Abstract
We present the first in-depth system integration study of in-plane hydrodynamic focusing in a microfluidic impedance cytometry lab-on-a-chip. The method relies on constricting the detection volume with non-conductive sheath flows and characterizing particles or cells based on changes in impedance. This approach represents an avenue of overcoming current limitations in sensitivity with translating cytometers to the point of care for rapid, low-cost blood analysis. While examples of integrated devices are present in the literature, no systematic study of the interplay between hydrodynamics and electrodynamics has been carried out as of yet. We develop analytical and numerical models to describe the impedimetric response of the sensor as a function of cellular characteristics, physical flow properties, and device geometry. We fabricate a working prototype lab-on-a-chip for experimental validation using latex particles. We find that ionic diffusion can be a critical limiting factor even at high Péclet number. Moreover, we explore geometric variations, revealing that the ionic diffusion-related distance between the center of the hydrodynamic focusing junction and the impedance measurement electrodes plays a dominant role. With our device, we demonstrate over fivefold enhancement in impedance signals and population separation with in-plane hydrodynamic focusing. It is only through such in-depth system studies, in both models and experiments, that optimal utilization of microsystem capabilities becomes possible.
Similar content being viewed by others
Abbreviations
- ⌀ (µm):
-
Cell diameter
- VA (µm):
-
Virtual aperture width
- Q s, Q f (µl/h):
-
Flow rates of sample and focus streams
- FR (1):
-
Flow ratio (sample flow over total flow)
- Z, Z empty (Ω):
-
Impedance, empty channel impedance
- |ΔZ| (%):
-
Relative change in impedance (impedance signal)
- Δ|ΔZ| (%):
-
Population separation
- δ (%):
-
Population spread
- w, w fc (µm):
-
Center (sample) and focus channel width
- h (µm):
-
Channel height
- l, g (µm):
-
Microelectrode length and gap
- d (µm):
-
Distance from focus junction to electrode
- med, el, DI, ion, cell, mem, cyt :
-
Subscripts denoting: medium, electrolyte component of medium, DI water component of medium, ions in the medium, cells in the medium, membrane component of cell, cytosol component of cell
- P, V (µm3):
-
Particle and electrical interaction volume
- Π (1):
-
Volume fraction of particle in electrical interaction volume
- σ (S/m):
-
Conductivity
- ε (F/m):
-
Permittivity
- C (1):
-
Normalized concentration
- IS (M):
-
Ionic strength
- λ D (µm):
-
Debye length
- C offset, C amp (1):
-
Concentration profile fit parameters
- s p, s s (µm):
-
Concentration profile fit parameters
- f, f s (Hz):
-
Signal and sampling frequency
- Pe, Re (1):
-
Péclet number, Reynolds number
- RMSE (%):
-
Root-mean-square error
References
Analog Devices (2005) AD5933 Impedance converter network analyzer. http://www.analog.com/en/products/rf-microwave/direct-digital-synthesis-modulators/ad5933.html
Asami K, Hanai T, Koizumi N (1980) Dielectric approach to suspensions of ellipsoidal particles covered with a shell in particular reference to biological cells. Jpn J Appl Phys 19:359–365. doi:10.1143/JJAP.19.359
Ateya DA, Erickson JS, Howell PB Jr et al (2008) The good, the bad, and the tiny: a review of microflow cytometry. Anal Bioanal Chem 391:1485–1498. doi:10.1007/s00216-007-1827-5
Bashir R (2004) BioMEMS: state-of-the-art in detection, opportunities and prospects. Adv Drug Deliv Rev 56:1565–1586. doi:10.1016/j.addr.2004.03.002
Ben-Yoav H, Dykstra PH, Gordonov T et al (2014) A microfluidic-based electrochemical biochip for label-free DNA hybridization analysis. J Vis Exp. doi:10.3791/51797
