Abstract
For the purpose of developing reflection pulse oximetry, we have analyzed the light propagation in tissue to calculate the photon-visit depth for reflected light and to estimate the measured intensities of transmitted and reflected light using the analytical solutions of the photon diffusion equation. The effects of the reflectance of the probe surface have been studied, and a low reflecting surface has been found to provide deeper photon-visit depth than a high reflecting surface for reflection pulse oximetry. Based on analytical estimation, we made reflection type probes, and conducted preliminary experiments to confirm the feasibility of this type of pulse oximetry.
Similar content being viewed by others
References
J. G. Webster: Design of Pulse Oximeters (IOP Publishing, Bristol, 1997).
A. Zourabian, A. Siegel, B. Chance, N. Ramanujan, M. Rode and D. A. Boas: J. Biomed. Opt. 5 (2000) 391.
T. J. Farrell, M. S. Patterson and B. Wilson: Med. Phys. 19 (1992) 879.
K. Furutsu and Y. Yamada: Phys. Rev. E 50 (1994) 3634.
M. S. Patterson, B. Chance and B. C. Wilson: Appl. Opt. 28 (1989) 2331.
W.-F. Cheong, S. A. Prahl and A. J. Welch: IEEE J. Quant. Electr. 26 (1990) 2166.
C. R. Simpson, M. Kohl, M. Essenpreis and M. Cope: Phys. Med. Biol. 43 (1998) 2465.
T. L. Troy and S. N. Thennadil: J. Biomed. Opt. 6 (2001) 167.
M. S. Patterson, S. Andersson-Engels, B. C. Wilson and E. K. Osei: Appl. Opt. 34 (1995) 22.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Matsushita, K., Aoki, K., Kakuta, N. et al. Fundamental Study of Reflection Pulse Oximetry. OPT REV 10, 482–487 (2003). https://doi.org/10.1007/s10043-003-0482-0
Received:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s10043-003-0482-0