Abstract
The decision limit (CCα), capability of detection (CCβ) and quantification limit (QL) are importance performance characteristics in method validation. The TLC-Scanner 3 from Camag provides the possibility to choose the slit dimension of light to determine the peak chromatogram of a substance. The influence of the slit dimension for determination of CCα, CCβ and QL of paracetamol has been carried out. Paracetamol was spotted onto plates of AL-TLC Si G 60 F254 by linomat 4 in the range of 50–400 ng/spot and 10–400 ng/band, then on twin chambers eluted with TAEA (toluene:acetone-ethanol:conc.ammonia, 45 + 45 + 7 + 3 v/v) for 45 mm. Eluted spots were scanned in different slit dimensions at 248 nm. The CCα, CCβ and QL of paracetamol were estimated through the linear regression (LRM) and signal-to-noise (S/N) methods. Slit lengths between 50 and 133 % of the band width of the spots, and with the noise factor of the slit under 2.6, produced good precision measurements of TLC-densitometry between plates, while slit lengths between 50 and 83 % of the band width of the spots introduced a higher sensitivity response of the detector. The estimated CCα, CCβ and QL were determined by how the data were collected, the analytical optical setting, and the usage method for the estimation of both validation parameters.
Similar content being viewed by others
References
Dolan JW (2006) The role of the signal-to-noise ratio in precision and accuration. How is the detection limit determined? LCGC EUROPE, 19 (1). http://www.chromatographyonline.com/lcgc/Column. Accessed: 20 May 2012
European Commission (2002) Commission Decision (EC) no 657/2002 of 12 August 2002 implementing Council Directive 96/23/EC concerning the performance of analytical methods and the interpretation of results. Off J Eur Commun L 221/8 (17 August 2002)
Funk W, Dammann V, Donnevert G (2007) Quality assurance in analytical chemistry. Wiley-VCH Verlag GmbH & Co. KGaA, Weinheim
Hahn-Deinstrop E (2006) Applied thin-layer chromatography best practice and avoidance of mistakes, 2nd revised and enlarged edn. Wiley-VCH, Weinheim
Harvey D (2000) Modern analytical chemistry. McGraw Hill, Toronto
Indrayanto G (2011) In: Srivastava MM (ed) High-performace thin-layer chromatography (HPTLC). Springer, Germany
Indrayanto G, Yuwono M, Suciati (2009) TLC: validation of analyses. In: Cazes J (ed) Encyclopedia of Chromatography, vol 1, 3rd edn. Marcel Dekker, New York, pp 2336–2339
United Nations (2009) Guidance for the validation of analytical methodology and calibration of equipment used for testing of illicit drugs in seized materials and biological specimens: a commitment to quality and continuous improvement. United Nations, New York
Verdon E, Hurtaud-Pessel D, Sanders P (2006) Evaluation of the limit of performance of an analytical method based on a statistical calculation of its critical concentrations according to ISO standard 11843: application to routine control of banned veterinary drug residues in food according to European Decision 657/2002/EC. Accred Qual Assur 11:58–62
Author information
Authors and Affiliations
Corresponding author
Additional information
Published in the topical collection Miniaturized and New Featured Planar Chromatography and Related Techniques with guest editor Paweł K. Zarzycki.
Rights and permissions
About this article
Cite this article
Wirasuta, I.M.A.G., Dewi, N.M.A.R., Cahyadi, K.D. et al. Studying Systematic Errors on Estimation Decision, Detection, and Quantification Limit on Micro-TLC. Chromatographia 76, 1261–1269 (2013). https://doi.org/10.1007/s10337-012-2376-4
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s10337-012-2376-4