Continuous monitoring of blood pressure by analyzing the blood flow sound of arteriovenous fistula in hemodialysis patients

  • Yuka Kamijo
  • Eiichiro Kanda
  • Hayato Horiuchi
  • Noriyuki Kounoue
  • Keisuke Ono
  • Keizo Maeda
  • Akane Yanai
  • Kazuya Honda
  • Ryuji Tsujimoto
  • Mai Yanagi
  • Yoshitaka Ishibashi
  • Masayuki Yoshida
Original article

Abstract

Background

Patients with end-stage renal disease undergoing hemodialysis (HD) have an elevated risk of cardiovascular disease-related morbidity and mortality. To prevent from such a life-threatening event, the continuous blood pressure (BP) monitoring system may contribute to detect BP decline in early stages and may help to do appropriate disposal. Our research team has introduced an electronic stethoscope (Asahi Kasei Co, Ltd., Tokyo, Japan), which translates sound intensity of Arteriovenous Fistula (AVF) to BP data using the technique of Fourier transformation that can predict continuous BP non-invasively. This study, we investigated whether electronic stethoscope-guided estimated BP (e-BP) would actually reflect systolic BP measured by sphygmomanometer (s-BP), and whether e-BP could predict fall of BP during HD.

Methods

Twenty-six patients who underwent HD treatment in our hospital were evaluated prospectively. We obtained sound intensity data from the electronic stethoscope which was equipped with the return line of HD. Then, the data were translated into e-BP data to be compared with s-BP. Correlation of total of 315 data sets obtained from each method was examined. An accuracy of diagnosis of intra-dialytic hypotension (IDH) was evaluated.

Results

Total of 315 data sets were obtained. A close correlation was observed between e-BP and s-BP (r = 0.887, p < 0.0001). Sensitivity and positive predictive value of predicted-BP for detection of IDH was 90 and 81.3%, respectively.

Conclusions

Electronic stethoscope-guided BP measurement would be helpful for real-time diagnosis of BP fall in HD patients. Further investigations are needed.

Keywords

Arteriovenous fistula Intra-dialytic hypotension Blood pressure monitoring Electro stethoscope Fourier transformation 

Notes

Acknowledgements

The authors gratefully acknowledge the work of past and present members of our department of our hospital. We also wish to thank the timely technically help given by Elekio, Asahi Kasei Co., Ltd., Tokyo, Japan.

Compliance with ethical standards

Conflicts of interest

The authors declare no conflict of interest associated with this manuscript.

Human and animal rights (with IRB approval number)

All procedures performed in studies involving human participants were in accordance with the ethical standards of our institutional research committee at which the studies were conducted (IRB approval number 380) and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.

Informed consent

Informed consent was obtained from all individual participants included in the study.