Berg HC (1993) Random walks in biology. Princeton University Press, Princeton
Bernabini C, Holmes D, Morgan H (2011) Micro-impedance cytometry for detection and analysis of micron-sized particles and bacteria. Lab Chip 11:407–412. doi:10.1039/c0lc00099j
Chen J, Xue C, Zhao Y et al (2015) Microfluidic impedance flow cytometry enabling high-throughput single-cell electrical property characterization. Int J Mol Sci 16:9804–9830. doi:10.3390/ijms16059804
Cheung K, Gawad S, Renaud P (2005) Impedance spectroscopy flow cytometry: on-chip label-free cell differentiation. Cytom A 65A:124–132. doi:10.1002/cyto.a.20141
Cheung KC, Di Berardino M, Schade-Kampmann G et al (2010) Microfluidic impedance-based flow cytometry. Cytometry A 77A:648–666. doi:10.1002/cyto.a.20910
DeBlois RW, Bean CP (1970) Counting and sizing of submicron particles by the resistive pulse technique. Rev Sci Instrum 41:909–916. doi:10.1063/1.1684724
Evander M, Ricco AJ, Morser J et al (2013) Microfluidic impedance cytometer for platelet analysis. Lab Chip 13:722–729. doi:10.1039/c2lc40896a
Frankowski M, Simon P, Bock N et al (2015) Simultaneous optical and impedance analysis of single cells: a comparison of two microfluidic sensors with sheath flow focusing. Eng Life Sci 15:286–296. doi:10.1002/elsc.201400078
Gawad S, Schild L, Renaud P (2001) Micromachined impedance spectroscopy flow cytometer for cell analysis and particle sizing. Lab Chip 1:76–82. doi:10.1039/b103933b
Gawad S, Cheung K, Seger U et al (2004) Dielectric spectroscopy in a micromachined flow cytometer: theoretical and practical considerations. Lab Chip 4:241–251. doi:10.1039/b313761a
Golden JP, Justin GA, Nasir M, Ligler FS (2012) Hydrodynamic focusing—a versatile tool. Anal Bioanal Chem 402:325–335. doi:10.1007/s00216-011-5415-3
Han X, van Berkel C, Gwyer J et al (2012) Microfluidic lysis of human blood for leukocyte analysis using single cell impedance cytometry. Anal Chem 84:1070–1075. doi:10.1021/ac202700x
Holmes D, Pettigrew D, Reccius CH et al (2009) Leukocyte analysis and differentiation using high speed microfluidic single cell impedance cytometry. Lab Chip 9:2881–2889. doi:10.1039/b910053a
Johnson AM, Sadoway DR, Cima MJ, Langer R (2005) Design and testing of an impedance-based sensor for monitoring drug delivery. J Electrochem Soc 152:H6–H11. doi:10.1149/1.1824045
Jung W, Han J, Choi J-W, Ahn CH (2015) Point-of-care testing (POCT) diagnostic systems using microfluidic lab-on-a-chip technologies. Microelectron Eng 132:46–57. doi:10.1016/j.mee.2014.09.024
Kottke-Marchant K, Davis B (2012) Laboratory hematology practice. Wiley, Chichester
Kunstmann-Olsen C, Hoyland JD, Rubahn H-G (2012) Influence of geometry on hydrodynamic focusing and long-range fluid behavior in PDMS microfluidic chips. Microfluid Nanofluid 12:795–803. doi:10.1007/s10404-011-0923-1
Lee G-B, Chang C-C, Huang S-B, Yang R-J (2006) The hydrodynamic focusing effect inside rectangular microchannels. J Micromech Microeng 16:1024–1032. doi:10.1088/0960-1317/16/5/020
Lide DR, Kehiaian HV (1994) CRC handbook of thermophysical and thermochemical data. CRC Press, Boca Raton
Mansor MA, Ahmad MR (2015) Single cell electrical characterization techniques. Int J Mol Sci 16:12686–12712. doi:10.3390/ijms160612686
Meyer MT, Subramanian S, Kim YW et al (2015) Multi-depth valved microfluidics for biofilm segmentation. J Micromech Microeng 25:95003. doi:10.1088/0960-1317/25/9/095003
Morgan H, Sun T, Holmes D et al (2007) Single cell dielectric spectroscopy. J Phys Appl Phys 40:61–70. doi:10.1088/0022-3727/40/1/S10
Nasir M, Price DT, Shriver-Lake LC, Ligler F (2010) Effect of diffusion on impedance measurements in a hydrodynamic flow focusing sensor. Lab Chip 10:2787–2795. doi:10.1039/C005257D
