Abstract.
A flow injection system was developed for on-line sorbent extraction preconcentration and flame atomic absorption spectrometric determination of cadmium in natural water samples. The non-charged cadmium complex with diethyl-dithiophosphate (DDPA) was formed on-line in 0.1 mol L−1 HNO3 and retained on the hydrophobic poly-chlorotrifluoroethylene (PCTFE) sorbent material. The adsorbed complex was eluted with isobutyl methylketone (IBMK) and injected directly into the nebulizer via a flow compensation unit. All major chemical and flow parameters affecting the complex formation adsorption and elution as well as interference were studied and optimized. By processing 2.4 mL of sample, the enhancement factor was 39 and the sampling frequency was 50 h−1. For 30 s preconcentration time the detection limit was 0.3 µg L−1 and the relative standard deviation at 5.0 µg L−1 Cd concentration level was 2.9%. The calibration curve was linear in the range 0.8–40.0 µg L−1. The accuracy of the method was estimated by analyzing a certified reference material NIST-CRM 1643d (Trace elements in water). Good recoveries were obtained for spiked natural-water and waste-water samples.
Similar content being viewed by others
References
World Health Organization (1998) Guidelines for drinking water quality health criteria and other supporting information, 2nd edn. Geneva, 2: 281
Z Fang (1995) Flow injection atomic absorption spectrometry John Wiley & Sons Ltd West Sussex, England
M F Enríquez-Domínguez M C Yebra-Biurrun M P Bermejo-Barrera (1998) ArticleTitleDetermination of cadmium in mussels by flame atomic absorption spectrometry with preconcentration on a chelating resin in a flow injection system Analyst 123 105 Occurrence Handle10.1039/a705748b
X Long R Chomchoei P Gala E H Hansen (2004) ArticleTitleEvaluation of a novel PTFE material for use as a means for separation and preconcentration of trace levels of metal ions in sequential injection (SI) and sequential injection lab-on-valve (SI-LOV) systems: determination of cadmium(II) with detection by electrothermal atomic absorption spectrometry (ETAAS) Anal Chim Acta 523 279 Occurrence Handle1:CAS:528:DC%2BD2cXns12it74%3D
A N Anthemidis G A Zachariadis J A Stratis (2001) ArticleTitleOn-line solid phase extraction system using PTFE packed column for the flame atomic absorption spectrometric determination of copper in water samples Talanta 54 935 Occurrence Handle10.1016/S0039-9140(01)00362-9 Occurrence Handle1:CAS:528:DC%2BD3MXktFGqsLc%3D
Y P De Pena M Gallego M Valcarcel (1997) ArticleTitleFullerene: a sensitive and selective sorbent for the continuous preconcentration and atomic absorption determination of cadmium J Anal At Spectrom 12 453 Occurrence Handle10.1039/a606990h
Z-L Fang T Guo B Welz (1991) ArticleTitleDetermination of cadmium, lead and copper in water samples by flame atomic-absorption spectrometry with preconcentration by flow-injection on-line sorbent extraction Talanta 38 613 Occurrence Handle10.1016/0039-9140(91)80144-O Occurrence Handle1:CAS:528:DyaK3MXkslyntbs%3D
J Wang E H Hansen M Miro (2003) ArticleTitleSequential injection-bead injection-lab-on-valve schemes for on-line solid phase extraction and preconcentration of ultra-trace levels of heavy metals with determination by electrothermal atomic absorption spectrometry and inductively coupled plasma mass spectrometry Anal Chim Acta 499 139 Occurrence Handle10.1016/S0003-2670(03)00509-9 Occurrence Handle1:CAS:528:DC%2BD3sXpt1KksL4%3D
E L Da Silva E M Ganzarolli E Carasek (2004) ArticleTitleUse of Nb2O5–SiO2 in an automated on-line preconcentration system for determination of copper and cadmium by FAAS Talanta 62 727 Occurrence Handle10.1016/j.talanta.2003.09.014 Occurrence Handle1:CAS:528:DC%2BD2cXhtVKks7g%3D
Z-H Wang Z-P Zhang Z-P Wang L-W Liu X-P Yan (2004) ArticleTitleAcrylic acid grafted polytetrafluoroethylene fiber as new packing for flow injection on-line microcolumn preconcentration coupled with flame atomic absorption spectrometry for determination of lead and cadmium in environmental and biological samples Anal Chim Acta 514 151 Occurrence Handle10.1016/j.aca.2004.03.049 Occurrence Handle1:CAS:528:DC%2BD2cXkt1Cktbw%3D
V Camel (2003) ArticleTitleSolid phase extraction of trace elements Spectrochim Acta Part B 58 1177 Occurrence Handle10.1016/S0584-8547(03)00072-7
