Reliability of a portable device for the detection of sleep bruxism

  • Andrea Deregibus
  • Tommaso CastroflorioEmail author
  • Andrea Bargellini
  • Cesare Debernardi
Original Article

Abstract

Objectives

The aim of the study was to assess the repeatability in detecting sleep bruxism (SB) episodes by combined surface electromyography and heart rate (HR) signals recorded by a compact portable device (Bruxoff®). SB episodes are preceded by a sudden HR change. Thus, HR detection increases the precision of automatic detection of SB.

Materials and methods

Ten healthy subjects (five women and five men; 30.2 ± 11.02 years) were selected for the study. Rhythmic masseter muscle activities, constituting the basic pattern of SB, were detected during three nights of recording during three different weeks with the Bruxoff device.

Results

The two-way ANOVA was not significant for SB episodes per night, SB episodes per hour, and heart frequency: no significant differences were observed during the three different nights of recording for each of the abovementioned variables (P > 0.05). The intraclass correlation coefficient showed a good reproducibility for SB episodes per night (69 %), SB per hour (74 %), and heart frequency (82 %). A poor reproducibility was revealed for the number of masseter contractions (53 %). The Pearson analysis showed the absence of a significant correlation between the number of masseter contractions per night and the number of SB episodes per night (r = −0.02, P = 0.91).

Conclusions

The Bruxoff device showed a good reproducibility of measurements of sleep bruxism episodes over time.

Clinical relevance

These findings are important in the light of the need for simple and reliable portable devices for the diagnosis of SB both in the clinical and research settings.

Keywords

Sleep bruxism Surface electromyography Masseter muscle Heart rate Rhythmic masticatory muscles activity 

Notes

Acknowledgments

The procedures were approved by the Lingotto Dental School Ethics Committee. All individuals gave their informed consent in accordance with the Helsinki Declaration. The research funding was provided by University of Torino.

Conflict of interest

The authors do not have any financial relation with Spes Medica and/or OTBioelettronica. There is no conflict of interest with any financial organization regarding the material discussed in the manuscript.

References

  1. 1.
    Lobbezoo F, Ahlberg J, Glaros AG, Kato T, Koyano K, Lavigne GJ, de Leeuw R, Manfredini D, Svensson P, Winocur E (2013) Bruxism defined and graded: an international consensus. J Oral Rehabil 40:2–4PubMedCrossRefGoogle Scholar
  2. 2.
    Lavigne GJ, Khoury S, Abe S, Yamaguchi T, Raphael K (2008) Bruxism physiology and pathology: an overview for clinicians. J Oral Rehabil 35:476–494PubMedCrossRefGoogle Scholar
  3. 3.
    Thorpy MJ (1990) ICSD—International Classification of Sleep Disorders: Diagnostic and Coding Manual. Diagnostic Classification Steering Committee, American Sleep Disorders AssociationGoogle Scholar
  4. 4.
    Carra MC, Huynh N, Lavigne G (2012) Sleep bruxism: a comprehensive overview for the dental clinician interested in sleep medicine. Dent Clin N Am 56:387–413PubMedCrossRefGoogle Scholar
  5. 5.
    Castroflorio T, Mesin L, Tartaglia GM, Sforza C, Farina D (2013) Use of electromyographic and electrocardiographic signals to detect sleep bruxism episodes in a natural environment. IEEE J Biomed Health Inform 17:994–1001PubMedCrossRefGoogle Scholar
  6. 6.
    Minakuchi H, Sakaguchi C, Hara ES, Maekawa K, Matsuka Y, Clark GT, Kuboki T (2012) Multiple sleep bruxism data collected using a self-contained EMG detector/analyzer system in asymptomatic healthy subjects. Sleep Breath 16:1069–1072PubMedCrossRefGoogle Scholar
  7. 7.
    Shochat T, Gavish A, Arons E, Hadas N, Molotsky A, Lavie P, Oksenberg A (2007) Validation of the BiteStrip screener for sleep bruxism. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 104:e32–e39PubMedCrossRefGoogle Scholar
  8. 8.
    Mainieri VC, Saueressig AC, Pattussi MP, Fagondes SC, Grossi ML (2012) Validation of the Bitestrip versus polysomnography in the diagnosis of patients with a clinical history of sleep bruxism. Oral Surg Oral Med Oral Pathol Oral Radiol 113:612–617PubMedCrossRefGoogle Scholar
  9. 9.
    Yamaguchi T, Abe S, Rompré PH, Manzini C, Lavigne GJ (2012) Comparison of ambulatory and polysomnographic recording of jaw muscle activity during sleep in normal subjects. J Oral Rehabil 39:2–10PubMedCrossRefGoogle Scholar
  10. 10.
    Gallo LM, Lavigne G, Rompré P, Palla S (1997) Reliability of scoring EMG orofacial events: polysomnography compared with ambulatory recordings. J Sleep Res 6:259–263PubMedCrossRefGoogle Scholar
  11. 11.
    Lavigne GJ, Rompré PH, Poirier G, Huard H, Kato T, Montplaisir JY (2001) Rhythmic masticatory muscle activity during sleep in humans. J Dent Res 80:443–448PubMedCrossRefGoogle Scholar
  12. 12.
    Lavigne GJ, Huynh N, Kato T, Okura K, Adachi K, Yao D, Sessle B (2007) Genesis of sleep bruxism: motor and autonomic-cardiac interactions. Arch Oral Biol 52:381–384PubMedCrossRefGoogle Scholar
  13. 13.
    Mizumori T, Sumiya M, Kobayashi Y, Inano S, Yatani H (2013) Prediction of sleep bruxism events by increased heart rate. Int J Prosthodont 26:239–243PubMedCrossRefGoogle Scholar
  14. 14.
    Lavigne GJ, Guitard F, Rompré PH, Montplaisir JY (2001) Variability in sleep bruxism activity over time. J Sleep Res 10:237–244PubMedCrossRefGoogle Scholar
  15. 15.
    Castroflorio T, Bracco P, Farina D (2008) Surface electromyography in the assessment of jaw elevator muscles. J Oral Rehabil 35:638–645PubMedCrossRefGoogle Scholar
  16. 16.
    Farina D, Cescon C (2001) Concentric-ring electrode systems for noninvasive detection of single motor unit activity. IEEE Trans Biomed Eng 48:1326–1334PubMedCrossRefGoogle Scholar
  17. 17.
    Rompré PH, Daigle-Landry D, Guitard F, Montplaisir JY, Lavigne GJ (2007) Identification of a sleep bruxism subgroup with a higher risk of pain. J Dent Res 86:837–842PubMedCrossRefGoogle Scholar
  18. 18.
    Bartko JJ (1966) The intraclass correlation coefficient as a measure of reliability. Psychol Rep 19:3–11PubMedCrossRefGoogle Scholar
  19. 19.
    Rivera-Morales WC, McCall WD Jr (1995) Reliability of a portable electromyographic unit to measure bruxism. J Prosthet Dent 73:184–189PubMedCrossRefGoogle Scholar

Copyright information

© Springer-Verlag Berlin Heidelberg 2013

Authors and Affiliations

  • Andrea Deregibus
    • 1
    • 2
  • Tommaso Castroflorio
    • 1
    Email author
  • Andrea Bargellini
    • 1
    • 2
  • Cesare Debernardi
    • 1
  1. 1.Department of Surgical Sciences, Specialization School of Orthodontics, Dental SchoolUniversity of TorinoTorinoItaly
  2. 2.Department of Surgical Sciences, Gnathology Unit, Dental SchoolUniversity of TorinoTorinoItaly

Personalised recommendations