Nasir M, Mott DR, Kennedy MJ et al (2011) Parameters affecting the shape of a hydrodynamically focused stream. Microfluid Nanofluid 11:119–128. doi:10.1007/s10404-011-0778-5
Nguyen J, Wei Y, Zheng Y et al (2015) On-chip sample preparation for complete blood count from raw blood. Lab Chip 15:1533–1544. doi:10.1039/C4LC01251H
Nieuwenhuis JH, Kohl F, Bastemeijer J et al (2004) Integrated Coulter counter based on 2-dimensional liquid aperture control. Sens Actuat B Chem 102:44–50. doi:10.1016/j.snb.2003.10.017
Pitts E, Tabor BE (1970) Concentration dependence of electrolyte conductance. Part 2—comparison of experimental data with the Fuoss-Onsager and Pitts treatments. Trans Faraday Soc 66:693–707. doi:10.1039/TF9706600693
Rodriguez-Trujillo R, Mills CA, Samitier J, Gomila G (2007) Low cost micro-Coulter counter with hydrodynamic focusing. Microfluid Nanofluid 3:171–176. doi:10.1007/s10404-006-0113-8
Rodriguez-Trujillo R, Castillo-Fernandez O, Garrido M et al (2008) High-speed particle detection in a micro-Coulter counter with two-dimensional adjustable aperture. Biosens Bioelectron 24:290–296. doi:10.1016/j.bios.2008.04.005
Shapiro HM (2003) Practical flow cytometry. Wiley, Hoboken
Siggaard-Andersen O, Durst RA, Maas AHJ (1984) Physicochemical quantities and units in clinical chemistry with special emphasis on activities and activity coefficients (Recommendations 1983). Pure Appl Chem 56:567–594. doi:10.1351/pac198456050567
Sun T, Morgan H (2010) Single-cell microfluidic impedance cytometry: a review. Microfluid Nanofluid 8:423–443. doi:10.1007/s10404-010-0580-9
Sun T, Gawad S, Green NG, Morgan H (2007a) Dielectric spectroscopy of single cells: time domain analysis using Maxwell’s mixture equation. J Phys Appl Phys 40:1–8. doi:10.1088/0022-3727/40/1/S01
Sun T, Green NG, Gawad S, Morgan H (2007b) Analytical electric field and sensitivity analysis for two microfluidic impedance cytometer designs. IET Nanobiotechnol 1:69–79. doi:10.1049/iet-nbt:20070019
Sun T, Green NG, Morgan H (2008) Analytical and numerical modeling methods for impedance analysis of single cells on-chip. Nano 3:55–63. doi:10.1142/S1793292008000800
Sun T, Bernabini C, Morgan H (2010) Single-colloidal particle impedance spectroscopy: complete equivalent circuit analysis of polyelectrolyte microcapsules. Langmuir 26:3821–3828. doi:10.1021/la903609u
van Berkel C, Gwyer JD, Deane S et al (2011) Integrated systems for rapid point of care (PoC) blood cell analysis. Lab Chip 11:1249–1255. doi:10.1039/C0LC00587H
Watkins N, Venkatesan BM, Toner M et al (2009) A robust electrical microcytometer with 3-dimensional hydrofocusing. Lab Chip 9:3177–3184. doi:10.1039/B912214A
Xuan X, Zhu J, Church C (2010) Particle focusing in microfluidic devices. Microfluid Nanofluid 9:1–16. doi:10.1007/s10404-010-0602-7
Acknowledgments
The authors would like to thank the Robert W. Deutsch Foundation and the Maryland Innovation Initiative for financial support. The authors also appreciate the support of the Maryland NanoCenter and its FabLab. The authors wish to thank Robert Dietrich for useful discussions.
Author information
Authors and Affiliations
Corresponding author
Additional information
This research was performed while H. Ben-Yoav was at the MEMS Sensors and Actuators Laboratory (MSAL), Institute for Systems Research, Department of Electrical and Computer Engineering, University of Maryland, College Park, MD 20742, USA.
Electronic supplementary material
Below is the link to the electronic supplementary material.
Rights and permissions
About this article
Cite this article
Winkler, T.E., Ben-Yoav, H. & Ghodssi, R. Hydrodynamic focusing for microfluidic impedance cytometry: a system integration study. Microfluid Nanofluid 20, 134 (2016). https://doi.org/10.1007/s10404-016-1798-y
Received:
Accepted:
Published:
DOI: https://doi.org/10.1007/s10404-016-1798-y