Auswirkung verschiedener Bohrlochdurchmesser auf die Haltekraft von Osteosyntheseschrauben

The effect of various pilot hole sizes on the holding power of osteosynthesis screws

  • Originalien
  • Published:
Mund-, Kiefer- und Gesichtschirurgie Aims and scope Submit manuscript

Zusammenfassung

Osteosyntheseschrauben können beim Eindrehen nach Überschreitung der maximalen Torsionsbelastung des Schraubenschafts abbrechen. Durch Verwendung eines dickeren Vorbohrers wird die Torsionsbelastung der Schrauben reduziert. Ziel dieser Studie war, in In-vitro-Untersuchungen zu ermitteln, bis zu welchem Durchmesser die Vorbohrung für Mikro- und Minischrauben erweitert werden kann, ohne daß die Haltekraft der Schrauben abnimmt. 1,5- und 2-mm-Titanschrauben wurden nach Vorbohrung mit Bohrern, deren Durchmesser 66–95% des äußeren Schraubendurchmessers betrug, in Holz (2 mm), PVC (2; 3; 4 mm), und Kortikalis aus Schweineunterkiefern (2; 3; 4 mm) eingedreht. Die Haltekraft der Schrauben wurde durch Messung des maximalen Drehmoments und durch Ausreißtests bestimmt; für jede Kombination aus Schraube, Bohrlochdurchmesser und Testmaterialdicke wurden 10 Versuche durchgeführt. Eine multiphasische Regressionsanalyse ermöglichte die Berechnung der kritischen Bohrlochdurchmesser, bei deren Überschreitung die Schraubenhaltekraft abfiel. Bei den Drehmomentmessungen der Mikroschrauben wurden kritische Bohrlochdurchmesser (KBD) zwischen 83 und 85% des äußeren Schraubendurchmessers ermittelt, der KBD der Minischrauben lag zwischen 80 und 90% des äußeren Schraubendurchmessers. In den Ausreißversuchen betrug der KBD der Mikroschrauben 83–89% des äußeren Schraubendurchmessers, der KBD der Minischrauben 79–91% des äußeren Schraubendurchmessers. Der Durchmesser des Vorbohrers konnte im Durchschnitt auf 85% des äußeren Schraubendurchmessers vergrößert werden, ohne die Haltekraft der Schrauben deutlich zu verringern.

Summary

The standard drill size for most osteosynthesis screws is about 75% of the screw’s external diameter. When screws are inserted in thick cortical bone, a small pilot hole size can result in high torsional stress, leading to screw fracturing. The aim of this study was to enlarge the drill size up to a critical pilot hole size, exceeding of which leads to a rapid decrease of the screw’s holding power. Titanium screws of diameter 1.5 and 2 mm were inserted in discs of PVC, wood and porcine mandibular bone with different thickness between 2 to 4 mm, using pilot hole sizes of 66– 95% of the screw’s external diameter. The maximum torque was recorded and pull-out tests were performed. Ten trials were conducted for each screw–pilot hole size–material combination, yielding a total of 1560 tests. A multiphase regression analysis was performed to calculate the critical pilot hole size (CPHS). In torque measurements, the CPHS of microscrews were between 83% and 85% of screw outer diameter (SOD) and the CPHS of miniscrews were between 80% and 90% of SOD. In pull-out analysis the CPHS of microscrews were between 83% and 89% of SOD and the CPHS of miniscrews were between 79% and 91% of SOD. The CPHS was thus found to be around 85% of the screw’s external diameter. Up to this critical point the pilot hole size may be increased without affecting the holding power of the screws.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Heidemann, W., Gerlach, K.L., Gröbel, K.H. et al. Auswirkung verschiedener Bohrlochdurchmesser auf die Haltekraft von Osteosyntheseschrauben. Mund Kiefer GesichtsChir 2, 136–140 (1998). https://doi.org/10.1007/s100060050048

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s100060050048

Schlüsselwörter

Key words

Navigation