher, S., Harrison, W., Sun, X., and Feng, H., “Cochlear implants: System design, integration, and evaluation,” IEEE Rev. Biomed. Eng., 1, 115-142 (2008).CrossRefGoogle Scholar
  4. 4.
    Clark, G., Cochlear implants: Fundamentals and Applications, Springer (2003).Google Scholar
  5. 5.
    North, R., “Neural interface devices: Spinal cord stimulation technology,” Proc. IEEE, 96, 1108-1119 (2008).CrossRefGoogle Scholar
  6. 6.
    Eldridge, P., Simpson, B. A., and Gilbart, J., “The role of rechargeable systems in neuromodulation,” Eur. Neurol. Rev., 6, No. 3, 187-192 (2011).CrossRefGoogle Scholar
  7. 7.
    Pelletier, B., Spiliopoulos, S., Finocchiaro, T., Graef, F., et al., “System overview of the fully implantable destination therapy – ReinHeart-total artificial heart,” Eur. J. Card.-Thor. Surg., 47, No. 1, 80-86 (2015).CrossRefGoogle Scholar
  8. 8.
    Weiland, J. D. and Humayun, M. S., “Visual prosthesis,” Proc. IEEE, 96, No. 7, 1076-1084 (2008).CrossRefGoogle Scholar
  9. 9.
    Li, X., Yang, Y., and Gao, Y., “Visual prosthesis wireless energy transfer system optimal modeling,” Biomed. Eng. Online, 13, No. 3, 1-11 (2014).Google Scholar
  10. 10.
    Burny, F., Donkerwolcke, M., Moulart, F., et al., “Concept, design and fabrication of smart orthopedic implants,” Med. Eng. Phys., 22, 469 (2000).CrossRefGoogle Scholar
  11. 11.
    Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “Design and evaluation of an inductive powering unit for implantable medical devices using GPU computing,” Prog. Electromagn. Res. B, 69, 61-73 (2016).CrossRefGoogle Scholar
  12. 12.
    Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “Space-frequency approach to design of displacement tolerant transcutaneous energy transfer system,” Prog. Electromagn. Res. M, 44, 91-100 (2015).Google Scholar
  13. 13.
    Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “Methods for compensation of coil misalignment in systems for inductive transcutaneous power transfer to implanted medical devices,” Biomed. Eng., 51, No. 1, 56-60 (2017).CrossRefGoogle Scholar
  14. 14.
    Zierhofer, C. M. and Hochmair, E.S., “Geometric approach for coupling enhancement of magnetically coupled coils,” IEEE Trans. Biomed. Eng., 43, No. 7, 708-714 (1996).CrossRefGoogle Scholar
  15. 15.
    Danilov, A. A., Mindubaev, E. A., and Selishchev, S. V., “The effect of transmitter coil size on the optimal implantation depth of receiver coil in transcutaneous inductive energy transfer systems,” Biomed. Eng., 51, No. 5, 354-357 (2017).CrossRefGoogle Scholar

Copyright information

© Springer Science+Business Media, LLC, part of Springer Nature 2018

Authors and Affiliations

  • K. O. Gurov
    • 1
  • E. A. Mindubaev
    • 1
  • S. V. Selishchev
    • 1
  • O. A. Surkov
    • 1
  1. 1.National Research University of Electronic Technology (MIET)ZelenogradRussia

Personalised recommendations