Summary
Background
We have previously reported that the intracranial pulse pressure amplitudes were elevated in idiopathic normal pressure hydrocephalus (NPH) patients responding to shunt surgery. Whether or not shunt implantation or adjustment of the shunt valve opening pressure modifies the intracranial pulse pressure amplitudes in NPH patients remains to be established. This report summarises our observations.
Patients and methods
Thirteen patients with NPH (idiopathic in nine and secondary in four) are presented in whom continuous intracranial pressure (ICP) monitoring was done before and after shunt implantation. In two, ICP monitoring was also done during adjustment of shunt valve opening pressure. The mean ICP and mean ICP wave amplitude (i.e. pulse pressure amplitudes) were determined in 6-s time windows.
Results
After shunt implantation there was a fall in both mean ICP and mean ICP wave amplitude; the reduction in the two ICP parameters correlated significantly. However, mean ICP in the supine position was normal (i.e. <15 mmHg) in 12 of 13 patients before shunt placement, and remained normal after shunting. According to our criteria, the mean ICP wave amplitudes were elevated before shunting in 12 of 13 patients and became “normalised” the day after shunting in nine patients. The reduction in mean ICP wave amplitude after shunt was highly significant at the group level. Moreover, adjustment of shunt valve opening pressure modified the levels of mean ICP wave amplitudes.
Conclusions
The present observations in 13 NPH patients indicate that shunt implantation reduces mean ICP wave amplitudes. Moreover, the level of reduction can be tailored by adjustment of the shunt valve opening pressure.
Similar content being viewed by others
References
Bergsneider M (2000) Management of hydrocephalus with programmable valves after traumatic brain injury and subarachnoid haemorrhage. Curr Opinion Neurol 13:661–664
Bergsneider M, Yang I, Hu X, McArthur DL, Cook SW, Boscardin WJ (2004) Relationship between valve opening pressure, body position, and intracranial pressure in normal pressure hydrocephalus: paradigm for selection of programmable valve pressure setting. Neurosurgery 55:851–859
Boon AJW, Tans JTJ, Delwel EJ, Egeler-Peerdeman SM, Hanlo PW, Wurzer HAL, Avezaat CJJ, deJong DA, Gooskens RHJM, Hermans J (1998) Dutch normal-pressure hydrocephalus study: randomised comparison of low- and medium-pressure shunts. J Neurosurg 88:490–495
Cook SW, Bergsneider M (2002) Why valve opening pressure plays a relatively minor role in the postural ICP response to ventricular shunts in normal pressure hydrocephalus: modelling and implications. Acta Neurochir Suppl 81:15–17
Czosnyka M, Pickard JD (2004) Monitoring and interpretation of intracranial pressure. J Neurol Neurosurg Psychiatry 75:813–821
Eide PK (2003) Quantitative analysis of continuous intracranial pressure recordings in symptomatic patients with extracranial shunts. J Neurol Neurosurg Psychiatry 74:231–237
Eide PK (2006) A new method for processing of continuous intracranial pressure signals. Med Eng Phys 28:579–587
Eide PK (2006) Intracranial pressure parameters in idiopathic normal pressure hydrocephalus patients treated with ventriculoperitoneal shunts. Acta Neurochir (Wien) 148:21–29
Eide PK, Brean A (2006) Intracranial pulse pressure amplitude levels determined during pre-operative assessment of subjects with possible idiopathic normal pressure hydrocephalus. Acta Neurochir (Wien) 148:1151–1156
Eide PK, Fremming AD (2003) Lack of relationship between resistance to cerebrospinal fluid outflow and intracranial pressure in normal pressure hydrocephalus. Acta Neurol Scand 108:381–388
Malm J, Kristensen B, Fagerlund M, Koskinen L-O, Ekstedt J (1995) Cerebrospinal fluid shunt dynamics in patients with idiopathic adult hydrocephalus syndrome. J Neurol Neurosurg Psychiatry 58:715–723
Sorteberg A, Eide PK, Fremming A (2004) A prospective study on the clinical effect of surgical treatment of normal pressure hydrocephalus: the value of hydrodynamic evaluation. Br J Neurosurg 18:149–157
Zemack G, Romner B (2000) Seven years of clinical experience with the programmable Codman Hakim valve: a retrospective study of 583 patients. J Neurosurg 92:941–948
Zemack G, Romner B (2002) Adjustable valves in normal-pressure hydrocephalus: a retrospective study of 218 patients. Neurosurgery 51:1392–1402
Author information
Authors and Affiliations
Corresponding author
Additional information
Comments
This paper describes further work from this group on the puslatility of the ICP waveform in patients who either have, or may benefit from, a shunt.
Previously they have shown that higher ICP pulse pressure amplitudes may be a useful index for positive shunt response.
This study addressed the question of whether shunt implantation reduces pre- pulse pressure amplitude. Although the numbers in this study are small this does appear to the case even whilst the mean value of ICP remained within normal range throughout.
This methodology appears to give useful clinical information. In comparison to other methods, such as B wave analysis, implementation is simpler and therefore there may be greater potential to have such monitoring and analysis in real time at the bedside.
As the authors point out the goal of finding the optimal shunt valve opening pressure is still elusive. The production of intelligent shunts where not only the opening pressure can be altered but also the ICP waveform externally measured and monitored should improve the way patients with hydrocephalus are managed.
Iain Chambers
The James Cook University Hospital
ICP monitoring and measurement of the mean ICP wave amplitude was performed in 13 patients with Normal Pressure Hydrocephalus before and one day after shunting. As one would expect shunting resulted in a reduction of both ICP and ICP wave amplitude. No relation was found, however, with clinical improvement. As clinical improvement may take weeks or months such a relation was not very likely. Therefore, leaving a pressure transducer inside one day after shunting does not seem useful in clinical practise.More interesting were the results of measuring ICP and wave amplitude during adjustments of the valve opening pressure. By gradual lowering of this opening pressure first a reduction of both parameters was seen later followed by diverging changes of mean ICP and mean ICP wave amplitude permitting fine tuning of the valve opening pressure and clinical improvement. Here the availability of continuous online data is advantage compared to measurement of CSF outflow resistance by an infusion test.
Jos Th J Tans
Medical Center Haaglanden, The Hague
Rights and permissions
About this article
Cite this article
Eide, P.K., Sorteberg, W. Changes in intracranial pulse pressure amplitudes after shunt implantation and adjustment of shunt valve opening pressure in normal pressure hydrocephalus. Acta Neurochir (Wien) 150, 1141–1147 (2008). https://doi.org/10.1007/s00701-008-0138-8
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00701-008-0138-8