Abstract
Background
A number of methods for coating implants with bioactive ceramics have been reported to improve osseointegration in bone, but the effects of bioactive ceramic coatings on the osseointegration of cancellous screws are not known. Accordingly, biomechanical and histomorphometric analyses of the bone-screw interface of uncoated cancellous screws and cancellous screws coated with four different bioactive ceramics were performed.
Methods
After coating titanium alloy cancellous screws with calcium pyrophosphate (CPP), CaO–SiO2–B2O3 glass-ceramics (CSG), apatite-wollastonite 1:3 glass-ceramics (W3G), and CaO–SiO2–P2O5–B2O3 glass-ceramics (BGS-7) using an enameling method, the coated and the uncoated screws were inserted into the proximal tibia and distal femur metaphysis of seven male mongrel dogs. The torque values of the screws were measured at the time of insertion and at removal after 8 weeks. The bone-screw contact ratio was analyzed by histomorphometry.
Results
There was no significant difference in the insertion torque between the uncoated and coated screws. The torque values of the CPP and BGS-7 groups measured at removal after 8 weeks were significantly higher than those of the uncoated group. Moreover, the values of the CPP and BGS-7 groups were significantly higher than the insertion torques. The fraction of bone-screw interface measured from the undecalcified histological slide showed that the CPP, W3G, and BGS-7 groups had significantly higher torque values in the cortical bone area than the uncoated group, and the CPP and BGS-7 groups had significantly higher torque values in the cancellous bone area than the uncoated group.
Conclusions
In conclusion, a cancellous screw coated with CPP and BGS-7 ceramic bonds directly to cancellous bone to improve the bone-implant osseointegration. This may broaden the indications for cancellous screws by clarifying their contribution to improving osseointegration, even in the cancellous bone area.
Similar content being viewed by others
References
Ducheyne P, Beight J, Cuckler J, Evans B, Radin S. Effect of calcium phosphate coating characteristics on early post-operative bone tissue ingrowth. Biomaterials. 1990;11:531–40.
Gottlander M, Johansson CB, Wennerberg A, Albrektsson T, Radin S, Ducheyne P. Bone tissue reactions to an electrophoretically applied calcium phosphate coating. Biomaterials. 1997;18:551–7.
Magyar G, Toksvig-Larsen S, Moroni A. Hydroxyapatite coating of threaded pins enhances fixation. J Bone Joint Surg Br. 1997;79:487–9.
Merolli A, Moroni A, Faldini C, Tranquilli Leali P, Giannini S. Histomorphological study of bone response to hydroxyapatite coating on stainless steel. J Mater Sci Mater Med. 2003;14:327–33.
Moroni A, Orienti L, Stea S, Visentin M. Improvement of the bone-pin interface with hydroxyapatite coating: an in vivo long-term experimental study. J Orthop Trauma. 1996;10:236–42.
Radin S, Ducheyne P, Berthold P, Decker S. Effect of serum proteins and osteoblasts on the surface transformation of a calcium phosphate coating: a physicochemical and ultrastructural study. J Biomed Mater Res. 1998;39:234–43.
Bauer TW, Geesink RC, Zimmerman R, McMahon JT. Hydroxyapatite-coated femoral stems. Histological analysis of components retrieved at autopsy. J Bone Joint Surg Am. 1991;73:1439–52.
Capello WN, D’Antonio JA, Manley MT, Feinberg JR. Hydroxyapatite in total hip arthroplasty. Clinical results and critical issues. Clin Orthop Relat Res. 1998;(355):200–11.
D’Antonio JA, Capello WN, Manley MT, Geesink R. Hydroxyapatite femoral stems for total hip arthroplasty: 10- to 13-year follow-up. Clin Orthop Relat Res. 2001;(393):101–11.
Ellies LG, Nelson DG, Featherstone JD. Crystallographic changes in calcium phosphates during plasma-spraying. Biomaterials. 1992;13:313–6.
