Abstract
Background
No reports have been published on detailed risk factors for rod fracture after spinal deformity correction and fusion. The purpose of this study was to analyze clinical and radiographic risk factors of rod fracture after long construct fusion for spinal deformity.
Methods
The survey subjects were 155 cases who were diagnosed with spinal deformity and underwent correction and fusion surgery with long construct instrumentation (>3 levels, average 10.3 levels) between July 2004 and June 2010. The subjects comprised 32 males and 123 females with a mean age of 19.0 (range 8–78) years. The mean Cobb angle was 61.0 ± 16.1° preoperatively and 25.7 ± 16.9° postoperatively. Univariate analysis and logistic regression analysis were performed.
Results
Rod fracture occurred in 8 of 155 cases (5.2 %). The mean period from surgery to rod fracture was 18.1 months (range 2–37). The level of fracture ranged from the thoracolumbar junction to the lumbosacral vertebrae. Six patients had fracture near the fused lower end and two patients had fracture at the thoracolumbar junction. Univariate analysis revealed that non-ambulatory status, preoperative kyphosis, small-diameter rods, multiple surgery, and use of iliac screws were significant risk factors for rod fracture. Sex, obesity, severity of preoperative scoliosis, and rod material were not significant risk factors. Logistic regression analysis revealed that use of iliac screws (odds ratio: 81.9, 95 % confidence interval: 7.2–935.0, p < 0.001) and small-diameter (<6 mm) rods (odds ratio: 16.3, 95 % confidence interval: 1.7–152.6, p = 0.015) were risk factors for rod fracture.
Conclusions
The incidence of rod fracture after long construct fusion for spinal deformity was 5.2 %. Iliac screw fixation and small-diameter rods were risk factors for rod fracture.
Similar content being viewed by others
References
Reames DL, Smith JS, Fu KM, Polly DW Jr, Ames CP, Berven SH, Perra JH, Glassman SD, McCarthy RE, Knapp RD Jr, Heary R, Shaffrey CI; Scoliosis Research Society Morbidity and Mortality Committee. Complications in the surgical treatment of 19,360 cases of pediatric scoliosis—a review of the Scoliosis Research Society Morbidity and Mortality database. Spine. 2011;36:1484–91.
Sansur CA, Smith JS, Coe JD, Glassman SD, Berven SH, Polly DW Jr, Perra JH, Boachie-Adjei O, Shaffrey CI. Scoliosis Research Society Morbidity and Mortality of adult scoliosis surgery. Spine. 2011;36:E593–7.
Bridwell KH, Baldus C, Berven S, Edwards C 2nd, Glassman S, Hamill C, Horton W, Lenke LG, Ondra S, Schwab F, Shaffrey C, Wootten D. Changes in radiographic and clinical outcomes with primary treatment adult spinal deformity surgeries from two years to three- to five-years follow-up. Spine. 2010;35:1849–54.
Watanabe K, Lenke LG, Bridwell KH, Kim YJ, Watanabe K, Kim YW, Kim YB, Hensley M, Stobbs G. Comparison of radiographic outcomes for the treatment of scoliotic curves greater than 100 degrees—wires versus hooks versus screws. Spine. 2008;33:1084–91.
Yang JS, Sponseller PD, Thompson GH, Akbarnia BA, Emans JB, Yazici M, Skaggs DL, Shah SA, Salari P, Poe-Kochert C; Growing Spine Study Group. Growing rod fractures -risk factors and opportunities for prevention. Spine. 2011;36:1639–44.
Erwin WD, Dickson JH, Harrington PR. Clinical review of patients with broken Harrington rods. J Bone Joint Surg Am. 1980;62:1302–7.
Okamoto T, Neo M, Fujibayashi S, Ito H, Takemoto M, Nakamura T. Mechanical implant failure in posterior cervical spine fusion. Eur Spine J. 2012;21:328–34.
Suda K, Ito M, Abumi K, Haba H, Taneichi H, Kaneda K. Radiological risk factors of pseudoarthrosis and/or instrument breakage after PLF with the pedicle screw system in isthmic spondylolisthesis. J Spinal Disord Tech. 2006;19:541–6.
Neo M, Fujibayashi S, Nakamura T. Mechanical failure of spinal instrumentation. Seikeigeka. 2009;60:886–93 (in Japanese).
Dick JC, Bourgeault CA. Notch sensitivity of titanium alloy, commercially pure titanium, and stainless steel spinal implants. Spine. 2001;26:1168–72.
O’Brien JR, Matteini L, Yu WD, Kebaish KM. Feasibility of minimally invasive sacropelvic fixation: percutaneous S2 alar iliac fixation. Spine. 2010;35:460–4.
Serhan H, Mhatre D, Newton P, Giorgio P, Sturm P. Would CoCr rods provide better correctional forces than stainless steel or titanium for rigid scoliosis curves? J Spinal Disord Tech. 2013;26:E70–4.
Modi HN, Suh SW, Hong JY, Yang JH. Posterior multilevel vertebral osteotomy for severe and rigid idiopathic and nonidiopathic kyphoscoliosis: a further experience with minimum two-year follow-up. Spine. 2011;36:1146–53.
Conflict of interest
The authors declare that they have no conflict of interest.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Akazawa, T., Kotani, T., Sakuma, T. et al. Rod fracture after long construct fusion for spinal deformity: clinical and radiographic risk factors. J Orthop Sci 18, 926–931 (2013). https://doi.org/10.1007/s00776-013-0464-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00776-013-0464-4