Introduction

Portable chest radiographs (CXRs) are crucial for in-patient imaging in hospitals [1]. The ongoing SARS-CoV-2 virus (coronavirus-19) pandemic has increased the use of portable radiography for acute care in some centers [2]. Provided images are acquired with acceptable image quality (IQ), performing portable CXRs at the patient's bedside has several advantages, including increased workflow efficiency [3] and patient acceptability [4], reducing the risk associated with transporting critically unwell or immobile patients [5], and limiting infection transmission risk by minimizing the frequency of interactions between infectious patients and staff [6, 7].

As the result of photon-electron interactions, Compton scattered photons are redirected on a trajectory misaligned to the primary beam. While the probability of Compton scattering interactions in tissue remains relatively constant over the diagnostic energy range, the likelihood of photoelectric interactions decreases with photon energy, resulting in a relatively greater chance of scatter interactions [8]. The overwhelming majority of photons produced by general radiography X-ray tubes during portable chest X-rays are high energy [8], resulting in a scattered to primary ratio (SPR) of between 2 and 4; meaning 2–4 times as much scatter can reach the detector as primary beam when imaging through the thorax [9,10,11]. The SPR increases with increased tube voltage, greater patient thickness, and larger X-ray field sizes, all of which are routinely found in the chest, abdominal, and spinal X-ray imaging [12]. As the pixel values (PV) within an image comprises both primary and scattered photons [12], scattered radiation degrades contrast within an image by dispersing the photons over the image receptor, generating PVs which do not represent the level of absorption the primary beam has undergone.

Several prospective scatter reduction methods exist to combat the IQ degradation caused by scattered radiation, including decreasing tube voltage, collimating tightly to the region of interest, or implementing the “air-gap” technique or slit-scanning technologies. However, in most situations, these methods are practically difficult to implement in portable CXRs and are rarely used. More commonly, an anti-scatter grid is introduced between the patient and the image receptor, which attenuates scattered photons based on the photon’s angle of incidence with the grid [13]. In this process, some primary beam photons are also attenuated by the interstitial or septal material [12]. Other drawbacks of anti-scatter grids include “grid cutoff” or artefacts resulting from stationary grids [14,15,16]. Imaging protocols involving a grid usually require the exposure parameters to be increased to compensate for the reduced number of photons reaching the detector [8]. The utilization of a grid in a CXR typically requires the dose to be increased by a factor of 2–3 [17]. Unsurprisingly, anti-scattered grid use for portable CXRs is inconsistent among sites [15, 18].

To overcome these practical constraints, several scatter correction post-processing algorithms (SCA) have been developed for use with portable radiography units, intending to reduce the effects of scattered radiation on IQ, including the Philips “Skyflow” [19,20,21,22,23], Fuji “Virtual Grid Software” [24,25,26], and Shimadzu SCA [27]. The detailed technical information in the Shimadzu SCA “White-Paper” is limited. However, the algorithm first requires a grid-less patient image acquired at a reference tube voltage (100 kVp). The algorithm then applies the predefined reconstruction kernel that estimates the scatter level within a grid-less patient image by comparing the pixel values within the patient image to an inbuilt object absent image acquired with the same exposure parameters. The estimated scatter contribution to the object absent image is then removed from the patient image, leaving the patient image where the scatter has been digitally suppressed [27].

Both “Skyflow” and “Virtual-Grid” software have been shown to improve the IQ in phantom and clinical images [19,20,21,22,23,24,25,26]. However, no studies assessing the efficacy of the Shimadzu SCA have been published. This study aimed to evaluate the efficacy of the Shimadzu SCA on imaging quality in phantom imaging and clinical CXRs.

Methods

Ethics

Local institutional low-risk ethics has been obtained (RES-21–0000-365Q).

Portable radiography unit

The portable X-rays were performed with the portable radiography unit Shimadzu Mobile Dart MX8, installed with proprietary SCA software (Shimadzu Corporation, Kyoto, Japan).

Objective image quality

Objective IQ was assessed using simulated radiographic acquisitions of a Kyoto Whole Body Adult PB-60 Phantom. The phantom’s thorax contains several anatomically realistic organs, including the lungs with pulmonary vessels, trachea, heart, and liver with hepatic and portal veins. The thorax portion of the phantom was used to simulate an average-sized patient’s portable CXR. Erect CXRs are most performed at our institution. Subsequently, the exposure conditions used in this study replicated local hospital protocols for erect portable AP CXRs performed on average-sized patients (100 kVp, and 180 cm source to image distance (SID)). The phantom was placed in the supine position, parallel to the floor, to allow for easier phantom alignment and created a replicable setup. During the first portion of acquired CXRs, two 1.5 cm × 1.5 cm × 1 mm of high purity (> 99.9%) aluminum attenuators were placed on the anterior surface of the phantoms chest to simulate low-contrast pathologies present within the chest. Due to its low effective atomic number, aluminum undergoes proportionally more Compton scattering interactions when compared to other attenuators making aluminum ideal to assess changes in the presence of scatter within a reconstructed image. The reason for placing the aluminum attenuators on the anterior surface of the phantom’s thorax was to generate the largest disbursement of scatter across the detector. The two attenuators were placed anterior to the lungs and heart regions over regions of uniform thickness and composition (Fig. 1a). The attenuator anterior to the lung was used to assess the algorithm’s ability to enhance high contrast details, while the attenuator anterior to the heart was used to evaluate the algorithm’s ability to enhance low-contrast details (Fig. 1b).

Fig. 1
figure 1

a Placement of the aluminum attenuators on the surface of the PBU-60 phantom (white vertical lines are reflection of shiny surface of the phantom). b Appearance on radiographic acquisition (arrow 1 showing the low-contrast attenuator over the phantom’s right heart border, arrow 2 showing the high contrast attenuator over the phantom’s left lung)

Patients of this phantom’s size would typically undergo examinations at approximately 1.4–1.8 mAs at our institution. As such, images were acquired at three mAs settings (1 mAs, 1.4 mAs, and 1.8 mAs). The aluminum attenuators were then removed, and the acquisitions were repeated with the same exposure increments. Each attenuator and tube current combination was repeated three times (18 total acquisitions). Images acquired with the same exposure (mAs) and placement of aluminum attenuators were combined and constituted a “set.”

The acquired raw images were each reconstructed with different SCA strengths enabled. The algorithm’s strength refers to the degree of scatter reduction, with “10” corresponding to maximum scatter reduction and “0” indicating no post-processing scatter reduction. Raw images being reconstructed multiple times ensured any difference in the images could be attributed to the reconstructions themselves rather than the image acquisitions. Each raw image was reconstructed with five different strengths (90 total reconstructions). The algorithm strengths carefully chosen for analysis encapsulated the spectrum of the SCA capabilities, with the minimum (level 0), maximum (level 10), and intermediate (levels 2, 5 and 8) strengths selected.

The objective IQ was assessed by the contrast to noise ratio (CNR). CNR is considered an essential objective image metric and correlated with the visibility of objects within the image [28, 29]. To assess this, 225 mm2 regions of interest (ROIs) were drawn inside the projection of two aluminum attenuators using ImageJ® (National Institute of Health, Maryland, USA). ImageJ calculated the mean pixel value (PV) and associated standard deviation (SD) within each of the ROIs. A script ensured that the location and size of each ROI was the same across each of the reconstructions. The mean PV and SDs were calculated from the distribution of PVs across the three images within the set.

To calculate CNR for a given mAs and algorithm strength, the set average PV of the images acquired with the aluminum attenuator was subtracted from the set average PV from the images acquired without the aluminum attenuator, divided by this set’s SD. Each CNR was then normalized by the value calculated for the non-scatter-corrected image acquired at 1.4 mAs. This value was selected as it most closely mimics the exposure and reconstruction parameters used in clinical practice for a patient of the phantom’s size.

Subjective image quality

Thirty consecutive adult portable chest X-rays performed in the emergency department, which utilized anti-scatter correction from the Shimadzu Mobile DaRt unit, were included. Images containing motion artifacts, an incomplete field of view, or pediatric studies were excluded. All included studies were acquired using the “Chest Lscape AP Scatter Correction 100kVp” protocol and at variable, mAs based on patient habitus. All thirty images were performed at approximately 180 cm source to image distance (SID). A “level 5” scatter correction was implemented based on results found in the objective portion of the study and after discussions with Senior Radiologists. It also reflected an intermediate level between high noise/high contrast and low noise/low contrast within the images. The same 30 image data sets were then reprocessed without the SCA by unselecting the “Scatter Correction” checkbox in the image’s reconstruction properties. As a result, 60 de-identified CXR images in total were created. Therefore, the images processed under the “no” SCA also had the same kV, mAs, and SID parameters. This meant images were obtained with the SCA and then retrospectively had the correction turned off accurately represented how the images would have appeared if the images were acquired without the algorithm selected during acquisition.

Table 1 indicates the patient characteristics and study indications. Patients presented with a diverse range of physical features and clinical indications.

