Abstract
Background
In Y-stent-assisted coil embolization for cerebral aneurysms, open or closed cell stents are used. Different microcatheters for coil insertion are available. We investigated which microcatheter could be navigated into an aneurysm through a Y-stent with different stents.
Methods
Double Neuroform open-cell stents or double Enterprise closed-cell stents were deployed in Y-configuration in a silicon model of a bifurcation aneurysm. Two endovascular neurosurgeons independently tried to navigate an SL-10 microcatheter for 0.010” coils or a PX Slim microcatheter for 0.020” Penumbra coils into the aneurysm through the Y-stent. In addition, we measured lengths of stent pores of the Y-stents with double Enterprise stents deployed in the model by micro-computed tomography.
Results
It was feasible to navigate an SL-10 microcatheter into the aneurysm through the Y-stent with Enterprise or Neuroform stents. Navigation of a PX Slim microcatheter was feasible in the Y-stents only with Neuroform stents. In the Y-stent with double Enterprise stents, the lengths of the second stent pores were significantly smaller than those of the first stent (0.41 ± 0.18 mm vs 0.69 ± 0.20 mm; P = 0.008). The SL-10 microcatheter was smaller than approximately 80 % of the stent pores of the first stent and 30 % of those of the second stent. The PX Slim microcatheter was smaller than 20 % of the stent pores of the first stent and 0 % of those of the second stent.
Conclusions
It was feasible to insert an SL-10 microcatheter into the aneurysm through Y-stents with Enterprise or Neuroform stents. Navigation of a PX Slim microcatheter for 0.020” Penumbra coils was feasible in Y-stents with Neuroform stents, but not with double Enterprise stents. The measurements of stent pores by micro-computed tomography supported this feasibility study. These results may be helpful to select appropriate stents and microcatheters in Y-stent-assisted coil embolization, especially in case of retreatments.
Similar content being viewed by others
References
Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2004) A novel endovascular treatment of a wide-necked basilar apex aneurysm by using a Y-configuration, double-stent technique. AJNR Am J Neuroradiol 25:509–512
Lozen A, Manjila S, Rhiew R, Fessler R (2009) Y-stent-assisted coil embolization for the management of unruptured cerebral aneurysms: report of six cases. Acta Neurochir (Wien) 151:1663–1672
Rohde S, Bendszus M, Hartmann M, Hähnel S (2010) Treatment of a wide-necked aneurysm of the anterior cerebral artery using two Enterprise stents in “Y”-configuration stenting technique and coil embolization: a technical note. Neuroradiology 52:231–235
Zhao KJ, Yang PF, Huang QH, Li Q, Zhao WY, Liu JM, Hong B (2012) Y-configuration stent placement (crossing and kissing) for endovascular treatment of wide-neck cerebral aneurysms located at 4 different bifurcation sites. AJNR Am J Neuroradiol 33:1310–1316
Kono K, Shintani A, Okada H, Terada T (2013) Preoperative simulations of endovascular treatment for a cerebral aneurysm using a patient-specific vascular silicone model. Neurol Med Chir (Tokyo) 53:347–351
Akgul E, Aksungur E, Balli T, Onan B, Yilmaz DM, Bicakci S, Erman T (2011) Y-stent-assisted coil embolization of wide-neck intracranial aneurysms. A single center experience. Interv Neuroradiol 17:36–48
Chalouhi N, Jabbour P, Gonzalez LF, Dumont AS, Rosenwasser R, Starke RM, Gordon D, Hann S, Tjoumakaris S (2012) Safety and efficacy of endovascular treatment of basilar tip aneurysms by coiling with and without stent-assistance: a review of 235 cases. Neurosurgery 71:785–794
Fargen KM, Mocco J, Neal D, Dewan MC, Reavey-Cantwell J, Woo HH, Fiorella DJ, Mokin M, Siddiqui AH, Turk AS, Turner RD, Chaudry I, Kalani YS, Albuquerque F, Hoh BL (2013) A multicenter study of stent-assisted coiling of cerebral aneurysms using a Y-configuration. Neurosurgery. doi:10.1227/NEU.0000000000000015
Johnson AK, Heiferman DM, Lopes DK (2013) Stent-assisted embolization of 100 middle cerebral artery aneurysms. J Neurosurg 118:950–955
Martínez-Galdámez M, Saura P, Saura J, Martínez A, De Campos JM, Pérez A (2012) Y-stent-assisted coil embolization of anterior circulation aneurysms using two Solitaire AB devices: a single center experience. Interv Neuroradiol 18:158–163
Spiotta AM, Gupta R, Fiorella D, Gonugunta V, Lobo B, Rasmussen PA, Moskowitz SI (2011) Mid-term results of endovascular coiling of wide-necked aneurysms using double stents in a Y configuration. Neurosurgery 69:421–429
Thorell WE, Chow MM, Woo HH, Masaryk TJ, Rasmussen PA (2005) Y-configured dual intracranial stent-assisted coil embolization for the treatment of wide-necked basilar tip aneurysms. Neurosurgery 56:1035–1040
Yavuz K, Geyik S, Cekirge S, Saatci I (2013) Double stent-assisted coil embolization treatment for bifurcation aneurysms: immediate treatment results and long-term angiographic outcome. AJNR Am J Neuroradiol. doi:10.3174/ajnr.A3464
Kono K, Shintani A, Yoshimura R, Okada H, Tanaka Y, Fujimoto T, Tomura N, Terada T (2013) Triple antiplatelet therapy with addition of cilostazol to aspirin and clopidogrel for Y-stent-assisted coil embolization of cerebral aneurysms. Acta Neurochir (Wien) 155:1549–1557
Kono K, Terada T (2013) Hemodynamics of 8 different configurations of stenting for bifurcation aneurysms. AJNR Am J Neuroradiol. 34:1980–1986
Cho YD, Park SW, Lee JY, Seo JH, Kang HS, Kim JE, Han MH (2012) Nonoverlapping Y-configuration stenting technique with dual closed-cell stents in wide-neck basilar tip aneurysms. Neurosurgery 70:244–249
Mascitelli JR, Polykarpou MF, Patel AA, Kamath AA, Moyle H, Patel AB (2012) Initial experience with Penumbra Coil 400 versus standard coils in embolization of cerebral aneurysms: a retrospective review. J Neurointerv Surg. doi:10.1136/neurintsurg-2012-010483
Milburn J, Pansara AL, Vidal G, Martinez RC (2013) Initial experience using the Penumbra coil 400: comparison of aneurysm packing, cost effectiveness, and coil efficiency. J Neurointerv Surg. doi:10.1136/neurintsurg-2012-010587
Conrad MD, Brasiliense LB, Richie AN, Hanel RA (2013) Y stenting assisted coiling using a new low profile visible intraluminal support device for wide necked basilar tip aneurysms: a technical report. J Neurointerv Surg. doi:10.1136/neurintsurg-2013-010818
Conflicts of interest
None.
Author information
Authors and Affiliations
Corresponding author
Electronic supplementary material
Below is the link to the electronic supplementary material.
(MPG 4654 kb)
Rights and permissions
About this article
Cite this article
Kono, K., Terada, T. Feasibility of insertion of a microcatheter through a Y-stent in coil embolization of cerebral aneurysms and its detailed geometry by micro-computed tomography. Acta Neurochir 156, 39–43 (2014). https://doi.org/10.1007/s00701-013-1925-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00701-013-1925-4