A Pilot Study on the Development of Remote Quality Control of Digital Mammography Systems in the NHS Breast Screening Programme

  • P. LooneyEmail author
  • M. D. Halling-Brown
  • J. M. Oduko
  • K. C. Young
Article

Abstract

In the UK, physicists and radiographers perform routine quality control (QC) of digital mammography equipment at daily, weekly and monthly intervals. The tests performed and tolerances are specified by standard protocols. The manual nature of many of the tests introduces variability due to the positioning of regions of interest (ROIs) and can be time consuming. The tools on workstations provided by manufacturers limit the range of analysis that radiographers can perform and do not allow for a standard set of tools and analysis because they are specific to a given manufacturer. Automated software provides a means of reducing the variability in the analysis and also provides the possibility of additional, more complex analysis than is currently performed on the daily, weekly and monthly checks by radiographers. To this end, a set of tools has been developed to analyse the routine images taken by radiographers. As well as automatically reproducing the usual measurements by radiographers more complex analysis is provided. A QC image collection system has been developed which automatically routes QC data from a clinical site to a centralised server for analysis. A Web-based interface has been created that allows the users to view the performance of the mammographic equipment. The pilot system obtained over 3000 QC images from seven X-ray units at a single screening centre over 2 years. The results show that these tools and methods of analysis can highlight changes in a detector over time that may otherwise go unnoticed with the conventional analysis.

Keywords

QC Mammography NHSBSP ImageJ QA 

Notes

Acknowledgments

This work has been funded by the NHS Breast Screening Programme. We would like to acknowledge the help of Prof. David Dance, the cooperation of the radiographers at the Jarvis Screening Centre and the staff at the Regional Radiation Protection Service at the Royal Surrey County Hospital.

References

  1. 1.
    Reiner B I: Automating quality assurance for digital radiography. J Am Coll Radiol 6(7):486–490,2009Google Scholar
  2. 2.
    Moores B, Charnock P, Ward M: Web-based tools for quality assurance and radiation protection in diagnostic radiology. Radiat Prot Dosim 139(1–3):422–429,2010Google Scholar
  3. 3.
    Grimes J, Leng S, McCollough C: Th-c-18a-07: A software tool for automated analysis of ct quality assurance phantoms. Med Phys 41(6):558–558,2014Google Scholar
  4. 4.
    Baxter G, Jones V, Milnes V, Oduko J, Philips V, Sellars S, Zoe V: Routine quality control tests for full field digital mammography systems: NHSBSP Report (1303),2013Google Scholar
  5. 5.
    Workman A, Castellano IA, Kulama E, Lawinski C, Marshall N, Young KC: Commissioning and routine testing of full field digital mammography systems: NHSBSP Report (0604), 2009Google Scholar
  6. 6.
    Jacobs J, Lemmens K, Nens J, Michielsen K, Marchal G, Bosmans H: One year of experience with remote quality assurance of digital mammography systems in the Flemish breast cancer screening program. In: Digital Mammography. Springer, 2008, pp 703–710Google Scholar
  7. 7.
    Bloomquist A, Jacobs J, Yaffe M: Su-e-i-87: Pilot testing of software for automated remote quality control of digital mammography equipment for use in the ontario breast screening program. Med Phys 38(6):3415–3416,2011Google Scholar
  8. 8.
    Young KC, Van Engen R, Bosmans H, Jacobs J, Zanca F: Quaxlity control in digital mammography. In: Digital Mammography. Springer, 2010, pp 33–54Google Scholar
  9. 9.
    Young K, Oduko J: Technical evaluation of Hologic Selenia Dimensions 2D Digital Breast Imaging System with software version 1.4.2: NHSBSP Report (1201), 2012Google Scholar

Copyright information

© Society for Imaging Informatics in Medicine 2015

Authors and Affiliations

  • P. Looney
    • 1
    Email author
  • M. D. Halling-Brown
    • 2
  • J. M. Oduko
    • 1
  • K. C. Young
    • 1
    • 3
  1. 1.NCCPMRoyal Surrey County HospitalGuildfordUK
  2. 2.Scientific Computing SectionRoyal Surrey County HospitalGuildfordUK
  3. 3.Department of PhysicsUniversity of SurreyGuildfordUK

Personalised recommendations