Assessing intraocular pressure by rebound tonometer in rats with an air-filled anterior chamber

  • Brief Communication
  • Published:
Japanese Journal of Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To compare rebound tonometer and cannulation as methods for measuring intraocular pressure (IOP) in rats.

Methods

The accuracy of the TonoLab rebound tonometer was determined in eight cannulated rat eyes. IOP was manipulated by changing air pressure from 20 to 100 mmHg at 10-mmHg intervals, and the IOP was measured with the rebound tonometer at each level. The average value of three repeated pressure readings was recorded. Correlation analysis and comparison with the Bland and Altman method were performed. The intraclass correlation coefficient was calculated to assess intraoperator variability.

Results

The IOP values measured with the TonoLab rebound tonometer were well correlated with the actual IOP (r 2 = 0.963, P = 0.01). The mean of the difference between the rebound tonometer and actual (cannulation) IOP was 7.41 ± 7.87%. The intraclass correlation coefficient was 0.9, indicating low intraoperator variability.

Conclusions

The rebound tonometer showed high accuracy and reliability for IOP measurement in rat eyes.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

References

  1. Kontiola AI, Goldblum D, Mittag T, Danias J. The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp Eye Res 2001;73:781–785.

    Article  CAS  PubMed  Google Scholar 

  2. Wang WH, Millar JC, Pang IH, Wax MB, Clark AF. Noninvasive measurement of rodent intraocular pressure with a rebound tonometer. Invest Ophthalmol Vis Sci 2005;46:4617–4621.

    Article  PubMed  Google Scholar 

  3. Joo CK, Choi JS, Ko HW, et al. Necrosis and apoptosis after retinal ischemia: involvement of NMDA-mediated excitotoxicity and p53. Invest Ophthalmol Vis Sci 1999;40:713–720.

    CAS  PubMed  Google Scholar 

  4. Danias J, Kontiola AI, Filippopoulos T, Mittag T. Method for the noninvasive measurement of intraocular pressure in mice. Invest Ophthalmol Vis Sci 2003;44:1138–1141.

    Article  PubMed  Google Scholar 

  5. Aihara M, Lindsey JD, Weinreb RN. Reduction of intraocular pressure in mouse eyes treated with latanoprost. Invest Ophthalmol Vis Sci 2002;43:146–150.

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ki Ho Park.

About this article

Cite this article

Lee, E.J., Park, K.H., Kim, D.M. et al. Assessing intraocular pressure by rebound tonometer in rats with an air-filled anterior chamber. Jpn J Ophthalmol 52, 500–503 (2008). https://doi.org/10.1007/s10384-008-0591-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-008-0591-1

Key Words

Navigation