Table 1 Patient characteristics

Two radiologists with 31 years and 7 years of plain radiography experience independently reviewed all 60 images in the same randomized order. The reviewers were blinded to which images had the SCA applied. The observers studied one image at a time on the standard Picture Archiving and Communication System (PACS) to prevent bias. Both observers viewed the images under the same lighting conditions on monitors that met the performance requirements for general radiography reporting [30]. The reviewers graded the overall IQ (contrast, noise, and perceived resolution) of the lung, hilar, mediastinum, pleural cavity, bone, and chest wall on a 5-point Likert scale which covers most of the anatomical structures on chest radiography. Additionally, each observer also graded the overall diagnostic confidence as either “confident” or “not confident” based on the observers’ ability to diagnose the images.

Statistical analysis

The Cohen’s kappa statistic with a 95% confidence interval (CI) was employed to assess the inter-observer reliability between observer 1 and observer 2 for each category. The inter-observer effect size was determined against the Koch and Landis criteria [31]. Values equal to 0 indicate no agreement, 0.01–0.20 as none to slight, 0.21–0.40 as fair, 0.41–0.60 as moderate, 0.61–0.80 as substantial, and 0.81–1.00 as almost perfect agreement.

The median and range were calculated based on combined scores from both observers. A Mann–Whitney U test with a 95% CI was employed to determine whether a statistical difference in IQ existed between the scatter-corrected and non-scatter-corrected images for each anatomical location.

Results

Objective image quality

Figures 2 and 3 show the normalized CNRs for the aluminum attenuators anterior to the lung (high-contrast enhancement) and the heart (low-contrast enhancement) region of the phantom. Both figures show higher relative CNRs for image sets acquired at greater mAs. Figure 2 show the selected algorithm strengths improved the CNR of the high contrast object for most mAs settings, with only “level 8” and “level 10” algorithm strengths degrading the CNR at the two lowest tube current–time settings. “Level 5” provided the greatest CNR enhancement, with an approximately 3% increase when compared to the non-scatter-corrected image across all mAs settings. Figure 3 shows the contrast of the attenuator in the (low contrast) heart region. Figure 3 indicates the greatest relative CNR enhancement was achieved using algorithm “level 10” and the lowest relative CNR was noted for images that had not been scatter-corrected. The SCA “level 2,” “level 5,” “level 8,” and “level 10” elevated the relative CNR within the low-contrast-attenuator by an average of 10%, 15%, 23%, and 32%, respectively, across each of the tube voltages. Figure 3 also indicates a higher relative CNR for all algorithm strengths at 1.4 mAs when compared to the relative CNR of images acquired at 1.8 mAs without the scatter correction applied.

Fig. 2
figure 2

Relative CNR as a function of algorithm strength and mAs for high contrast attenuator

Fig. 3
figure 3

Relative CNR as a function of algorithm strength and mAs for low contrast attenuator

Subjective image quality comparison

Interobserver agreement

Table 2 indicates high interobserver agreement levels in scores provided for different regions by the two observers. When graded against the Koch and Landis criteria, all regions had a “substantial” level of interobserver agreement.

Table 2 Cohen’s kappa statistic for interobserver agreement for all anatomical regions

Subjective image quality analysis

A series of Mann–Whitney U tests showed the scatter-corrected images received statistically significantly higher ratings (Table 3). As noted in the mean rank and U-statistics, the observers recorded the SCA had the largest improvement in IQ in the mediastinum and lung regions. Bone and hilar regions were recorded as the lowest areas of improvement. All images with and without the SCA were considered diagnostically adequate for both reviewers to make an accurate diagnosis and were without image artifacts (Fig. 4).

Table 3 Mann–Whitney U test of the non-scatter-corrected and scatter-corrected images
Fig. 4
figure 4

Mobile digital frontal chest radiograph of a 27-year-old obese female patient demonstrated better demarcation of pulmonary vessels and hilar structures, and sharper interfaces of mediastinal/lung and lung/bone interfaces despite the body habitus on the image with SCA (right) as compared to the image without SCA (left)

Table 4 presents the IQ changes in individual anatomical parts by comparing the scatter-corrected image to the non-scatter-corrected image of each patient. An “improved” IQ for a given region was defined as a higher rating attributed to the scatter-corrected image than the matching non-scatter-corrected image. Similarly, a “reduced” image was defined as the scatter-corrected image attributed with a lower rating than the non-scatter-corrected image. The SCA improved the overall subjective contrast, noise, and perceived resolution across all anatomical regions in 50% of the images. The SCA only reduced subjective IQ in the lung, hilar, mediastinum, and chest wall in less than 2% of images and bone and pleural regions in less than 7% of cases.

