Abstract
In orthopaedics, conventional implant plates such as locking compression plate (LCP) made from non-biodegradable materials play a vital role in the fixation to support bone fractures, but also create a complication such as stress shielding. These again require a painful surgery to remove/replace after they have healed as it does not degrade into the physiological environment (PE). Currently, there has already been enough discovery of biodegradable materials that, despite being mechanically inefficient compared to non-biodegradable materials, can completely be biodegraded in PE during and after healing to avoid such problems. While there has been insufficient research on the design of biodegradable implant plates, the implementation of which may help achieve the goal with an effort of high mechanical strength. A novel design of biodegradable embossed locking compression plate (BELCP) is designed for biodegradable materials to approach superior mechanical performance and complete degradation over time, considering all such parameters and factors. For biomechanical evaluation, four-point bending test (4PBT), axial compressive and tensile test (ACTT) and torsion test (TT) have been performed on LCP, BELCP and its continuously degraded forms made of biodegradable material (Mg-alloy) using finite element method. BELCP has found 50%, 100% and 100% higher mechanical performance and safer in 4PBT, ACTT and TT, respectively, than LCP. Moreover, BELCP has also observed safe during continuous degradation up to 6 months after implantation under these three tests, considering an approximate sustained degradation rate of about 4 mm/year. Even Mg-alloy made BELCP can be sufficient and safer to support fractured bone than SS-alloy made LCP, but not Ti-alloy made LCP. BELCP can be a successful biodegradable bone implant plate after human/animal trials in the future.
Similar content being viewed by others
References
Abd Aziz AU, Abdul Wahab AH, Abdul Rahim RA, Abdul Kadir MR, Ramlee MH (2020) A finite element study: Finding the best configuration between unilateral, hybrid, and ilizarov in terms of biomechanical point of view. Injury 51(11):2474–2478. https://doi.org/10.1016/j.injury.2020.08.001
Abdul Wahab AH, Wui NB, Abdul Kadir MR, Ramlee MH (2020) Biomechanical evaluation of three different configurations of external fixators for treating distal third tibia fracture: finite element analysis in axial, bending and torsion load. Comput Biol Med 127:104062. https://doi.org/10.1016/j.compbiomed.2020.104062
Arnone JC, El-Gizawy AS, Crist BD, Rocca GJD, Ward CV (2013) Computer-aided engineering approach for parametric investigation of locked plating systems design. J Med Devices Trans ASME 7(2):021001. https://doi.org/10.1115/1.4024644
ASTM F382–17 (2017) Standard specification and test method for metallic bone plates. ASTM International, West Conshohocken
Bel JC (2019) Pitfalls and limits of locking plates. Orthop Traumatol Surg Res 105(1):S103–S109. https://doi.org/10.1016/j.otsr.2018.04.031
Biswas JK et al (2019) Design factors of lumbar pedicle screws under bending load: a finite element analysis. Biocybern Biomed Eng 39(1):52–62. https://doi.org/10.1016/j.bbe.2018.10.003
Bui HP, Tomar S, Courtecuisse H, Cotin S, Bordas SPA (2018) Real-time error control for surgical simulation. IEEE Trans Biomed Eng 65(3):596–607. https://doi.org/10.1109/TBME.2017.2695587
Burkhart TA, Andrews DM, Dunning CE (2013) Finite element modeling mesh quality, energy balance and validation methods: a review with recommendations associated with the modeling of bone tissue. J Biomech 46(9):1477–1488. https://doi.org/10.1016/j.jbiomech.2013.03.022
Chandra G, Pandey A (2020) Preparation strategies for Mg-alloys for biodegradable orthopaedic implants and other biomedical applications: a review. Irbm. https://doi.org/10.1016/j.irbm.2020.06.003
Chandra G, Pandey A (2021a) Design and analysis of biodegradable buttress threaded screws for fracture fixation in orthopedics: a finite element analysis. Biomed Phys Eng Express 7(4):45010. https://doi.org/10.1088/2057-1976/ac00d1
Chandra G, Pandey A (2021b) Design approaches and challenges for biodegradable bone implants: a review. Expert Rev Med Devices 18(7):629–648. https://doi.org/10.1080/17434440.2021.1935875
Chandra G, Pandey A (2022) Effectiveness of laddered embossed structure in a locking compression plate for biodegradable orthopaedic implants. J Biomater Appl 36(7):1213–1230. https://doi.org/10.1177/08853282211058945
