Studying Systematic Errors on Estimation Decision, Detection, and Quantification Limit on Micro-TLC

  • Original
  • Published:
Chromatographia Aims and scope Submit manuscript

Abstract

The decision limit (CCα), capability of detection (CCβ) and quantification limit (QL) are importance performance characteristics in method validation. The TLC-Scanner 3 from Camag provides the possibility to choose the slit dimension of light to determine the peak chromatogram of a substance. The influence of the slit dimension for determination of CCα, CCβ and QL of paracetamol has been carried out. Paracetamol was spotted onto plates of AL-TLC Si G 60 F254 by linomat 4 in the range of 50–400 ng/spot and 10–400 ng/band, then on twin chambers eluted with TAEA (toluene:acetone-ethanol:conc.ammonia, 45 + 45 + 7 + 3 v/v) for 45 mm. Eluted spots were scanned in different slit dimensions at 248 nm. The CCα, CCβ and QL of paracetamol were estimated through the linear regression (LRM) and signal-to-noise (S/N) methods. Slit lengths between 50 and 133 % of the band width of the spots, and with the noise factor of the slit under 2.6, produced good precision measurements of TLC-densitometry between plates, while slit lengths between 50 and 83 % of the band width of the spots introduced a higher sensitivity response of the detector. The estimated CCα, CCβ and QL were determined by how the data were collected, the analytical optical setting, and the usage method for the estimation of both validation parameters.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Dolan JW (2006) The role of the signal-to-noise ratio in precision and accuration. How is the detection limit determined? LCGC EUROPE, 19 (1). http://www.chromatographyonline.com/lcgc/Column. Accessed: 20 May 2012

  2. European Commission (2002) Commission Decision (EC) no 657/2002 of 12 August 2002 implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results. Off J Eur Commun L 221/8 (17 August 2002)

  3. Funk W, Dammann V, Donnevert G (2007) Quality assurance in analytical chemistry. Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim

    Google Scholar 

  4. Hahn-Deinstrop E (2006) Applied thin-layer chromatography best practice and avoidance of mistakes, 2nd revised and enlarged edn. Wiley-VCH, Weinheim

  5. Harvey D (2000) Modern analytical chemistry. McGraw Hill, Toronto

    Google Scholar 

  6. Indrayanto G (2011) In: Srivastava MM (ed) High-performace thin-layer chromatography (HPTLC). Springer, Germany

    Google Scholar 

  7. Indrayanto G, Yuwono M, Suciati (2009) TLC: validation of analyses. In: Cazes J (ed) Encyclopedia of Chromatography, vol 1, 3rd edn. Marcel Dekker, New York, pp 2336–2339

  8. United Nations (2009) Guidance for the validation of analytical methodology and calibration of equipment used for testing of illicit drugs in seized materials and biological specimens: a commitment to quality and continuous improvement. United Nations, New York

    Google Scholar 

  9. Verdon E, Hurtaud-Pessel D, Sanders P (2006) Evaluation of the limit of performance of an analytical method based on a statistical calculation of its critical concentrations according to ISO standard 11843: application to routine control of banned veterinary drug residues in food according to European Decision 657/2002/EC. Accred Qual Assur 11:58–62

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to I Made Agus Gelgel Wirasuta.

Additional information

Published in the topical collection Miniaturized and New Featured Planar Chromatography and Related Techniques with guest editor Paweł K. Zarzycki.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Wirasuta, I.M.A.G., Dewi, N.M.A.R., Cahyadi, K.D. et al. Studying Systematic Errors on Estimation Decision, Detection, and Quantification Limit on Micro-TLC. Chromatographia 76, 1261–1269 (2013). https://doi.org/10.1007/s10337-012-2376-4

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s10337-012-2376-4

Keywords

Navigation