R Ma W Van Mol F Adams (1994) ArticleTitleDetermination of cadmium, copper and lead in environmental samples. An evaluation of flow injection on-line sorbent extraction for flame atomic absorption spectrometry Anal Chim Acta 285 33 Occurrence Handle10.1016/0003-2670(94)85005-4 Occurrence Handle1:CAS:528:DyaK2cXhsFantLc%3D
E J Dos Santos A B Herrmann A S Ribeiro A J Curtius (2005) ArticleTitleDetermination of Cd in biological samples by flame AAS following on-line preconcentration by complexation with O,O-diethyldithiophosphate and solid phase extraction with Amberlite XAD-4 Talanta 65 593 Occurrence Handle10.1016/j.talanta.2004.07.025 Occurrence Handle1:CAS:528:DC%2BD2cXpslyjurk%3D
D Kara A Fisher S J Hill (2005) ArticleTitlePreconcentration and determination of trace elements with 2,6-diacetylpyridine functionalized Amberlite XAD-4 by flow injection and atomic spectroscopy Analyst 130 1518 Occurrence Handle10.1039/b508841k Occurrence Handle1:CAS:528:DC%2BD2MXhtV2gsbzE
V A Lemos P X Baliza (2005) ArticleTitleAmberlite XAD-2 functionalized with 2-aminothiophenol as a new sorbent for on-line preconcentration of cadmium and copper Talanta 67 564 Occurrence Handle10.1016/j.talanta.2005.03.012 Occurrence Handle1:CAS:528:DC%2BD2MXpsFyltbk%3D
C R T Tarley M A Z Arruda (2004) ArticleTitleA sensitive method for cadmium determination using an on-line polyurethane foam preconcentration system and thermospray flame furnace atomic absorption spectrometry Anal Sci 20 961 Occurrence Handle10.2116/analsci.20.961 Occurrence Handle1:CAS:528:DC%2BD2cXlt1KltLs%3D
A N Anthemidis G A Zachariadis J A Stratis (2002) ArticleTitleOn-line preconcentration and determination of copper, lead and chromium(VI) using unloaded polyurethane foam packed column by flame atomic absorption spectrometry in natural waters and biological samples Talanta 58 831 Occurrence Handle10.1016/S0039-9140(02)00373-9 Occurrence Handle1:CAS:528:DC%2BD38XotVyqtr4%3D
G A Zachariadis A N Anthemidis P G Bettas J A Stratis (2002) ArticleTitleDetermination of lead by on-line solid phase extraction using a PTFE micro-column and flame atomic absorption spectrometry Talanta 57 919 Occurrence Handle10.1016/S0039-9140(02)00132-7 Occurrence Handle1:CAS:528:DC%2BD38Xksl2gsbs%3D
Z-L Fang S Xu L Dong W Li (1994) ArticleTitleDetermination of cadmium in biological materials by flame atomic absorption spectrometry with flow-injection on-line sorption preconcentration Talanta 41 2165 Occurrence Handle10.1016/0039-9140(94)00204-5 Occurrence Handle1:CAS:528:DyaK2MXjsFeku74%3D
A S Souza W N L dos Santos L C Ferreira (2005) ArticleTitleApplication of Box-Behnken design in the optimisation of an on-line pre-concentration system using knotted reactor for cadmium determination by flame atomic absorption spectrometry Spectrochim Acta Part B 60 737 Occurrence Handle10.1016/j.sab.2005.02.007
A N Anthemidis K-I G Ioannou (2006) ArticleTitleEvaluation of polychlorotrifluoroethylene as sorbent material for on-line solid phase extraction systems: determination of copper and lead by flame atomic absorption spectrometry in water samples Anal Chim Acta 575 126 Occurrence Handle10.1016/j.aca.2006.05.072 Occurrence Handle1:CAS:528:DC%2BD28XnsFOrtL0%3D
Anthemidis A N, Koussoroplis S-J V (2006) Determination of chromium(VI) and lead in water samples by on-line sorption preconcentration coupled with flame atomic absorption spectrometry using a PCTFE-beads packed column. Talanta (in press)
D Pozebon V L Dressler A J Curtius (1998) ArticleTitleDetermination of copper, cadmium, lead, bismuth and selenium(IV) in sea-water by electrothermal vaporization inductively coupled plasma mass spectrometry after on-line separation J Anal At Spectrom 13 363 Occurrence Handle10.1039/a707849h Occurrence Handle1:CAS:528:DyaK1cXisF2kt7g%3D
Author information
Authors and Affiliations
Corresponding author
Additional information
Correspondence: Aristidis N. Anthemidis, Laboratory of Analytical Chemistry, Department of Chemistry, Aristotle University, GR-Thessaloniki 54124, Greece
Rights and permissions
About this article
Cite this article
Anthemidis, A., Karapatouchas, CP. Flow injection on-line hydrophobic sorbent extraction for flame atomic absorption spectrometric determination of cadmium in water samples. Microchim Acta 160, 455–460 (2008). https://doi.org/10.1007/s00604-007-0775-2
Received:
Accepted:
Published:
Issue Date: