Clinical evaluation of stereotactic brain biopsies with an MKM-mounted instrument holder

  • Clinical Article
  • Published:
Acta Neurochirurgica Aims and scope Submit manuscript

Summary

Object. The aim of this study was to assess the clinical usefulness and accuracy of robot-assisted frameless stereotactic brain biopsies with a recently introduced MKM-mounted instrument holder.

Methods. Twenty-three patients with intracranial lesions participated in this study. Depending on the size of the intracranial lesion, fiducials for image-to-patient co-ordinate transformation consisted either of bone screws or adhesive markers. Shortly after surgery, postoperative MRI-imaging was performed to demonstrate the location of the biopsy site. These images were compared with the preoperative images to assess the biopsy localisation error.

Results. Postoperative biopsy sites could be demonstrated in six patients with bone screws and in 14 with adhesive markers. These two subgroups yielded average biopsy localisation errors of 3.3 mm (SD 1.7 mm) and 4.5 mm (SD 2.0 mm) respectively. This difference was not statistically significant. One biopsy was located in a liquefied haematoma. All others yielded pathological tissue. There were two postoperative haemorrhages, of which only one was temporarily symptomatic. There was no mortality in the first 30 days after surgery.

Conclusions. Robot-assisted frameless point-stereotactic techniques represent an alternative to frame-based techniques for the performance of stereotactic biopsies.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Published online September 18, 2003

Rights and permissions

Reprints and permissions

About this article

Cite this article

Willems, P., Noordmans, H., Ramos, L. et al. Clinical evaluation of stereotactic brain biopsies with an MKM-mounted instrument holder. Acta Neurochir 145, 889–897 (2003). https://doi.org/10.1007/s00701-003-0112-4

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00701-003-0112-4

Navigation