Abstract
Purpose
To compare rebound tonometer and cannulation as methods for measuring intraocular pressure (IOP) in rats.
Methods
The accuracy of the TonoLab rebound tonometer was determined in eight cannulated rat eyes. IOP was manipulated by changing air pressure from 20 to 100 mmHg at 10-mmHg intervals, and the IOP was measured with the rebound tonometer at each level. The average value of three repeated pressure readings was recorded. Correlation analysis and comparison with the Bland and Altman method were performed. The intraclass correlation coefficient was calculated to assess intraoperator variability.
Results
The IOP values measured with the TonoLab rebound tonometer were well correlated with the actual IOP (r 2 = 0.963, P = 0.01). The mean of the difference between the rebound tonometer and actual (cannulation) IOP was 7.41 ± 7.87%. The intraclass correlation coefficient was 0.9, indicating low intraoperator variability.
Conclusions
The rebound tonometer showed high accuracy and reliability for IOP measurement in rat eyes.
References
Kontiola AI, Goldblum D, Mittag T, Danias J. The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res 2001;73:781–785.
Wang WH, Millar JC, Pang IH, Wax MB, Clark AF. Noninvasive measurement of rodent intraocular pressure with a rebound tonometer. Invest Ophthalmol Vis Sci 2005;46:4617–4621.
Joo CK, Choi JS, Ko HW, et al. Necrosis and apoptosis after retinal ischemia: involvement of NMDA-mediated excitotoxicity and p53. Invest Ophthalmol Vis Sci 1999;40:713–720.
Danias J, Kontiola AI, Filippopoulos T, Mittag T. Method for the noninvasive measurement of intraocular pressure in mice. Invest Ophthalmol Vis Sci 2003;44:1138–1141.
Aihara M, Lindsey JD, Weinreb RN. Reduction of intraocular pressure in mouse eyes treated with latanoprost. Invest Ophthalmol Vis Sci 2002;43:146–150.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Lee, E.J., Park, K.H., Kim, D.M. et al. Assessing intraocular pressure by rebound tonometer in rats with an air-filled anterior chamber. Jpn J Ophthalmol 52, 500–503 (2008). https://doi.org/10.1007/s10384-008-0591-1
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10384-008-0591-1