Table 4 The change in IQ across included studies with the SCA applied for each region in %

Discussion

The ongoing COVID-19 pandemic poses a significant challenge to frontline health networks. Portable CXRs allow increased patient throughput and minimize the chance of virus spread during patient transit to the radiology department [7]. It offers a more feasible alternative for critically unwell patients. However, this increased uptake of portable CXRs puts increased focus on optimizing imaging, particularly IQ.

IQ refers to the exactness with which objects or structures are reproduced on a radiographic image. In chest radiographs, IQ refers to the ability to depict all anatomical structures clearly and accurately within the image, such as the lung, hilar, mediastinum, pleural, bone, and chest wall, and interpret any disease or pathology associated with these structures. Higher IQ typically allows for a more confident clinical diagnosis, which can impact the patient’s level of care. This study is the first to assess, objective or subjectively the impact of the Shimadzu SCA on IQ.

Both objective and subjective assessments indicated improved IQ of structures within radiographs utilizing the SCA. Objectively, the SCA algorithm increased the relative CNR for both measurement positions and across most mAs and algorithm strengths. This is likely attributed to the increased contrast instigated by the subtraction of scattered radiation within the phantom image. This contrast increase had the greatest benefit in the low-contrast cardiac region, where the difference in PV between the attenuator and the background was minimal. For the high-contrast attenuator superimposing the lung, the relative CNR increase was more modest, with level 5 increasing the relative CNR by an average of 3% compared to images reformatted without the SCA. “Level 8” and “level 10” were shown to have a detrimental effect on the CNR at low mAs values in regions of high contrast. This may be due to an enhancement of noise within the ROI, which was not proportional to the increase in contrast. However, “level 8” and “level 10” SCA strength provide positive CNR enhancement when utilizing a higher dose (1.8 mAs) acquisition. A greater enhancement of low contrast pathologies has also been noticed in other objective studies evaluating Philips Skyflow [32].

Subjectively, both observers agreed the SCA enhanced the overall IQ, contrast, noise, and apparent resolution in all anatomical regions of clinical patient CXRs, and they did not comment on any detrimental effect from significant image noise or artifacts. While the level of improvement was variable across anatomical parts, the SCA improves the lung and mediastinum in over 68% of scores. The hilar, pleural, bone and chest wall scores increased in 55%, 67%, 50%, and 57% of image sets, respectively. In less than 2% of images, there was an IQ reduction in the lung, hilar, mediastinum, and chest wall. In 7% of scores, one reviewer indicated the bone and pleural regions exhibited a reduction in IQ after applying the SCA; however, this IQ reduction was not indicated in the second reviewers scores. The impact of SCA on the IQ improvement on the bony regions was the least amongst all anatomical parts, which was likely due to the algorithm being based on the scattering properties of soft tissue and not bone. Also, due to a higher atomic number and a lower relative electron density associated with bone, proportionately fewer photons undergo Compton scattering when passing through, resulting in bone having a lower impact on the scatter and, therefore, less affected by the algorithm.

While it was determined the SCA increased the overall IQ of the clinical portable CXRs, it was also noted that all images within the non-scatter-corrected image set were of adequate diagnostic quality. The objective analysis indicated the SCA improved relative CNR within the attenuators. Figures 2 and 3 indicate a greater relative CNR was achieved for both attenuators when images were acquired with 1.4 mAs and reconstructed with a SCA strength of “level 5” compared to images acquired with a 1.8 mAs exposure without SCA reprocessing. This finding suggests that images utilising the SCA could be performed with a lower mAs [17], reducing the patient’s radiation exposure, and decreasing their theoretical stochastic risk [33]. Extrapolating such a dose reduction for all patients who receive portable CXRs would result in a substantial cumulative dose reduction to patients and staff tasked with performing the portable CXRs. However, a prospective study is required to confirm these findings.

This research has several limitations which should be considered in future research. This study only assessed the results of the algorithm at a single institution with a small number of patients in the subjective assessment. Other institutions may have different imaging protocols, impacting the SCA’s effect in the clinical images. Future studies assessing the SCA should evaluate its potential use in other anatomical regions such as the abdomen. Furthermore, despite this study indicating the SCA has provided an overall improvement for all patients within the present cohort, the level of improvement will likely vary across patient habitus sizes, particularly for patients with less typical physiques. Additionally, the objective analysis performed in this study was limited to a single phantom size. Therefore, future studies may aim to determine whether optimization is possible by adjusting SCA algorithm strengths based on the patient’s physical characteristics and whether the SCA is feasible for use in pediatric patients.

Conclusion

This study is the first to assess the effect of the SCA on the IQ of portable CXRs. Both objective and subjective measures showed the SCA improved the overall IQ of the portable CXR without increasing the patient’s radiation dose. The SCA significantly enhanced the overall IQ of all anatomical structures for various patient sizes and pathologies.