In-vitro performance of CAD/CAM-fabricated implant-supported temporary crowns

  • Martin Rosentritt
  • Philipp Raab
  • Sebastian Hahnel
  • Matthias Stöckle
  • Verena Preis
Original Article

Abstract

Objectives

The aim of this study was to investigate the in-vitro performance and fracture resistance of a temporary computer-aided designed and computer-aided manufactured polymethylmethacrylate (CAD/CAM-PMMA) material as implant or tooth-supported single crown with respect to the clinical procedure (permanently bonded/temporarily cemented).

Materials and methods

Sixty-four crowns were fabricated on implants or human molar teeth simulating (a) labside procedure on prefabricated titanium-bonding base ([TiBase] implant crown bonded in laboratory, screwed chairside), (b) labside procedure ([LAB] standard abutment and implant crown bonded in laboratory, screwed chairside), (c) chairside procedure ([CHAIR] implant crown bonded to abutment), and (d) reference ([TOOTH] crowns luted on prepared human teeth). Crowns were made of a CAD/CAM-PMMA temporary material (TelioCAD, Ivoclar-Vivadent). For investigating the influence of fixation, half of the crowns were permanently (P) or temporarily (T) bonded. Combined thermal cycling and mechanical loading (TCML) was performed simulating a 5-year clinical situation. Fracture force was determined. Data were statistically analyzed (Kolmogorov-Smirnov test, one-way ANOVA; post hoc Bonferroni, α = 0.05).

Results

All restorations survived TCML without visible failures. Fracture results varied between 3034.3 (Tooth-P) and 1602.9 N (Tooth-T) [TOOTH], 1510.5 (TiBase-P) and 963.6 N (TiBase-T) [TiBase], 2691.1 (LAB-P) and 2064.5 N (LAB-T) [LAB], and 1609.4 (Chair-P) and 1253.0 N (Chair-T) [CHAIR]. Tested groups showed significantly (p < 0.001) different fracture values. Failure pattern was characterized by fractures in mesial-distal, buccal-oral, or mixed (mesial-distal/buccal-oral) directions, with differences for the individual groups.

Conclusions

Temporary CAD/CAM crowns showed no different in-vitro performance but provided different fracture results that depended on cementation, screw channel, and type of abutment.

Clinical relevance

All bonded and screwed PMMA crowns were in a range where clinical application seems not restricted.

Keywords

CAD/CAM composites PMMA, temporary Provisional, chewing simulation Implant crown Abutment Fracture resistance 

Notes

Acknowledgements

We would like to thank the manufacturers for providing the materials.

Compliance with ethical standards

Conflict of interest

Prof. Dr. Martin Rosentritt declares that he has third-party research projects with all dental companies.

Dr. Philipp Raab declares that he has no conflict of interest.

Prof. Dr. Sebastian Hahnel declares that he has no conflict of interest.

Dr. Matthias Stöckle declares that he has no conflict of interest.

PD Dr. Verena Preis declares that she has no conflict of interest.

Funding

No.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

Informed consent

For this type of study, formal consent is not required.

