Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery

  • Clinical Investigation
  • Published:
Japanese Journal of Ophthalmology Aims and scope Submit manuscript

Abstract

Purpose

To compare the accuracy of postoperative pachymetry between Orbscan II (Bausch & Lomb) scanning-slit corneal topography/pachymetry and the Pentacam (Oculus) rotating Scheimpflug camera.

Methods

Central corneal thickness (CCT) was determined in 24 patients (48 eyes) before and after laser in situ keratomileusis (LASIK) or Epipolis LASIK (Epi-LASIK) procedures. All eyes were examined by Orbscan II and Pentacam prior to refractive surgery and at the first, fourth, and twelfth week postoperatively. The residual CCT (RCCT) measured by each instrument was compared to the theoretical RCCT.

Results

On the first, fourth, and twelfth week after the refractive surgery, the mean RCCT measurements by Orbscan II were 413 ± 72, 435 ± 65, and 440 ± 69 µm, respectively. Those of Pentacam were 434 ± 51, 436 ± 53, and 438 ± 50 µm, respectively. Orbscan II measurements at the postoperative first and fourth week were significantly smaller than the theoretical RCCT (P < 0.01, P < 0.01, paired t test). There was no statistical difference between the theoretical RCCT and the Pentacam measurements at any stage.

Conclusion

The Orbscan II measurement values of postoperative corneas tended to be thinner than the theoretical values, but not those of the Pentacam.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Amano S, Honda N, Amano Y, et al. Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography. Ophthalmology 2006;113:937–941.

    Article  PubMed  Google Scholar 

  2. Wei RH, Lim L, Chan WK, Tan DT. Evaluation of Orbscan II corneal topography in individuals with myopia. Ophthalmology 2006;113:177–183.

    Article  PubMed  Google Scholar 

  3. Lee JY, Kim JH, Kim HM, Song JS. Comparison of anterior chamber depth measurement between Orbscan IIz and ultrasound biomicroscopy. J Refract Surg 2007;23:487–491.

    PubMed  Google Scholar 

  4. O’Donnell C, Maldonado-Codina C. Agreement and repeatability of central thickness measurement in normal corneas using ultrasound pachymetry and the OCULUS Pentacam. Cornea 2005;24:920–924.

    Article  PubMed  Google Scholar 

  5. Nemeth G, Vajas A, Kolozsvari B, Berta A, Modis L Jr. Anterior chamber depth measurements in phakic and pseudophakic eyes: Pentacam versus ultrasound device. J Cataract Refract Surg 2006;32:1331–1335.

    Article  PubMed  Google Scholar 

  6. Buehl W, Stojanac D, Sacu S, Drexler W, Findl O. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol 2006;141:7–12.

    Article  PubMed  Google Scholar 

  7. Konstantopoulos A, Hossain P, Anderson DF. Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis? Br J Ophthalmol 2007;91:551–557.

    Article  PubMed  PubMed Central  Google Scholar 

  8. Barkana Y, Gerber Y, Elbaz U, et al. Central corneal thickness measurement with the Pentacam Scheimpflug system, optical lowcoherence reflectometry pachymeter, and ultrasound pachymetry. J Cataract Refract Surg 2005;31:1729–1735.

    Article  PubMed  Google Scholar 

  9. Lackner B, Schmidinger G, Pieh S, Funovics MA, Skorpik C. Repeatability and reproducibility of central corneal thickness measurement with Pentacam, Orbscan, and ultrasound. Optom Vis Sci 2005;82:892–899.

    Article  PubMed  Google Scholar 

  10. Suzuki S, Oshika T, Oki K, et al. Corneal thickness measurements: scanning-slit corneal topography and noncontact specular microscopy versus ultrasonic pachymetry. J Cataract Refract Surg 2003;29:1313–1318.

    Article  PubMed  Google Scholar 

  11. Tkachov SI, Lautenschlager C, Ehrich D, Struck HG. Changes in the lens epithelium with respect to cataractogenesis: light microscopic and Scheimpflug densitometric analysis of the cataractous and the clear lens of diabetics and non-diabetics. Graefes Arch Clin Exp Ophthalmol 2006;244:596–602.

    Article  PubMed  Google Scholar 

  12. Matsuda J, Hieda O, Kinoshita S. Quantification of corneal opacity after refractive corneal surgery using the anterior segment analyzer (in Japanese). Nippon Ganka Gakkai Zasshi (J Jpn Ophthalmol Soc) 2007;111:447–453.

    Google Scholar 

  13. Fujioka M, Nakamura M, Tatsumi Y, Kusuhara A, Maeda H, Negi A. Comparison of Pentacam Scheimpflug camera with ultrasound pachymetry and noncontact specular microscopy in measuring central corneal thickness. Curr Eye Res 2007;32:89–94.

    Article  PubMed  Google Scholar 

  14. Kawana K, Tokunaga T, Miyata K, Okamoto F, Kiuchi T, Oshika T. Comparison of corneal thickness measurements using Orbscan II, non-contact specular microscopy, and ultrasonic pachymetry in eyes after laser in situ keratomileusis. Br J Ophthalmol 2004;88:466–468.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  15. Prisant O, Calderon N, Chastang P, Gatinel D, Hoang-Xuan T. Reliability of pachymetric measurements using Orbscan after excimer refractive surgery. Ophthalmology 2003;110:511–515.

    Article  PubMed  Google Scholar 

  16. Boscia F, La Tegola MG, Alessio G, Sborgia C. Accuracy of Orbscan optical pachymetry in corneas with haze. J Cataract Refract Surg 2002;28:253–258.

    Article  PubMed  Google Scholar 

  17. Iskander NG, Anderson Penno E, Peters NT, Gimbel HV, Ferensowicz M. Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures. J Cataract Refract Surg 2001;27:681–685.

    Article  CAS  PubMed  Google Scholar 

  18. Ho T, Cheng AC, Rao SK, Lau S, Leung CK, Lam DS. Central corneal thickness measurements using Orbscan II, Visante, ultrasound, and Pentacam pachymetry after laser in situ keratomileusis for myopia. J Cataract Refract Surg 2007;33:1177–1182.

    Article  PubMed  Google Scholar 

  19. Wang Z, Chen J, Yang B. Posterior corneal surface topographic changes after laser in situ keratomileusis are related to residual corneal bed thickness. Ophthalmology 1999;106:406–409; discussion 409–410.

    Article  CAS  PubMed  Google Scholar 

  20. Liu Z, Huang AJ, Pflugfelder SC. Evaluation of corneal thickness and topography in normal eyes using the Orbscan corneal topography system. Br J Ophthalmol 1999;83:774–778.

    Article  CAS  PubMed  PubMed Central  Google Scholar 

  21. Seitz B, Torres F, Langenbucher A, Behrens A, Suarez E. Posterior corneal curvature changes after myopic laser in situ keratomileusis. Ophthalmology 2001;108:666–672; Discussion 673.

    Article  CAS  PubMed  Google Scholar 

  22. Seiler T, Koufala K, Richter G. Iatrogenic keratectasia after laser in situ keratomileusis. J Refract Surg 1998;14:312–317.

    CAS  PubMed  Google Scholar 

  23. Kamiya K, Miyata K, Tokunaga T, Kiuchi T, Hiraoka T, Oshika T. Structural analysis of the cornea using scanning-slit corneal topography in eyes undergoing excimer laser refractive surgery. Cornea 2004;23:S59–64.

    Article  PubMed  Google Scholar 

  24. Fakhry MA, Artola A, Belda JI, Ayala MJ, Alio JL. Comparison of corneal pachymetry using ultrasound and Orbscan II. J Cataract Refract Surg 2002;28:248–252.

    Article  PubMed  Google Scholar 

  25. Tomidokoro A, Oshika T, Amano S, Higaki S, Maeda N, Miyata K. Changes in anterior and posterior corneal curvatures in keratoconus. Ophthalmology 2000;107:1328–1332.

    Article  CAS  PubMed  Google Scholar 

  26. Tanabe T, Oshika T, Tomidokoro A, et al. Standardized colorcoded scales for anterior and posterior elevation maps of scanning slit corneal topography. Ophthalmology 2002;109:1298–1302.

    Article  PubMed  Google Scholar 

  27. Miyata K, Tokunaga T, Nakahara M, et al. Residual bed thickness and corneal forward shift after laser in situ keratomileusis. J Cataract Refract Surg 2004;30:1067–1072.

    Article  PubMed  Google Scholar 

  28. Baek T, Lee K, Kagaya F, Tomidokoro A, Amano S, Oshika T. Factors affecting the forward shift of posterior corneal surface after laser in situ keratomileusis. Ophthalmology 2001;108:317–320.

    Article  CAS  PubMed  Google Scholar 

  29. Yoshida T, Miyata K, Tokunaga T, Tanabe T, Oshika T. Difference map or single elevation map in the evaluation of corneal forward shift after LASIK. Ophthalmology 2003;110:1926–1930.

    Article  PubMed  Google Scholar 

  30. Hashemi H, Mehravaran S. Corneal changes after laser refractive surgery for myopia: comparison of Orbscan II and Pentacam findings. J Cataract Refract Surg 2007;33:841–847.

    Article  PubMed  Google Scholar 

  31. Nawa Y, Masuda K, Ueda T, Hara Y, Uozato H. Evaluation of apparent ectasia of the posterior surface of the cornea after keratorefractive surgery. J Cataract Refract Surg 2005;31:571–573.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Jumpei Matsuda.

About this article

Cite this article

Matsuda, J., Hieda, O. & Kinoshita, S. Comparison of central corneal thickness measurements by Orbscan II and Pentacam after corneal refractive surgery. Jpn J Ophthalmol 52, 245–249 (2008). https://doi.org/10.1007/s10384-008-0550-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10384-008-0550-x

Key Words

Navigation