References

  1. 1.
    Masakane I, Nakai S, Ogata S, Kimata N, Hanafusa N, Hamano T, et al. An overview of regular dialysis treatment in Japan (as of 31 December 2013). Ther Apher Dial. 2015;19:540–74.CrossRefPubMedGoogle Scholar
  2. 2.
    Flythe JE, Kimmel SE, Brunelli SM. Rapid fluid removal during dialysis is associated with cardiovascular morbidity and mortality. Kidney Int. 2011;79:250–7.CrossRefPubMedGoogle Scholar
  3. 3.
    KDOQI: K/doqi clinical practice guidelines for cardiovascular disease in dialysis patients. Am J Kidney Dis. 2005; 45:S1–S153.Google Scholar
  4. 4.
    Heinrich WL. Hemodynamic instability during hemodialysis. Kidney Int. 1986;30:605–12.CrossRefGoogle Scholar
  5. 5.
    Bos WJ, Bruin S, van Olden RW, Keur I, Wesseling KH, Westerhof N, et al. Cardiac and hemodynamic effects of hemodialysis and ultrafiltration. Am J Kidney Dis. 2000;35:819–26.CrossRefPubMedGoogle Scholar
  6. 6.
    Palmer BF, Henrich WL. Recent advances in the prevention and management of intradialytic hypotension. J Am Soc Nephrol. 2008;19:8–11.CrossRefPubMedGoogle Scholar
  7. 7.
    Stefánsson BV, Brunelli SM, Cabrera C, Rosenbaum D, Anum E, Ramakrishnan K, et al. Intradialytic hypotension and risk of cardiovascular disease. Clin J Am Soc Nephrol. 2014;9:2124–32.CrossRefPubMedPubMedCentralGoogle Scholar
  8. 8.
    Peng RC, Yan WR, Zhang NL, Lin WH, Zhou XL, Zhang YT. Cuffless and continuous blood pressure estimation from the heart sound signals. Sensors (Basel). 2015;15:23653–66.CrossRefPubMedPubMedCentralGoogle Scholar
  9. 9.
    Owen PJ, Priestman WS, Sigrist MK, Lambie SH, John SG, Chesterton LJ, et al. Myocardial contractile function and intradialytic hypotension. Hemodial Int. 2009;13:293–300.CrossRefPubMedGoogle Scholar
  10. 10.
    Mizumasa T, Hirakata H, Yoshimitsu T, Hirakata E, Kubo M, Kashiwagi M, et al. Dialysis-related hypotension as a cause of progressive frontal lobe atrophy in chronic hemodialysis patients: a 3-year prospective study. Nephron Clin Pract. 2004;97:c23–30.CrossRefPubMedGoogle Scholar
  11. 11.
    Kersh ES, Kronfield SJ, Unger A, Popper RW, Cantor S, Cohn K. Autonomic insufficiency in uremia as a cause of hemodialysis-induced hypotension. N Engl J Med. 1974;290:650–3.CrossRefPubMedGoogle Scholar
  12. 12.
    Poldermans D, Man in ‘t Veld AJ, Rambaldi R, Van Den Meiracker AH, Van Den Dorpel MA, Rocchi G, Boersma E, et al. Cardiac evaluation in hypotension-prone and hypotension-resistant hemodialysis patients. Kidney Int. 1999;56:1905–11.CrossRefPubMedGoogle Scholar
  13. 13.
    Kooman JP, Gladziwa U, Böcker G, van Bortel LM, van Hooff JP, Leunissen KM. System in hemodynamics during ultrafiltration and bicarbonate dialysis. Kidney Int. 1992;42:718–26.CrossRefPubMedGoogle Scholar
  14. 14.
    Song JH, Park GH, Lee SY, Lee SW, Lee SW, Kim MJ. Effect of sodium balance and the combination of ultrafiltration profile during sodium profiling hemodialysis on the maintenance of the quality of dialysis and sodium and fluid balances. J Am Soc Nephrol. 2005;16:237–46.CrossRefPubMedGoogle Scholar
  15. 15.
    Depner TA. Assessing adequacy of hemodialysis: urea modeling. Kidney Int. 1994;45:1522–35.CrossRefPubMedGoogle Scholar
  16. 16.
    Sego S. Crit line: instrument to aid fluid removal during hemodialysis. Anna J. 1997;24:347–8.PubMedGoogle Scholar
  17. 17.
    Hinchliffe RJ, Kirk B, Bhattacharjee D, Roe S, Jeffcoate W, Game F. The effect of haemodialysis on transcutaneous oxygen tension in patients with diabetes-a pilot study. Nephrol Dial Transplant. 2006;21:1981–3.CrossRefPubMedGoogle Scholar
  18. 18.
    Agishi T, Sakauchi M, Hoshino T, Akamatsu M, Sato T, Kawashima N. Shift in tissue oxygen pressure possibly preceding abrupt blood pressure drop. Jpn J Clin Dial. 2006;22:573–8.Google Scholar
  19. 19.
    Munguia M M, Vasquez P, Mattsson E, Mandersson B. Acoustical detection of venous stenosis in hemodialysis patients using principal component analysis. Conf Proc IEEE Eng Med Biol Soc. 2010;3654–7.Google Scholar
  20. 20.
    Bartels A, Harder D. Noninvasive determination of systolic blood-pressure by heart-sound pattern-analysis. Clin Phys Physiol Meas. 1992;13:249–56.CrossRefPubMedGoogle Scholar

Copyright information

© Japanese Society of Nephrology 2017

Authors and Affiliations

  • Yuka Kamijo
    • 1
    • 2
  • Eiichiro Kanda
    • 2
    • 3
  • Hayato Horiuchi
    • 4
  • Noriyuki Kounoue
    • 1
  • Keisuke Ono
    • 1
  • Keizo Maeda
    • 1
  • Akane Yanai
    • 1
  • Kazuya Honda
    • 1
  • Ryuji Tsujimoto
    • 1
  • Mai Yanagi
    • 1
  • Yoshitaka Ishibashi
    • 1
  • Masayuki Yoshida
    • 2
  1. 1.Department of NephrologyJapanese Red Cross Medical CenterTokyoJapan
  2. 2.Department of Life Science and Bioethics, Graduate School of MedicineTokyo Medical and Dental UniversityTokyoJapan
  3. 3.Department of NephrologyTokyo Kyosai HospitalTokyoJapan
  4. 4.Division of Medical Engineering TechnologyJapanese Red Cross Medical CenterTokyoJapan

Personalised recommendations