References

  1. 1.
    Gratton DG, Aquilino SA (2004) Interim restorations. Dent Clin N Am 48:487–497CrossRefGoogle Scholar
  2. 2.
    Rayyan MM, Aboushelib M, Sayed NM, Ibrahim A, Jimbo R (2015) Comparison of interim restorations fabricated by CAD/CAM with those fabricated manually. J Prosthet Dent 114:414–419CrossRefPubMedGoogle Scholar
  3. 3.
    Lodding DW (1997) Long-term esthetic provisional restorations in dentistry. Curr Opin Cosmet Dent 4:16–21PubMedGoogle Scholar
  4. 4.
    Wiegand A, Stucki L, Hoffmann R, Attin T, Stawarczyk B (2015) Repairability of CAD/CAM high-density PMMA- and composite-based polymers. Clin Oral Investig 19:2007–2013CrossRefPubMedGoogle Scholar
  5. 5.
    Keul C, Müller-Hahl M, Eichberger M, Liebermann A, Roos M, Edelhoff D, Stawarczyk B (2014) Impact of different adhesives on work of adhesion between CAD/CAM polymers and resin composite cements. J Dent 42:1105–1114CrossRefPubMedGoogle Scholar
  6. 6.
    Attia A, Abdelaziz KM, Freitag S, Kern M (2006) Fracture load of composite resin and feldspathic all-ceramic CAD/CAM crowns. J Prosthet Dent 95:117–123CrossRefPubMedGoogle Scholar
  7. 7.
    Wimmer T, Ender A, Roos M, Stawarczyk B (2013) Fracture load of milled polymeric fixed dental prostheses as a function of connector cross-sectional areas. J Prosthet Dent 110:288–295CrossRefPubMedGoogle Scholar
  8. 8.
    Stawarczyk B, Özcan M, Trottmann A, Schmutz F, Roos M, Hämmerle C (2013) Two-body wear rate of CAD/CAM resin blocks and their enamel antagonists. J Prosthet Dent 109:325–332CrossRefPubMedGoogle Scholar
  9. 9.
    Seiss M, Nitz S, Kleinsasser N, Buters JT, Behrendt H, Hickel R, Reichl FX (2007) Identification of 2,3-epoxymethacrylic acid as an intermediate in the metabolism of dental materials in human liver microsomes. Dent Mater 23:9–16CrossRefPubMedGoogle Scholar
  10. 10.
    Bähr N, Keul C, Edelhoff D, Eichberger M, Roos M, Gernet W, Stawarczyk B (2013) Effect of different adhesives combined with two resin composite cements on shear bond strength to polymeric CAD/CAM materials. Dent Mater J 32:492–501CrossRefPubMedGoogle Scholar
  11. 11.
    Keul C, Kohen D, Eichberger M, Roos M, Gernet W, Stawarczyk B (2015) The effect of different pretreatment methods of PMMA-based crowns on the long-term tensile bond strength to dentin abutments. Clin Oral Investig 19:35–43CrossRefPubMedGoogle Scholar
  12. 12.
    Stawarczyk B, Basler T, Ender A, Roos M, Ozcan M, Hämmerle C (2012) Effect of surface conditioning with airborne-particle abrasion on the tensile strength of polymeric CAD/CAM crowns luted with self-adhesive and conventional resin cements. J Prosthet Dent 107:94–101CrossRefPubMedGoogle Scholar
  13. 13.
    Güth JF, Almeida E, Silva JS, Ramberger M, Beuer F, Edelhoff D (2012) Treatment concept with CAD/CAM-fabricated high-density polymer temporary restorations. J Esthet Restor Dent 24:310–318CrossRefPubMedGoogle Scholar
  14. 14.
    Güth JF, Almeida E, Silva JS, Beuer FF, Edelhoff D (2012) Enhancing the predictability of complex rehabilitation with a removable CAD/CAM-fabricated long-term provisional prosthesis: a clinical report. J Prosthet Dent 107:1–6CrossRefPubMedGoogle Scholar
  15. 15.
    Hebel KS, Gajjar RC (1997) Cement-retained versus screw-retained implant restorations: achieving optimal occlusion and esthetics in implant dentistry. J Prosthet Dent 77:28–35CrossRefPubMedGoogle Scholar
  16. 16.
    Weber HP, Kim DM, Ng MW, Hwang JW, Fiorellini JP (2006) Peri-implant soft-tissue health surrounding cement- and screw-retained implant restorations: a multi-center, 3-year prospective study. Clin Oral Implants Res 17:375–379CrossRefPubMedGoogle Scholar
  17. 17.
    Chu KM, Tredwin CJ, Setchell DJ, Hems E (2005) Effect of screw hole filling on retention of implant crowns. Eur J Prosthodont Restor Dent 13:154–158PubMedGoogle Scholar
  18. 18.
    Lewis S (1995) Anterior single-tooth implant restorations. Int J Periodontics Restorative Dent 15:30–41PubMedGoogle Scholar
  19. 19.
    Saba S (1997) Anatomically correct soft tissue profiles using fixed detachable provisional implant restorations. J Can Dent Assoc 63(767–768):770Google Scholar
  20. 20.
    Breeding LC, Dixon DL (1996) Transfer of gingival contours to a master cast. J Prosthet Dent 75:341–343CrossRefPubMedGoogle Scholar
  21. 21.
    Neale D, Chee WW (1994) Development of implant soft tissue emergence profile: a technique. J Prosthet Dent 71:364–368CrossRefPubMedGoogle Scholar
  22. 22.
    Capp NJ (1985) The diagnostic use of provisional restorations. Restor Dent 1(92):94–98Google Scholar
  23. 23.
    Moscovitch MS, Saba S (1996) The use of a provisional restoration in implant dentistry: a clinical report. Int J Oral Maxillofac Implants 11:395–399PubMedGoogle Scholar
  24. 24.
    Rosentritt M, Hmaidouch R, Behr M, Handel G, Schneider-Feyrer S (2011) Fracture resistance of zirconia FPDs with adhesive bonding versus conventional cementation. Int J Prosthodont 24:168–171PubMedGoogle Scholar
  25. 25.
    Rosentritt M, Behr M, Gebhard R, Handel G (2006) Influence of stress simulation parameters on the fracture strength of all-ceramic fixed-partial dentures. Dent Mater 22:176–182CrossRefPubMedGoogle Scholar
  26. 26.
    Rosentritt M, Behr M, van der Zel JM, Feilzer AJ (2009) Approach for valuating the influence of laboratory simulation. Dent Mater 25:348–352CrossRefPubMedGoogle Scholar
  27. 27.
    Rosentritt M, Siavikis G, Behr M, Kolbeck C, Handel G (2008) Approach for valuating the significance of laboratory simulation. J Dent 36:1048–1053CrossRefPubMedGoogle Scholar
  28. 28.
    Menini M, Conserva E, Tealdo T, Bevilacqua M, Pera F, Signori A, Pera P (2013) Shock absorption capacity of restorative materials for dental implant prostheses: an in vitro study. Int J Prosthodont 26:549–556CrossRefPubMedGoogle Scholar
  29. 29.
    Pjetursson BE, Brägger U, Lang NP, Zwahlen M (2007) Comparison of survival and complication rates of tooth-supported fixed dental prostheses (FDPs) and implant-supported FDPs and single crowns (SCs). Clin Oral Implants Res 18(Suppl 3):97–113CrossRefPubMedGoogle Scholar
  30. 30.
    Shembish FA, Tong H, Kaizer M, Janal MN, Thompson VP, Opdam NJ, Zhang Y (2016) Fatigue resistance of CAD/CAM resin composite molar crowns. Dent Mater 32:499–509CrossRefPubMedPubMedCentralGoogle Scholar
  31. 31.
    de Kok P, Kleverlaan CJ, de Jager N, Kuijs R, Feilzer AJ (2015) Mechanical performance of implant-supported posterior crowns. J Prosthet Dent 114:59–66CrossRefPubMedGoogle Scholar
  32. 32.
    Conserva E, Menini M, Tealdo T, Bevilacqua M, Ravera G, Pera F, Pera P (2009) The use of a masticatory robot to analyze the shock absorption capacity of different restorative materials for prosthetic implants: a preliminary report. Int J Prosthodont 22:53–55PubMedGoogle Scholar
  33. 33.
    Jung RE, Pjetursson BE, Glauser R, Zembic A, Zwahlen M, Lang NP (2008) A systematic review of the 5-year survival and complication rates of implant-supported single crowns. Clin Oral Implants Res 19:119–130CrossRefPubMedGoogle Scholar
  34. 34.
    Bijjargi S, Chowdhary R (2013) Stress dissipation in the bone through various crown materials of dental implant restoration: a 2-D finite element analysis. J Investig Clin Dent 4:172–177CrossRefPubMedGoogle Scholar
  35. 35.
    Trulsson M, Gunne HS (1998) Food-holding and -biting behavior in human subjects lacking periodontal receptors. J Dent Res 77:574–582CrossRefPubMedGoogle Scholar
  36. 36.
    Silva GC, Cornacchia TM, de Magalhães CS, Bueno AC, Moreira AN (2014) Biomechanical evaluation of screw- and cement-retained implant-supported prostheses: a nonlinear finite element analysis. J Prosthet Dent 112:1479–1488CrossRefPubMedGoogle Scholar
  37. 37.
    Wittneben JG, Millen C, Bragger U (2014) Clinical performance of screw- versus cement-retained fixed implant-supported reconstructions - a systematic review. Int J Oral Maxillofac Implants 29(Suppl):84–98CrossRefPubMedGoogle Scholar
  38. 38.
    Torrado E, Ercoli C, Al Mardini M, Graser GN, Tallents RH, Cordaro L (2004) A comparison of the porcelain fracture resistance of screw-retained and cement-retained implant-supported metal-ceramic crowns. J Prosthet Dent 91:532–537CrossRefPubMedGoogle Scholar
  39. 39.
    Gressler May L, Kelly JR, Bottino MA, Hill T (2015) Influence of the resin cement thickness on the fatigue failure loads of CAD/CAM feldspathic crowns. Dent Mater 31:895–900CrossRefPubMedGoogle Scholar
  40. 40.
    Rosentritt M, Hahnel S, Engelhardt F, Behr M, Preis V (2016) In vitro performance and fracture resistance of CAD/CAM-fabricated implant supported molar crowns. Clin Oral Investig. doi: 10.1007/s00784-016-1898-9 Google Scholar
  41. 41.
    Preis V, Kammermeier A, Handel G, Rosentritt M (2016) In vitro performance of two-piece zirconia implant systems for anterior application. Dent Mater 32:765–774CrossRefPubMedGoogle Scholar
  42. 42.
    Varga S, Spalj S, Lapter Varga M, Anic Milosevic S, Mestrovic S, Slaj M (2011) Maximum voluntary molar bite force in subjects with normal occlusion. Eur J Orthod 3:427–433CrossRefGoogle Scholar

Copyright information

© Springer-Verlag Berlin Heidelberg 2017

Authors and Affiliations

  • Martin Rosentritt
    • 1
  • Philipp Raab
    • 1
  • Sebastian Hahnel
    • 1
  • Matthias Stöckle
    • 1
  • Verena Preis
    • 1
  1. 1.Department of Prosthetic DentistryUniversity Medical Center RegensburgRegensburgGermany

Personalised recommendations