Fazan F, Marquis PM. Dissolution behavior of plasma-sprayed hydroxyapatite coatings. J Mater Sci Mater Med. 2000;11:787–92.
Inagaki M, Kameyama T. Phase transformation of plasma-sprayed hydroxyapatite coating with preferred crystalline orientation. Biomaterials. 2007;28:2923–31.
Klein CP, Patka P, van der Lubbe HB, Wolke JG, de Groot K. Plasma-sprayed coatings of tetracalciumphosphate, hydroxyl-apatite, and alpha-TCP on titanium alloy: an interface study. J Biomed Mater Res. 1991;25:53–65.
Lee JH, Ryu HS, Lee DS, Hong KS, Chang BS, Lee CK. Biomechanical and histomorphometric study on the bone-screw interface of bioactive ceramic-coated titanium screws. Biomaterials. 2005;26:3249–57.
Koo KH, Hwang CJ, Lee JH, Chang BS, Lee CK. Treatment of bone defects in rabbit tibiae using CaO–SiO2–P2O5–B2O3 bioactive ceramics: radiological, biomechanical, and histological evaluation. Tissue Eng Reg Med. 2009;6:811–8.
Lee JH, Jeung UO, Jeon DH, Chang BS, Lee CK. Quantitative comparison of novel CaO–SiO2–P2O5–B2O3 glass–ceramics (BGS-7) with hydroxyapatite as bone graft extender in rabbit ilium. Tissue Eng Reg Med. 2010;7:540–7.
Lee JH, Chang BS, Jeung UO, Park KW, Kim MS, Lee CK. The first clinical trial of beta-calcium pyrophosphate as a novel bone graft extender in instrumented posterolateral lumbar fusion. Clin Orthopaedic Surg. 2011 (in press).
Lee JH, Chang BS, Ryu HS, Lee CK. A 90-day subchronic toxicity study of beta-calcium pyrophosphate in rat. Drug Chem Toxicol. 2009;32:277–82.
Kitsugi T, Yamamuro T, Nakamura T, Kotani S, Kokubo T, Takeuchi H. Four calcium phosphate ceramics as bone substitutes for non-weight-bearing. Biomaterials. 1993;14:216–24.
Lee JH, Ryu HS, Seo JH, Chang BS, Lee CK. A 90-day intravenous administration toxicity study of CaO–SiO(2)–P(2)O(5)–B(2)O(3) glass–ceramics (BGS-7) in rat. Drug Chem Toxicol. 2010;33:38–47.
Bigi A, Fini M, Bracci B, Boanini E, Torricelli P, Giavaresi G, Aldini NN, Facchini A, Giardino R. The response of bone to nanocrystalline hydroxyapatite-coated Ti13Nb11Zr alloy in an animal model. Biomaterials. 2008;29:1730–6.
Mouzin O, Soballe K, Bechtold JE. Loading improves anchorage of hydroxyapatite implants more than titanium implants. J Biomed Mater Res. 2001;58:61–8.
Lee JH, Lee CK, Chang BS, Ryu HS, Seo JH, Hong KS, Kim H. In vivo study of novel biodegradable and osteoconductive CaO–SiO2–B2O3 glass–ceramics. J Biomed Mater Res. 2006;77:362–9.
Acknowledgment
This study was supported by a Grant-in-Aid from the Korean Ministry of Knowledge Economy/KEIT (No. 10033623, Improvement of bone quality surrounding insertion hole for dental implant).
Conflict of interest
No benefits of any form have been received or will be received from any commercial party related directly or indirectly to the subject of this article.
Author information
Authors and Affiliations
Corresponding author
About this article
Cite this article
Lee, J.H., Nam, H., Ryu, HS. et al. Bioactive ceramic coating of cancellous screws improves the osseointegration in the cancellous bone. J Orthop Sci 16, 291–297 (2011). https://doi.org/10.1007/s00776-011-0047-1