Chandra G, Pandey A (2020) Biodegradable bone implants in orthopedic applications: a review. Biocybern Biomed Eng 40(2):596–610. https://doi.org/10.1016/j.bbe.2020.02.003
Chandra G, Pandey A, Tipan N (2021) Longitudinally centered embossed structure in the locking compression plate for biodegradable bone implant plate: a finite element analysis. Comput Methods Biomech Biomed Engin 25(6):603–618. https://doi.org/10.1080/10255842.2021.1970145
Chandra G, Pandey A, Pandey S (2020) Design of a biodegradable plate for femoral shaft fracture fixation. Med Eng Phys 81:86–96. https://doi.org/10.1016/j.medengphy.2020.05.010
Chung CY (2018) A simplified application (APP) for the parametric design of screw-plate fixation of bone fractures. J Mech Behav Biomed Mater 77:642–648. https://doi.org/10.1016/j.jmbbm.2017.10.025
Cronier P, Pietu G, Dujardin C, Bigorre N, Ducellier F, Gerard R (2010) The concept of locking plates. Orthop Traumatol Surg Res. https://doi.org/10.1016/j.otsr.2010.03.008
Cui Z, Li W, Cheng L, Gong D, Cheng W, Wang W (2019) Effect of nano-HA content on the mechanical properties, degradation and biocompatible behavior of Mg-Zn/HA composite prepared by spark plasma sintering. Mater Charact 151(April):620–631. https://doi.org/10.1016/j.matchar.2019.03.048
Fan J et al (2013) Microstructure, mechanical properties, in vitro degradation and cytotoxicity evaluations of Mg-1.5Y-1.2Zn-0.44Zr alloys for biodegradable metallic implants. Mater Sci Eng C 33(4):2345–2352. https://doi.org/10.1016/j.msec.2013.01.063
Fouda N, Mostafa R, Saker A (2019) Numerical study of stress shielding reduction at fractured bone using metallic and composite bone-plate models. Ain Shams Eng J. https://doi.org/10.1016/J.ASEJ.2018.12.005
Ganadhiepan G et al (2019) The effects of dynamic loading on bone fracture healing under ilizarov circular fixators. J Biomech Eng 141(5):051005. https://doi.org/10.1115/1.4043037
Ghimire S, Miramini S, Richardson M, Mendis P, Zhang L (2018a) Role of dynamic loading on early stage of bone fracture healing. Ann Biomed Eng 46(11):1768–1784. https://doi.org/10.1007/s10439-018-2083-x
Ghimire S, Miramini S, Richardson M, Mendis P, Zhang L (2019) Effects of dynamic loading on fracture healing under different locking compression plate configurations: a finite element study. J Mech Behav Biomed Mater 94:74–85. https://doi.org/10.1016/j.jmbbm.2019.03.004
Hou L et al (2016) Microstructure, mechanical properties, corrosion behavior and biocompatibility of As-extruded biodegradable Mg–3Sn–1Zn–0.5Mn alloy. J Mater Sci Technol 32(9):874–882. https://doi.org/10.1016/j.jmst.2016.07.004
Jamil W, Allami M, Choudhury MZ, Mann C, Bagga T, Roberts A (2008) Do orthopaedic surgeons need a policy on the removal of metalwork? A descriptive national survey of practicing surgeons in the United Kingdom. Injury 39(3):362–367. https://doi.org/10.1016/j.injury.2007.10.028
Koç M, Tahmasebifar A, Evis Z, Sezer N, Kayhan SM (2018) Review of magnesium-based biomaterials and their applications. J Magnes Alloy 6(1):23–43. https://doi.org/10.1016/j.jma.2018.02.003
Li Z et al (2017) The synergistic effect of trace Sr and Zr on the microstructure and properties of a biodegradable Mg-Zn-Zr-Sr alloy. J Alloys Compd 702:290–302. https://doi.org/10.1016/j.jallcom.2017.01.178
Li J et al (2020a) Materials evolution of bone plates for internal fixation of bone fractures: a review. J Mater Sci Technol 36:190–208. https://doi.org/10.1016/j.jmst.2019.07.024
Li C et al (2020b) Design of biodegradable, implantable devices towards clinical translation. Nat Rev Mater 5(1):61–81. https://doi.org/10.1038/s41578-019-0150-z
Mehta CH, Narayan R, Nayak UY (2019) Computational modeling for formulation design. Drug Discov Today 24(3):781–788. https://doi.org/10.1016/j.drudis.2018.11.018
Miller DL, Goswami T (2007) A review of locking compression plate biomechanics and their advantages as internal fixators in fracture healing. Clin Biomech 22(10):1049–1062. https://doi.org/10.1016/j.clinbiomech.2007.08.004
Radha R, Sreekanth D (2017) Insight of magnesium alloys and composites for orthopedic implant applications – a review. J Magnes Alloys 5(3):286–312. https://doi.org/10.1016/j.jma.2017.08.003
Rho J, Kuhn-Spearing L, Zioupos P (1998) Mechanical properties and the hierarchical structure of bone. Med Eng Phys 20(2):92–102. https://doi.org/10.1016/S1350-4533(98)00007-1
Rosa N, Tavares SMO, Carbas RJC, Simoes R, Magalhães FD, Marques AT (2020) Methodology for bone–implant stiffness evaluation. Exp Mech. https://doi.org/10.1007/s11340-020-00654-w
Rouhi G, Hamedani MA (2012) A brief introduction into orthopaedic implants. Res Chapter, pp. 1–19
Ruggiero A, D’Amato R, Affatato S (2019) Comparison of meshing strategies in THR finite element modelling. Materials (basel) 12(14):1–11. https://doi.org/10.3390/ma12142332
Sanchez AHM, Luthringer BJC, Feyerabend F, Willumeit R (2015) Mg and Mg alloys: how comparable are in vitro and in vivo corrosion rates? A review. Acta Biomater 13(3):16–31. https://doi.org/10.1016/j.actbio.2014.11.048
Slone RM, Heare MM, Vander Griend RA, Montgomery WJ (1991) Orthopedic fixation devices. Radiographics 11(5):823–847. https://doi.org/10.1148/radiographics.11.5.1947319
Song J, She J, Chen D, Pan F (2020) Latest research advances on magnesium and magnesium alloys worldwide. J Magnes Alloy 8(1):1–41. https://doi.org/10.1016/j.jma.2020.02.003
Stahel PF, Alfonso NA, Henderson C, Baldini T (2017) Introducing the ‘Bone-Screw-Fastener’ for improved screw fixation in orthopedic surgery: a revolutionary paradigm shift? Patient Saf Surg 11(1):1–8. https://doi.org/10.1186/s13037-017-0121-5
Steiner M, Claes L, Ignatius A, Simon U, Wehner T (2014) Disadvantages of interfragmentary shear on fracture healing - mechanical insights through numerical simulation. J Orthop Res 32(7):865–872. https://doi.org/10.1002/jor.22617
Subasi O, Oral A, Lazoglu I (2019) A novel adjustable locking plate (ALP) for segmental bone fracture treatment. Injury 50(10):1612–1619. https://doi.org/10.1016/j.injury.2019.08.034
Thapa N, Prayson M, Goswami T (2015) A failure study of a locking compression plate implant. Case Stud Eng Fail Anal 3:68–72. https://doi.org/10.1016/j.csefa.2015.03.004
Trojanová Z, Drozd Z, Lukáč P (2018) Superplastic behaviour of selected magnesium alloys. Magnes Alloy Sel Issue. https://doi.org/10.5772/intechopen.79752
Venezuela J, Dargusch MS (2019) The influence of alloying and fabrication techniques on the mechanical properties, biodegradability and biocompatibility of zinc: a comprehensive review. Acta Biomater 87:1–40. https://doi.org/10.1016/j.actbio.2019.01.035
Wang M, Yang N (2017) A review of bioregulatory and coupled mechanobioregulatory mathematical models for secondary fracture healing. Med Eng Phys 48:90–102. https://doi.org/10.1016/j.medengphy.2017.06.031
Wisanuyotin T, Sirichativapee W, Paholpak P, Kosuwan W, Kasai Y (2020) Optimal configuration of a dual locking plate for femoral allograft or recycled autograft bone fixation: a finite element and biomechanical analysis. Clin Biomech 80:105156. https://doi.org/10.1016/j.clinbiomech.2020.105156
Witte F (2015) Reprint of: the history of biodegradable magnesium implants: a review. Acta Biomater 23:S28–S40. https://doi.org/10.1016/j.actbio.2015.07.017
Yan Y et al (2017) Effects of Zn concentration and heat treatment on the microstructure, mechanical properties and corrosion behavior of as-extruded Mg-Zn alloys produced by powder metallurgy. J Alloys Compd 693:1277–1289. https://doi.org/10.1016/j.jallcom.2016.10.017
Yang Y et al (2020) Mg bone implant: Features, developments and perspectives. Mater Design 185:108259. https://doi.org/10.1016/j.matdes.2019.108259
Yun Y et al (2009) Revolutionizing biodegradable metals. Mater Today 12(10):22–32. https://doi.org/10.1016/S1369-7021(09)70273-1
Zeng W, Liu Y, Hou X (2020) Biomechanical evaluation of internal fixation implants for femoral neck fractures: a comparative finite element analysis. Comput Methods Programs Biomed 196:105714. https://doi.org/10.1016/j.cmpb.2020.105714
Zhang L, Richardson M, Mendis P (2012) Role of chemical and mechanical stimuli in mediating bone fracture healing. Clin Exp Pharmacol Physiol 39(8):706–710. https://doi.org/10.1111/j.1440-1681.2011.05652.x
Zheng YF, Gu XN, Witte F (2012) Biodegradable metals. Mater Sci Eng R 77:1–34. https://doi.org/10.1007/978-1-4614-3942-4_5
Funding
None.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
No conflicts to declare in any way.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Chandra, G., Pandey, A. Biomechanical evaluation on a novel design of biodegradable embossed locking compression plate for orthopaedic applications using finite element analysis. Biomech Model Mechanobiol 21, 1371–1392 (2022). https://doi.org/10.1007/s10237-022-01596-z
Received:
Accepted:
Published:
Issue Date: